AED 5070-ABE Operation Manual

AED 5070-ABE Operation Manual

For use with philips heartstart models: onsite/home (m5066a, m5068a) frx (861304)

Advertisement

5070-ABE Battery
For use with Philips™HeartStart™
Models: OnSite/Home (M5066A, M5068A)
FRx (861304)
Operation Manual
Revision 3.5

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 5070-ABE and is the answer not in the manual?

Questions and answers

Summary of Contents for AED 5070-ABE

  • Page 1 5070-ABE Battery For use with Philips™HeartStart™ Models: OnSite/Home (M5066A, M5068A) FRx (861304) Operation Manual Revision 3.5...
  • Page 2 Capacity: When new a minimum of 200 shocks or 4 hours of operating time at 25°C (77°F). Actual battery operating life depends on device settings, usage, and environmental factors. Charge Time: A new battery typically takes 8 seconds to charge the AED to maximum energy. Batteries with reduced capacity will require additional time to charge the AED.
  • Page 3 Use one of these solutions to clean the plastic housing of the battery: soapy water, denatured ethanol, or 91% isopropyl alcohol. If the AED indicates that the battery is low and needs to be replaced please recycle or dispose of the lithium battery in accordance with all federal, country, state, and local laws. Help our environment,...
  • Page 5 Troubleshooting If the AED begins to chirp with a flashing “i” button, it is NOT necessarily an indication that the battery is low. Press the “i” button for the AED to identify why the device is chirping. The pads are not tested on battery installation so often times if the AED begins to chirp within a few weeks following the installation of a new battery it is an indication that the pads need to be replaced.

Table of Contents