Page 1
Model 8001 Automated Canister Sampler Operations and Maintenance Manual Version 2.01 February 19, 2016 2016 ATEC, Malibu, CA 90265-2112...
Page 3
The implied warranties of fitness for any particular purpose are expressly limited to a term of one year. Under no circumstances shall ATEC be liable to purchaser or any other person for any incidental, special, or consequential damages, whether arising out of breach of warranty, breach of contract, or otherwise.
RUNNING THE SYSTEM Hardware Setup Display and Function Tabs Date and Time Site Identification Canister Sampling Information Purge Flow Abort Function Tabs PROGRAMMING THE MODEL 8001 Time/Date Setup Schedule Leak Check Data Power Failures Manual Mode Advanced Mode TRIGGERED SAMPLING OPTION...
Page 6
Mass Flow Controller Calibration MODEL 8001 PARTS LIST APPENDIX A – PROCEDURE FOR PRESSURE TRANSDUCER AND FLOWMETER CALIBRATION APPENDIX B INSTRUCTION MANUAL FOR MASS FLOW CONTROLLER Atmospheric Technology...
1. Introduction This manual covers the installation, operation and routine maintenance of the ATEC Model 8001 Automated Canister Sampler. This instrument is a microcomputer controlled sampler that can be programmed to draw ambient air into canisters for subsequent analysis of Volatile Organic Compounds (VOC’s) according to the methods outlined in EPA Methods TO-14A and TO-15.
A microcomputer with a touchscreen display provides control and data acquisition for the automated sampling. The Model 8001 can also be ordered with an optional pressurization pump. The pump is equipped with a by-pass valve to regulate the pressure. The purge pump is...
Page 10
not used in this application because the pressure pump will purge the flow system prior to sampling. 2.1.2.1 Solenoid Valves One 15VDC solenoid valve is used for each port. The normally closed valve is mounted between the sampling manifold and the canister. There is one solenoid valve which isolates the manifold from the inlet line for leak checking and inlet line purging.
Page 11
2.1.2.4 Mass Flow Controllers A mass flow controller with a nominal range of 0-10 cc/min is used to control the sampling flowrate in each canister. If the optional collocated Channel 2 and Channel 3 are present, each channel would have a dedicated mass flow controller usually in the 0-10 cc/min range.
The normal display will illuminate after the 20 second startup. The Model 8001 uses a 7.0” SVGA LCD color touch screen display to show current operating status and to enter information into the computer.
Model 8001 Version 2.00 Time/Date Setup Schedules Data Leak Check SOP Main Manual Advanced Monday, November 26, 2001 18:58 East River Site Flow Pressure Status 11.8 Canister Ch. 1 Port 1 Finished Port 2 11.5 Finished Sampling Port 3 3.67 10.40...
Sampling—Indicates a channel is sampling Waiting—Indicates a channel has been programmed to sample Purging—Indicates a channel is purging prior to sampling Idle—Indicates a channel has not been programmed to sample Aborted—Operator terminated sampling Finished—A successful sample has been collected without errors Used Can—Canister initial pressure above limit Press Tol—Canister pressure did not meet expected pressure within a given tolerance...
Advanced allows the changing of operational parameters. 4.1 Time/Date The time and date that is stored in the Model 8001 can be changed using the Time/Date tab. When this tab is touched, the control panel screen will appear which enables the changing of time, date, time zone, and daylight savings time adjustment.
Page 16
The nominal canister volume in liters is need to determine the flow rate. Used Can Limit The Model 8001 checks the pressure of each canister prior to sampling. pressure specified in the used can limit is used to determine if a used or leaking canister has been installed.
Page 17
Site Label Touching the site label box brings up an alphanumeric keypad that can be used to input a site specific label (e.g. East River Site No. 2) that appears on the main screen. Remote IP Address If the sampler is configured with the Network Option, the IP address of the remote tablet or PC that is to connect to the sampler must be entered in this box.
to terminate the routine and to save the new settings. The touch screen should be re-calibrated whenever it becomes difficult to activate buttons on the screen. Pressing the Set button saves the data that was entered in the Set-up screens and returns the program to the main screen.
time, duration, and trigger option can be entered. If the port sampling time is to follow immediately after the previous port and is of the same duration, the information can be entered automatically by checking the box labeled “Follow the previous port”.
“Yes” to the query that appears when exiting the Data screen. The Next button on the 8001 Data screen will advance to the next data screen or Exit can be used to return to the main screen.
4.7 SOP The SOP tab provides a Standard Operating Procedure for installing, leak checking, and scheduling the Model 8001. The SOP software guides the user through the required procedures and enables inexperienced operators to successfully program the sampler.
Advanced tab. This portion of setup should only be used by individuals familiar with the operation and calibration of components used in the Model 8001. Changing the values in the Advanced Setup will alter the accuracy of the instrument.
Page 23
Number of Ports (password control) The Model 8001 has 8, 16, or 32 ports. If an expansion module is added, the numbers of ports can be increased up to 64. The total number of ports must be specified here. If the Model 8001 does not detect the expansion module, the number of ports will default to 8.
5. Triggered Sampling Option If the triggered sampling option is installed, sampling can be initiated independently for each port/ channel by a switch closure on the rear panel connector shown below. To use the trigger option, in Schedules or SOP the Triggered Operation check box must be selected for each port/channel to be triggered.
Page 25
When a trigger is sensed, the following can occur: If the pump is off (i.e. nothing is being sampled), the port/channel will start sampling after establishing MFC zeros (1 minute of time) and purging the inlet lines. The duration of purging is specified in Setup. If the purge time is set to zero and the pump is off, sampling will start one minute from the time of the trigger to allow the collection of MFC zero data.
Page 26
6. Network Connection If the Network Option has been included in the sampler, the following steps will needed to be taken to enable communication between the sampler and a PC or tablet on the same Local Area Network (LAN). Initial Network Setup Before communication is started, the IP addresses of the sampler and the remote PC/tablet must be determined.
Page 27
4. On the Sampler, enter the Remote IP Address in the appropriate text box. 5. On the Sampler, press the Enable Network button to start communication. 6. On the PC/tablet, start VNC Viewer. Enter the “Local IP Address” in the box labeled “VNC Server:”...
Page 28
These steps do not have to be done again. If a different PC/Tablet is to be used, the “Remote IP Address” will have to updated in the sampler. If communication is no longer needed, clicking the “Disable Network” button the the Setup screen will terminate communication.
7. Maintenance This section describes routine maintenance procedures for the Model 8001 Canister Sampler. All other repair work should be performed by trained personnel. 7.1 Pressure Pump To remove the pressure pump from the sampler, remove the top cover. Remove the inlet tube and disconnect the electrical connector.
Page 30
7.4 Pressure Relief Valve Adjustment The pressure relief valve controls the output pressure of the pump. The following steps can be used to change the pressure setting. 1. Insert Allen wrench slightly into set screw inside outlet and turn counter clockwise to unlock.
Page 31
8. Model 8001 Parts List Components of the Model 8001 are shown in the following Parts List. This list does not include common parts such as fuses, fittings, or fasteners. Chassis 8001-100-1 Front Panel 8001-100-2 Rear Panel 8001-100-3 Mass Flow Controller (Specify Range)
Page 33
Appendix A – Procedure for Pressure Transducer and Flowmeter Calibration Flowmeter Calibration Procedure 1. Enter the Manual screen and without the pump running, observe the average flow controller flow rate (the flow rate will oscillate). If it is not 0.00±0.02, manually adjust the flow controller output by turning the "zero"...
Page 34
Pressure Transducer Calibration Procedure 1. For the pressure transducer to be calibrated, set the slope to 1.0 and the intercept to 0.0 in the Advanced option. 2. Attach an evacuated canister (absolute pressure < 0.05 psia) to a port and in Manual mode open the port valve to evacuate the manifold and equilibrate it to the canister pressure.
Page 35
Appendix B Instruction Manual for Mass Flow Controller Atmospheric Technology...
Page 36
TELEDYNE HASTINGS INSTRUMENTS INSTRUCTION MANUAL 200/202 SERIES FLOWMETERS/CONTROLLERS I S O 9 0 0 1 C E R T I F I E D...
Page 37
Manual Print History The print history shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between revisions, contain important change information that the user should incorporate immediately into the manual.
1. General Information The Hastings HFM-200 mass flow meter and HFC-202 flow controller are designed to accurately measure and control mass flow over the range of 10 sccm to 30 slm, without corrections or compensations for gas pressure and temperature with an accuracy of better than ±1% FS. Hastings mass flow instruments do not require any periodic maintenance under normal operating conditions with clean gases.
1.4. Optional 4-20 mA Current Output An option to the standard 0 - 5 VDC output is the 4 - 20 mA current output that is proportional to flow. The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flow meter. The current loop output is useful for remote applications where pickup noise could substantially affect the stability of the voltage output or long cable runs where cable resistance would cause a voltage signal to decay.
1.5.2. Interconnecting Cables Cables are available from Hastings, in various lengths, to connect from the 15 pin "D" connector on the back of the Power Pod directly to any of the 200 series and 300 series flow instruments (including digital versions).
2. Installation and Operation This section contains the steps necessary to install a new flow meter/controller into operation as quickly and easily as possible. Please read the following thoroughly before attempting to install the instrument. 2.1. Receiving Inspection Carefully unpack the Hastings HFM-200/HFC-202 series instrument and any accessories that have also been ordered.
2.4. Mechanical Connections 2.4.1. Standard Configuration The flow meter may be mounted in any position as long as the direction of gas flow through the instrument follows the arrow marked on the bottom of the flow meter case label. The preferred orientation is with the inlet and outlet fittings in a horizontal plane (if operating with a dense gas or at high pressures the instrument must be installed horizontally).
EXAMPLE: Suppose a 4” LFE is selected. The length of 4” diameter tubing or pipe, upstream of the sensor inlet tap, will be a minimum of 20“. The minimum length of 4” tubing or pipe downstream from the sensor outlet tap will be 4”. The length of additional 4” inlet tubing required for the 4”...
24 Volt Connections Refer to the diagram to the right when connecting 24 Volt units. Connect the positive lead of the power supply to pin 7 of the DE-9 connector and negative lead to pin 4. The supply input is diode protected such that reversing the input polarity will not damage the instrument.
2.6.2. Zero Check Turn the power supply on if not already energized. Allow for a 1 hour warm-up. Stop all flow through the instrument and wait 2 minutes. Caution: Do not assume that all metering valves completely shut off the flow. Even a slight leakage will cause an indication on the meter and an apparent zero shift. For the standard 0 - 5 VDC output, adjust the zero potentiometer located on the lower outlet side of the flow meter until the meter indicates zero.
The “COMMAND” pot adjusts the Analog command signal sent to the flow controller. The setting for each controller connected to the power supply can be observed. (Depending on how the power supply was set up, the display could indicate in flow units or percent of full scale). Hastings Power Supply Feature Guide Feature Model 40...
2.7.3. Operation with an external sensor (Fig. 2.2) In some instances, it might be desirable to use an external Fig 2.2 sensor to provide process information to the control circuitry in the flow controller. For example, you might want to control the pressure in a vacuum system by adjusting the rate at which the system is backfilled with a gas.
controller circuit establishes control at the new command point. This jumper does not affect the system gain and will not dampen out oscillations. To adjust the response, you need a means of producing a step change in the command voltage from 10% of full scale to 100% of full scale.
3. Theory of Operation This section contains a functional description of Hastings flow controllers. Detailed schematics and parts lists can be obtained by contacting Hastings using the contact information found at the end of this document. In this section and other sections throughout this manual, when a power supply is mentioned, it is assumed that the customer has a Hastings Power Supply.
relatively constant over wide ranges of temperature and pressure, the flow meter may be calibrated directly in mass units for those gases. Changes in gas composition usually only require application of a simple multiplier to the air calibration to account for the difference in heat capacity and thus the flow meter is capable of measuring a wide variety of gases.
3.5. Valve The control valve is an “automatic metering solenoid” valve. While most solenoids operate in either the fully open or fully closed state, the automatic metering solenoid valve is designed to control flow (see Figure 3.5). A spring, connected to the plunger assembly, holds a magnetic plunger tightly against an orifice to shut off flow.
4. Maintenance This section contains service and calibration information. Some portions of the instrument are delicate. Use extreme care when servicing the flow controller. 4.1. Authorized Maintenance With proper care in installation and use, the flow controller will require little or no maintenance. If maintenance does become necessary, most of the instrument can be cleaned or repaired in the field.
CAUSE: Plugged orifice. ACTION: Verify the presence of a 10-50 psig pressure across the instrument. If present, shut off gas supply and power supply. Remove orifice per Section 4.9. Examine orifice. If plugged, clean or replace as applicable. Reassemble valve. SYMPTOM: Flow meter reads other than 0.00 VDC with no flow, or there is a small flow when flow meter reads 0.00 VDC.
4.3.2. Miscellaneous adjustments Periodically, during normal operation, the ZERO should be checked and adjusted when required. If the instrument is not shutting completely off when the Valve Override CLOSE function is active, or a command of zero flow has been given, then the orifice may require approximately 1/8 turn clockwise. 4.4.
B) If P >2P , use formula 1; otherwise use formula 2. C) Use a consistent set of units for pressure, flow, and density (i.e all lengths, masses, times in the same units, cm, ft, kg, sec etc.) 1 liter/minute = 1.667 x 10 /sec, 1 gm/liter = 1 kg/m , 1 psia = 6895 kg/m*sec...
Page 58
Lubricate the O-rings slightly with a silicone based grease, and the threads with anti-galling compound. Push the orifice into its hole and screw it in until it is flush with the instrument base. Apply pressure to the inlet side of the instrument. Enable the Valve Override CLOSE function or unplug the instrument.
5. Warranty and Repair 5.1. Warranty Repair Policy Hastings Instruments warrants this product for a period of one year from the date of shipment to be free from defects in material and workmanship. This warranty does not apply to defects or failures resulting from unauthorized modification, misuse or mishandling of the product.
Need help?
Do you have a question about the 8001 and is the answer not in the manual?
Questions and answers