TABLE OF CONTENTS Specification Installation Rack Mounted Cooling Filament Module Electrical Power Input Grounds Power Supply to Filament Module Filament Transformer Connection Power Supply Control Connectors Remote I/O Control 2.10 Remote Interface 2.11 Remote Wiring 2.11.1 Gun Control 2.11.2 High Voltage Control 2.11.3 Customer EPO Controls and Operation...
Page 4
TABLE OF CONTENTS CONTINUED Gun Operation Using Control Panel 3.8.1 Gun Interlocks 3.8.2 Gun Ready 3.8.3 Gun On 3.8.4 Gun Supply Metering 3.8.5 Gun Control Indicators 3.8.6 Remote Operation "Gun Supply" 3.8.7 Gun "Position" Control 3.8.8 Bias Control 3.8.9 Emission Current Control Theory Of Operation Overall Control "High Voltage"...
SPECIFICATION INPUT POWER Model CV-6SL 208V Input Voltage : 280V±10% Frequency : 50/60Hz Phase : Three - 4 Wire Current : 32 Amps Model CV-6SL 400V Input Voltage : 400V±10% For operation from 0ºC to 40ºC. Input as low as 342 VAC may be accommodated but unit must be operated at less than 35ºC due to reduced fan...
Page 7
Indicators : KV Overvoltage Fault : KV Overcurrent Fault : KV Out of Reg : ARC Counter Fault : Rail Fault : Step Start Fault : PWR OK : HV ON : Gun ON : PS Fault Main Circuit Breaker : Emergency Off (mechanical latch)
All the power required for the system inputs at this point. A mating plug is provided for the customers input power wires, four female pins and cable clamp are included as part of the assembly. The CV-6SL can be ordered for the following input voltage sources. MODEL #CV-6SL 208V #6024-6010...
Compatibility with the controls and monitoring of the original CV-6S is provided via J101 (HV control) and J103 (Gun Control) connectors. The local controls are separated into a high voltage control section and a gun control section. All cables described in the next two sections are available as separate items from the CV-6SL...
Remote I/O control Ribbon DB-25 Name (Ref) Input +24 VDC (from Interlock chain). Connect To Momentary (Normally Open GUN ON + Switch) Output from Momentary contact to +24VDC (Normally Open Switch). Momentary Close to GUN ON - turn Gun On Input +24 VDC (from Interlock chain).
2.10 Remote Interface: The remote interface designed and supplied with the CV-6S may be connected to the CV-6SL using the J-101 and J-102 cables: HV CONTROL GUN CONTROL 25 COND. W101 15 COND. W102 RIBBON RIBBON J-101 J-103 POWER SUPPLY CHASSIS...
Page 14
Connecting the Gun Control-System I/O Cables(s) (PN B-2-371-184: 10 Ft.) (PN B-2-371-225: 20 Ft.) Follow the procedure described below in connecting the gun control system I/O cable(s). Refer also to page 11A shows the pinout for the signals that can be interfaced to the gun control unit via its system I/O connector (G1-J13).
2.11 Remote Wiring 2.11.1 Gun Control Digital Interlock Inputs (Required) INTERLOCK WIRE FUNCTION TANK 13, 14 O, BK Prevents the gun from being switched on unless all vacuum system doors and covers are closed and locked so that personnel cannot come into contact with high voltage within the vacuum cubicle.
FILAMENT 26, 27 BN, BK This feature is not connected in the CV-6SL. BIAS SET The only Bias control is on the front panel of the CV-6SL Digital Outputs SIGNAL J2 PIN # WIRE FUNCTION PAIR GUN READY 1, 2...
Digital Outputs SIGNAL HV-J11 WIRE FUNCTION PAIR HV READY 1, 5 R, BK Indicates that all HV control interlocks are made and that the HV GO OFF*/ENABLE loop is closed; the HV is ready to be switched on. HV IS ON 2, 6 W, BK Indicates that the high voltage is on.
10 Turn Pot accessed through Recessed opening. The potentiometer is set at a current below the normal operating current. The bias functions as a preheat or idle for the filament. This is the only control that will affect the bias setting on the CV-6SL. 3.1.4 Emergency Off Switch: When pushed, opens the interlock and immediately turns off the HV and Gun.
High Voltage "Ready" The high voltage power supply is in a "ready" to turn on condition when the interlock chain is closed and the internal faults are reset. The internal faults circuits located on the High Voltage regulator circuit board 9120028-010 are summed and if the Control Panel is used, displayed on the high voltage control as HV-I7.
Remote Operation "High Voltage" when using the Control Panel Remote Inputs The remote interface has the following functions available: High Voltage OFF Contact Opening HV-J2 G-BK High Voltage ON Contact Closure HV-J2 BL-BK High Voltage Reference Command 0-10V = 0-10kV Output High Voltage Monitor 0-10V = 0-10kV Output Current Monitor 0-6V = 0-600 Milliamperes H.V.
3.8.2 Gun "Ready" When all the interlocks are satisfied, "Closed", the circuit is ready for the gun to be turned ON. The indicator in the gun off switch, G1-I2, will be illuminated. The circuits that must be satisfied for gun operation are: Remote Off G1-J2 Green-Black...
3.8.5 Gun Control Indicators The LED indicators are powered in a series chain. If the first interlock in the chain is broken, all indicators that follow are not satisfied. The sequence is: Tank G1-I7 Vacuum G1-I6 Auxiliary G1-I5 Water G1-I4 Position G1-I3 When each of the above five interlocks are closed, their respective LED's are illuminated.
#6024 - 6106-0 Filament Supply Without Position Supply 3.8.8 Bias Control G1-R3 The Bias Control does not work with the CV-6SL. The Bias Pot on the Front Panel of the CV- 6SL is the only active Bias control 3.8.9 "Emission Current" Control G1-R2 The emission current control located on the control panel provides a reference for emission current operation.
THEORY OF OPERATION Overall The power supply provides three power sources to the CV-6S System. High Voltage Power - 10kV DC - 600mA Filament Power - 10V AC - 75 Amperes Position Power 15V DC - 3 Amperes (Optional) Each power supply operates independently with the addition of emission current control by the filament supply.
Page 27
The switching between the operational modes is performed automatically on the Gun Supply Circuit Card. The high voltage power supply initiates the switching when the high voltage is turned on and operating in its regulation window. Position Power (Optional) The position supply uses the rail voltage provided on the input to the Gun Supply Card. The position supply is low power, 15 volts at 45 watts.
Control "High Voltage" (Optional) HV-S4 Meter Select Switch: The switch selects between power supply output voltage and power supply total current. The voltage readout maximum is 10.0kV and the current maximum is 600mA. HV-S3 Remote-Local Switch: The switch transfers the following functions to the remote interface: High Voltage ON H.V.
Control "Gun" (Optional) G1-S4 Selector Switch: Provides a readout of the three gun related circuits. The emission current 0-600mA is high voltage current and requires that both the gun and high voltage power supplies be ON. The filament current 0-75.0 amperes provides a readout with Gun ON.
G1-S1 Gun Off: The gun off switch opens to break the latch on the gun ON holding circuit. The switch's function is the same as an interlock opening. G1-I2 Gun Ready: The gun is ready for turn on after closing a chain of interlocks. See Paragraph 5 for Gun On and the list of interlocks.
Page 33
CB-2: The circuit breaker protects the gun power supply. CB-3: The circuit breaker protects the primary of the control step-down transformer T-2. This circuit provides power for all +24V DC and -24V DC in the system. All control power originates through this breaker.
High Voltage Regulator 9120028-010 The regulator for the high voltage power supply is the heart of the power supply. It controls, regulates the output voltage and current, communicates with the outside system, protects the power supply from internal and external faults. There are three interface connectors: J5 Connection to components in the chassis J7 Ribbon cable to the HV inverter power card J4 Ribbon cable to the Gun Supply card...
Page 35
monitor as R-103 on the input to the buffer U-16 Pin 10 provides adjustment. Pin 25 on J-6 connects to the kV readout circuit in the control panel. mA Monitor: The 5 volt signal from the external current monitoring shunt is raised to 10 volts at the output of the mA feedback buffer U-21 Pin 14.
Page 36
Latched Faults: The latched faults are seen in the circuit around U24. The latched faults are: Overvoltage Fault Slow Overload Fault Arc Counter Fault Rail Fault Each latched fault has a sensitivity adjustment and an LED readout on the board that identifies that each respective fault caused the power supply shut down.
+ 15V DC Regulated U-12 - 15V DC Regulated U-14 + 10V DC Regulated U-13 + 5V DC Regulated U-1 mA Feedback Output: An output signal is provided on P1 that is used as a ground referenced power supply current feedback signal to P2 pins 2, 3 on the Gun Card: C-6-1899-98. 0-600 MA power supply current equals 0-6V DC output on P1 pins 2, 3.
feedback signal 7.5V = 75 amperes enters the filament circuit error amp at U-1 Pin 9 by way of R-25. The reference voltage originates with the bias potentiometer on the gun control panel and enters the board on P-1 Pin 1. This circuit controls the filament with high voltage off.
Page 39
The high voltage output PCB assembly contains – Tuning Capacitors C-22 to C-31 Doubler Capacitors C-5 and C-6 Filter Capacitor C-7 Feedback Divider C-9 and R-2 With Series Res R-6 Output Series Resistors R-1A, 1B, 1C, 1D, 1E, 1F Low End Current Shunt R-3A, 3B The metering signals connect to jack J-4.
Filament Transformer Module The filament power supply provides an A.C. voltage for filament operation. The frequency of operation is 34 kHz. The filament transformer provides 10 volts at 75 amperes maximum to the filament for operation. The filament transformer has a single turn secondary providing the power. The 40 mills thickness of copper connects to the two brass studs located in the G-10 output plate.
Interconnect Cables: There are five interconnect cables in the power supply chassis that make connection between the circuit boards and the circuit boards and the jacks on the rear panel. Make sure they are in place and making connection. There are two shield ribbon cables connecting between the power supply chassis and the control. Make sure the connections are in place and bolted in position.
The gun can be turned ON at anytime (before or after high voltage turn ON). Emission current regulation will not take place until the high voltage is turned ON and in regulation. Make sure all the interlock lights are illuminated. Contact closures are required on the external interface.
Page 78
PARTS LIST CV-6SL 400V PART # DESCRIPTION 6024-6009-0 POWER SUPPLY, CV-6SL 400V 1-GUN BRC-10-600N-380-3PH-50-SATIS POWER SUPPLIES INCLUDE: * FILAMENT SUPPLY PCB, CV-XS (W/OUT POSITION SUPPLY) (C-6-1899-98-D) * ONE-GUN HV OUTPUT PCB (Made up of 2 bds, the 48 bd & 56 bd)