20. WIT-601A Series syringe pump cannot be used in parallel with other brands of pumps. 21. The alarm sound pressure range of WIT-601A Series pump is 47-71 db, and the noise in the operating environment cannot exceed 60 db.
Page 5
24 The infusion pump is apply to 20d/ml, 60d/ml tubing, if the tubing type was set in Abnormal, there is a big risk of accuracy deviation. Please make sure the tubing type settings are correct! If not be adjusted based on different tubing type and different drug liquid, it may cause large deviations in infusion accuracy.
Chapter 1 Overview 1.1 Scope of application It is applicable to accurate, constant and continuous intravenous infusion in medical institutions. 1.2 Technical Parameters IV Set Options /Size: Open System for common widely-used brand of 10d/ml、15d/ml、20d/ml、60d/ml (3.4mm~4.5mm), Also can be customized for dedicated IV set Function Option With Drug Lab.
Page 7
to save & query Self-Test safety real time self-test function to ensure the pump working properly Bubble Detected by ultrasonic sensor with adjust setting 5 level, 50μL ~ 800μL, Infrared sensor to detect the number of drops, monitoring the status in case the empty Drop Sensor bottles Bubble, Door open, Finish, Pause timeout, Down Occlusion, Up Occlusion, Lower...
Main information displayed on LCD Battery status Working status Machine status bar Keypad lock status Infusion set info AC status Flow rate Sound status Infusion status VTBI Pressure status Time needed Volume infused (1) Machine status bar: battery status icon, sound status icon, networking status icon, AC supply status icon, infusion status icon, pressure status icon, IV Set brand, specification, and fluid warmer temperature.
Chapter 3 Assembling Packing list Please check the follow contents when unpacking: Infusion pump body……………………………... 1 set (1) Drop sensor……………………………………… 1 set (2) Power cord……………………………………… 1 set (3) User’s manual………………………………….. 1 copy (4) Pump Installation Horizontal installation on bracket Figure 5 - 10 -...
Page 12
Vertical installation on bracket Installation on cylinder - 11 -...
Page 13
Tips: infusion pump can be placed and used on a flat top, such as bedside cabinet top; do not move a working pump. Infusion pump socket description (1) Plug the corresponding plugs into the rear sockets of WIT-601 series infusion pump according to the icon shown on pump back. Notes: if the drop sensor is not connected, “drop sensor failure will...
IV Set Installation (1) Under normal conditions, insert an ordinary PVC IV set into the outlet of infusion bottle/bag. (2) Slightly squeeze Murphy’s dropper with fingers, to make its 1/3 volume fill with liquid (see figure 9) Notes:To ensure normal drip, never fill the dropper with water, otherwise the infusion pump cannot detect drip drops and will cause alarm;...
Chapter 4 Operation Function key description Function keys concentrate upon the front housing, operated easily, the functions are as follows: Description Function key On/Off Used for boot and shutdown, long press for 1 second to boot, and long press 1 second to shut down on the condition of no infusion.
Operation Flowchart Infusion operational processes Pump installation Power on IV Set installation Parameters setting Purge bubbles Venous puncture Start infusion Power On Connect the power wire, then the light indicator lights up, the AC power is connected, long press the power button for 3 seconds to start up.
Volume mode operation Infusion mode selection Press in stop, display infusion mode interface. If the present mode is volume mode, so not need to select mode. Choose “volume mode” with up or down key. Press , then return to infusion interface after volume mode is chosen. ...
Rate/Time Move the cursor to corresponding input boxes of rate and time with the up or down key. Set rate. Set time. Press Simple Rate Move the cursor to VTBI input box with the up or down key, long press for 1 seconds.
Program mode Custom can himself define 8 stages of program mode to infuse automatically. Allow to suspend programming in process of infusion, but KVO would be launched during programming pause. If the rate of previous program is less than 1ml/h, then the KVO begins according to the rate of previous program; if the rate of previous program is greater than 1ml/h, then the KVO begins according to the rate of 1ml/h.
Page 20
Pause stage program Move the cursor to corresponding input boxes of time, VTBI and rate with the up or down keys. Set time needed to pause. Set VTBI to zero. Set rate to zero. Press ...
Dose/weight mode Infusion mode selection Press in starting up status, display infusion mode interface. Use the up and down keys to choose “weight mode”. Press key. Notes:If the present interface is weight mode, can press directly to enter parameters input interface of weight mode.
Import the drug library Connect the pump with PC by cable and use WIT software to manage the drug library Export & Import Export the drug library Connect the pump with PC by cable and use WIT software to manage the drug library Export & Import ...
Page 23
Input drug concentration. Press down key. Input weight (if needed). Press down key. Input VTBI, Press key. Press to begin drug mode infusion. The operation of unknown drug concentration Drug selection. Use up or down keys to choose drug needed, press ...
Chapter 5 Special functions Bolus Auto Bolus Parameters setting of Bolus Bolus Press in process of infusion Set Bolus rate Set Bolus VTBI Press to begin Bolus. Return to normal infusion after Bolus completed. Manual Bolus ...
IV set management IV set selection Infusion set management Infusion set selection Brand selection Choose infusion set type Exit to main interface Press Choose “IV Set” menu, press Choose “Brand” menu, press Choose IV Set brand needed, press ...
Page 26
IV set calibration Prepare for calibration Because there is a large size deviation among different brands or different types of infusion set, should employ 0.9% saline or pure water to calibrate, set the infusion rate to normal infusion rate, like certain department often employs the infusion rate as 60ml/h-120ml/h, then can take the median 90ml/h as the calibration rate, so can guarantee the accuracy in the condition of normal infusion after calibration, the steps for calibration as follows:...
Page 27
Interface of calibration proceeds Input real output Press confirm key Press Choose “IV Set” menu, press Choose “Adjust” , press Choose “Current IV Set” or “Other”, press . (if choose other tube types, complete tube type selection according to the demands) ...
Choose IV Set Input characteristic value and press confirm Long press the upward key. Input the code, press Input the characteristic value and press to complete calibration. Press to exit to calibration interface. Add IV set If the brand of tube type actually used is not in the brand list of machine, user can add infusion set brand by tube type add operation.
Remove IV Set If need to remove needless tube type brands in the tube type list, user can remove infusion set brand through tube type remove operation. The operation as follows: Choose menu Choose brand need be deleted Prompt “delete or not” Prompt “deletion completed” Choose “Delete brand”...
Chapter 6 General operation Stop Press to stop infusion. Power off Press to stop infusion. Press until the screen off. Unplug the AC power cord. Standby Press “ ” in stop status, until the screen turns dark. ...
Clear the VI Move the cursor to the VI box, press . The VI is reset. Battery life check In process of infusion, can check the runtime of battery under present infusion rate. Press Choose “battery life”, press ...
Page 32
press key again.then the keypad will unlock, System parameters setup System parameters in pause status Press Choose “system parameters”, press Sensitivity setting of up 3 level options: high, median and low occlusion detection When not easy to prompt occlusion alarm but it please ...
Page 33
Time setting Set system time of machine Can input year, month, date; hour, minute, second Brightness adjustment of Use up and down keys to adjust the brightness TFT setting System parameters check or setting in process of infusion Press ...
Input the date need to be checked, press Choose the item need to be checked, press Press to exit. Export history record Connect the pump with PC by cable and use WIT software to manage the history record Export & Import - 33 -...
Chapter 8 Security characteristic and alarm Trouble shooting Infusion level Alarm message Possible failure Check and repair stop Door open High Pump door is open Close the pump door and lock. Infusion set is not properly Properly install the infusion set and make sure it is installed mounted to the slot of bubble detector Flick the infusion tube or press Bolus key to purge the...
Page 36
Can close pressure pre-alarm if not needed Check for strong electromagnetic interference Bottom pressure sensor is around, re-start infusion pump after removal of interfered Abnormal down High the source of interference pressure Bottom pressure sensor Pressure sensor damaged, contact ...
Check Checkpoints Check procedure stage Cracks, broken, Check whether the housing of WIT-601 series infusion pump is intact, whether Before use deformed cracks, broken and deformed exist. Boot and do not connect AC, the icon“ ” flashes and accompanies light-sound ...
Chapter 10 Electrical specification AC supply AC supply 100 V - 240 V ~/50-60 Hz Class I AC power Maximum power 15VA Fuse of protection 2 x 1A Fuse Battery power supply Battery 9.6V 2000mA –nickel-metal hydride battery specification Weight About 150g...
Chapter 11 Storage and transport Product size, weight L x W x H 130 x120x195mm Weight 2.0Kg Screen size 3.2 inches Storage environment Environment temperature: -20℃~55℃; relative humidity: 10%~95%; atmospheric pressure: 70kPa~ 106kPa。 Forbidden to store in the following environments: ...
User's Guide and Manufacturer’s Statement of Anti-electromagnetic Interference: apply to all the equipments and systems WIT-601A Infusion Pump is intended to be used in the following specified electromagnetic environments. Buyers or users should ensure that the pump is used in such environments.
Page 42
User's Guide and Manufacturer’s Statement of Anti-electromagnetic Interference: applicable for all non-life-sustaining equipment and systems WIT-601A Infusion Pump is intended to be used in the following specified electromagnetic environments. Buyers or users should ensure that the pump is used in such environments.
Page 43
WIT-601A Infusion Pump is higher than RF compliance level, it is necessary to check whether the pump runs normally. If not, compensation measures may be necessary, such as re-orientation and positioning for the infusion pump.
Infusion accuracy characteristic Horn-shaped curve represents the changes of maximum and minimum bias in infusion pump. The testing programs for these results have been included in EN60601-2-24 (IV set Drag Heart 20 drops/ml standard PVC set). The following graphs and curves show the overall characteristics of the infusion pump in tests. For more information, please contact the supplier.
System statement WIT – 601 can be attached to other medical electronic equipments, the person who composes this medical electronic system should be responsible for the requirement of compliance. The equipments connected with WIT – 601 should comply with their respective IEC or ISO standards (i.e.