Neb-u-Tyke Speedster ® Nebulizer Compressor Item JB0112-067 User Manual Read this manual before operating the nebulizer. Save this manual for future reference. Info: The most current version of this manual can be found online at www.grahamfield.com. Federal Law (USA) restricts this device to sale by or on the order of a physician.
INTRODUCTION This user manual contains important information and safety precautions for the John Bunn Neb-u-Tyke Speedster Nebulizer Compressor. Before using your Nebulizer Compressor, please read and understand this entire user manual. Take special note of all safety precautions marked DANGER and WARNING. Save this user manual for future reference.
IMPORTANT SAFETY PRECAUTIONS When using electrical products, especially when children are present, always follow basic safety precautions, including the following: DANGER: Indicates a potential hazard situation or unsafe practice that, if not avoided, will result in death or serious personal injury. WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in death or serious personal injury.
Page 5
WARNING: To reduce the risk of burns, electrocution, fire or personal injury: 1. Never leave this device unattended when plugged in. 2. This product contains small parts that may present a choking hazard to small children, and the tubing may present a strangulation hazard. Always use close supervision when this device is used by or near children or those who require close supervision.
Page 6
13. This device has a polarized plug (one blade is wider than the other). As a safety feature, this plug will fit into a polarized outlet only one way. If the plug does not fit fully into the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. Do not modify the plug in any way. Do not attempt to defeat this safety feature. 14. Unplug the device before filling or cleaning the nebulizer. 15. This device is oil-free. Do not lubricate. 16. Do not remove the jet nebulizer while the device is operating. 17.
Neb-u-Tyke Speedster mouthpiece Nebulizer Compressor T adapter Info: Neb-u-Tyke Speedster replacement parts are available at www.grahamfield.com or from your Graham-Field authorized distributor. Item No. Description JB77200 Neb-u-Tyke Speedster Nebulizer Compressor complete with Garage Bag JB0112-067 Neb-u-Tyke Speedster Nebulizer Compressor without Garage Bag...
SETUP Info: Before initial operation, clean the nebulizer cup assembly following the instructions given in MAINTENANCE/Cleaning the Nebulizer Cup. WARNING: Ensure that the nebulizer compressor is unplugged during setup. 1. Before each use, inspect the compressor and nebulizer cup assembly for damage or wear; replace as needed. 2.
Page 9
7. Insert mouthpiece into large end of T adapter T adapter. mouthpiece Info: The reservoir tube, shown in the illustration earlier in this manual, acts as assemble a holding tank for the mist, reducing the mouthpiece and T adapter amount lost during treatment. Its use is optional.
OPERATION s NOTICE: The Neb-u-Tyke Speedster Nebulizer Compressor is designed for intermittent use only. Do not operate it continuously for more than 30 minutes for a single use without turning it off and allowing a cooling period of least 30 minutes.
Safety Overload s NOTICE: Do not exceed 30 minutes of continuous operation. 1. The motor of this device has a built-in thermal overload protector. Should the motor overheat, the protector will automatically shut off the motor. Should this occur, move the ON/OFF switch to the OFF position and allow the motor to cool down for approximately 30 minutes before turning it on again.
MAINTENANCE Cleaning the Compressor’s Outer Case WARNING: Electrical shock hazard. Do not remove the outer case of the device. All disassembly and maintenance of the device must be performed by a qualified service technician. Refer service to qualified service personnel. WARNING: This device does not require oil. Do not attempt to lubricate any internal parts.
Replacing the Filter s NOTICE: Change the filter once every 30 days or when filter turns gray. Info: Additional filters can be purchased at www.grahamfield.com or from your Graham-Field authorized distributor. 1. Ensure that ON/OFF switch is in the OFF position. Open the filter cap located next to the ON/OFF switch. 2.
TROUBLESHOOTING If your Neb-u-Tyke Speedster Nebulizer Compressor fails to function, consult the following Troubleshooting guide. If the problem persists, consult your Graham-Field authorized distributor. Problem Cause and Solution Device doesn’t Check if plug is properly fit into a grounded electrical outlet.
LIMITED WARRANTY SCOPE OF WARRANTY GF Health Products, Inc. (“GF”) warrants to the original purchaser only that it will replace or repair components, at GF’s sole discretion, that are defective in material or workmanship under normal use and service. All warranties are conditioned upon the proper use of the products strictly in accordance with good commercial practice and applicable GF instructions and manuals, including proper use and maintenance.
Page 17
ENTIRE WARRANTY, EXCLUSIVE REMEDY AND CONSEQUENTIAL DAMAGES DISCLAIMER THIS WARRANTY IS GF’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. GF MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
Page 18
Manufactured for: GF Health Products, Inc. 2935 Northeast Parkway Atlanta, Georgia 30360 USA Telephone: 770-368-4700 Fax: 770-368-2386 www.grahamfield.com Made in China...
Page 19
Neb-u-Tyke Speedster ® Compresor de Nebulizador Modelo JB0112-067 Manual de Usuario Lea antes de operar el nebulizador. Guarde este manual para referencia en el futuro. Info: La más reciente versión de este manual podrá ser encontrado en www.grahamfield.com. La Ley Federal (USA) restringe este aparato para ventas por la orden de su médico.
INTRODUCCIÓN Este manual del usuario contiene información importante y precauciones de seguridad para el John Bunn Neb-u-Tyke Speedster Compresor de Nebulizador. Antes de usar su Compresor de Nebulizador, favor de leer y entender el manual de usuario. Tome nota especial en todas las precauciones marcadas PELIGRO y ADVERTENCIA.
DIRECTIVAS IMPORTANTES DE SEGURIDAD PELIGRO: Indica una situación de peligro o una practica insegura que, si no es evitada, resultará en muerte o heridas serias a la persona. ADVERTENCIA: Indica una situación de peligro o una practica insegura que, si no es evitada, puede resultar en muerte o heridas serias a la persona.
Page 23
ADVERTENCIA: Para reducir el riesgo de quemadura, electrocución, fuego o heridas a la persona. 1. Este producto nunca debería ser desatendido mientras esté conectado. 2. Este producto contiene piezas pequeñas que presenta riesgos de asfixia para niños, y la manga puede presentar riesgo de estrangulación. Supervisión cercana es necesa- ria cuando este producto esté...
Page 24
12. Conecte este producto solamente en un enchufe de pared correcto. 13. Este aparato tiene un enchufe polarizado (un aspa es más ancha que la otra). Para reducir el riesgo de electrocución, este enchufe es diseñado para poder ser posicionado de una sola manera en una toma eléctrica polarizada.
Info: Neb-u-Tyke Speedster piezas reemplazadas están disponibles en www.grahamfield.com o de su distribuidor autorizado de Graham-Field. Modelo Descripción JB77200 Neb-u-Tyke Speedster Compresor de Nebulizador completo con Bolso del Garaje JB0112-067 Neb-u-Tyke Speedster Compresor de Nebulizador sin Bolso del Garaje JB76220 Bolso del Garaje solamente JB0112-067-INS-LAB-RevC14...
ENSAMBLAJE Info: Antes de operación inicial, limpie la taza del nebulizador siguiendo las instrucciones escritos en la sección Mantenimiento/Limpieza de la Taza del Nebulizador. ADVERTENCIA: Antes de conectar el cordón de electricidad, asegure que el interruptor de ON/OFF está en la posición OFF.
Page 27
7. Conecte la boquilla a la parte adaptador T grande del adaptador T. boquilla Info: La manga de reserva, en la ilustración superior del manual, es ensamblaje de como un tanque de retención para adaptador T y boquilla reducir la cantidad de medicamento perdido durante tratamiento.
OPERACIÓN s AVISO: El Neb-u-Tyke Speedster Compresor de Nebulizador está diseñado para uso solamente por periodos. No debe operarlo continuamente por más de 30 minutos en un uso sin apagarlo y dejar que se enfríe por lo menos 30 minutos.
Seguridad en caso de sobrecarga s AVISO: No debe exceder más de 30 minutos de operación continuo. 1. Este compresor contiene un dispositivo de protección termal. Si el motor comienza a sobrecalentarse, se apagará automáticamente. Si esto ocurre, posicione el interruptor a la posición OFF y permita que el motor se enfrié...
MANTENIMIENTO Limpieza de la Carcasa del Compresor ADVERTENCIA: Riesgo de electrocución. No remueve carcasa externa de este aparato. Todo desensamblaje y mantenimiento de este aparato debe ser realizado por un técnico de servicio cualificado. ADVERTENCIA: Este aparato no necesita aceite. No lubrique piezas internas. ADVERTENCIA: Desenchufe el aparato antes de limpiar.
Instrucciones para el cambio del filtro de aire s AVISO: Cambie el filtro cada 30 días o cuando el filtro tenga un color gris. Info: Filtros adicionales pueden ser comprados en www.grahamfield.com o de su distribuidor autorizado de Graham-Field. 1. Remueva la tapa del filtro de aire localizado al lado del interruptor ON/OFF.
DETECTANDO PROBLEMAS Si su Neb-u-Tyke Speedster Compresor de Nebulizador falla, consulte a la guía de Detectando Problemas debajo. Si el problema continua, llame a su distribuidor autorizado de Graham-Field. Problema Causa y Solución El aparato no Verifique si el enchufe está conectado correctamente en la toma funciona eléctrica apropiada.
Cada usuario debe ser preparado en la operación y seguridad de este producto antes de uso. Dentro las reglas normales en este documento, el John Bunn JB0112-067 Compresor Nebulizador y sus componentes a estar libre de defectos de materiales o mano de obra según listado debajo:...
Page 35
GARANTÍA ENTERO, REMEDIO EXCLUSIVO Y DESCARGO DE DAÑOS CONSIGUIENTE ESTA GARANTIA ES LA UNICA GARANTIA DE GF Y ES EN VEZ DE TODAS OTRAS GARANTIAS, EXPRESIDAS O IMPLICITADAS. GF NO HACE GARANTIAS ENTENTIDOS DE CUAL QUIER MANERA INCLUYENDO GARANTIAS ENTENDIDOS DE COMERCIALIZACION O IDONEIDAD PARA UN PROPOSITO PARTICULAR.
Page 36
Fabricado para: GF Health Products, Inc. 2935 Northeast Parkway Atlanta, Georgia 30360 USA Telephone: 770-368-4700 Fax: 770-368-2386 www.grahamfield.com Fabricado en China...
Need help?
Do you have a question about the Neb-u-Tyke Speedster and is the answer not in the manual?
Questions and answers