Only the surface echo is displayed.
The sensitivity is poor.
The picture does not change even if the Gain or Far Gain Control dial is turned.
Is TX POWER D set?
Is TVG CURVE setting changed
from the user setting?
Does Picture Display on the
screen change by turning
Gain/Far Gain Control dial?
                                                                                                             
TR PCB J1-No.2: 12V & J1-No.5: 70V
(or 85Volt in case Power Supply
                                                                                                                         
Are the output waveforms are normal?
Are the output waveforms
normal when replacing
Transducer with Dummy?
*
Soundome
* DUMMY RESISTANCE
120Ω/0.0057μF   
(180kHz) 
Use the over 10W and 1.5kV pressure-resistant dummy resistance.
Use the dummy resistance just for the confirmation of the output waveform.
Do not flow the current for a long time. Otherwise it will cause fire.
NO
YES
NO
YES
NO
YES
NO   NO
PCB-85V)
YES
          YES
NO
NO
YES
160Ω/375μH
   (80kHz)
Set TX POWER D in
FUNCTION SET, MENU2.
Reset TVG curve setting in
FUNCTION SET, MENU2.
* To be continued to the next page.
* OUTPUT WAVEFORM 
  180kHz:500Vp-p
(or 600Vp-p in case Power Supply PCB-85V)
80kHz:535Vp-p
(or 650Vp-p in case Power Supply PCB-85V)
-15-
Hull Unit portion
Power Supply PCB
TR PCB
(Transducer−GND)