Supplementary instructions
External housing
VEGAWAVE
Document ID: 32357

Advertisement

Table of Contents
loading

Summary of Contents for Vega VEGAWAVE

  • Page 1 Supplementary instructions External housing VEGAWAVE Document ID: 32357...
  • Page 2: Table Of Contents

    Setup ..........................15 Maintenance ........................... 16 How to proceed if a repair is necessary ................16 Dismount..........................17 Dismounting steps......................17 Disposal ......................... 17 Supplement ..........................18 Technical data ........................ 18 Dimensions ........................20 Editing status: 2021-04-13 External housing • VEGAWAVE...
  • Page 3: About This Document

    Symbols used Document ID This symbol on the front page of this instruction refers to the Docu- ment ID. By entering the Document ID on www.vega.com you will reach the document download. Information, note, tip: This symbol indicates helpful additional infor- mation and tips for successful work.
  • Page 4: For Your Safety

    That is why we have introduced an environment management system with the goal of continuously improving company environmental pro- tection. The environment management system is certified according to DIN EN ISO 14001. Please help us fulfil this obligation by observing the environmental instructions in this manual: • Chapter " Storage and transport" • Chapter " Disposal" External housing • VEGAWAVE...
  • Page 5: Product Description

    The instrument version "External housing" consists of the instrument Components - external housing with direct cable housing. outlet The instrument housing consists of the base element, the screwed cover for the electronics or connection compartment and the housing base. External housing • VEGAWAVE...
  • Page 6 The metal sensor housing cannot be opened. The electrical connec- tion must be carried out on the instrument housing. The instrument housing consists of the base element, the screwed cover for the electronics or connection compartment and the housing base. External housing • VEGAWAVE...
  • Page 7 The housing base is produced in two different materials. The version depends on the selected material of the instrument housing. • Instrument housing of plastic - housing base: plastic • Instrument housing of stainless steel - housing base: stainless steel External housing • VEGAWAVE...
  • Page 8 To make it easier to guide the cable through cable ducts or cable rout- ing chains, for example, the plug of the short cable is supplied loose. The short cable is always 1.5 m (4.92 ft) long, the long cable can be selected in several fixed lengths. External housing • VEGAWAVE...
  • Page 9 Cable plug, short (enclosed) Pin 1 - rd Pin 2 - ye Pin 3 - bk Pin 4 - gn Pin 5 -gn/ye Fig. 5: Plug mounting Plug insert Cable Coupling sleeve Seal Clamping basket Pressure screw External housing • VEGAWAVE...
  • Page 10: Principle Of Operation

    Dispose of the packaging material via specialised recycling companies. • Storage and transport temperature see chapter " Supplement - Storage and transport temperature Technical data - Ambient conditions" • Relative humidity 20 … 85 % External housing • VEGAWAVE...
  • Page 11: Mounting

    1. Mark the holes according to the following drilling template housing 2. Depending on the mounting surface, fasten the wall mounting plate with 4 screws 90 mm (3.54") 70 mm (2.76") 8 mm (0.12") (0.32") Fig. 6: Hole pattern - wall mounting plate (external housing) External housing • VEGAWAVE...
  • Page 12: Mounting - External Housing (Instrument Housing)

    With the plastic housing, the four screws of the base may only be screwed in hand tight. Exceeding the max. torque specified in chapter " Technical data" can damage the wall mounting plate. Mounting - external housing (instrument housing) → Mount the sensor to the bottom, depending on the process fitting Note: When mounting the sensor, note the instructions in the operating instructions manual of the sensor. External housing • VEGAWAVE...
  • Page 13: Connect The Sensor To The External Housing

    1 2 3 4 Fig. 7: Connection of the sensor in the socket housing Yellow Black Green Green/yellow (ground terminal) 2. Connect the cables to the terminals in the socket housing. External housing • VEGAWAVE...
  • Page 14 3. Tighten the compression nut of the cable entry gland. The seal ring must completely encircle the cable The electrical connection of the electronics module is described in the operating instructions manual of the sensor. External housing • VEGAWAVE...
  • Page 15: Setup

    6 Setup Setup Setup Setup is carried out according to the operating instructions manual of the respective sensor. External housing • VEGAWAVE...
  • Page 16: Maintenance

    Clean the instrument and pack it damage-proof • Attach the completed form and, if need be, also a safety data sheet outside on the packaging • Please contact the agency serving you to get the address for the return shipment. You can find the agency on our home page www.vega.com. External housing • VEGAWAVE...
  • Page 17: Dismount

    Pass the instrument directly on to a spe- cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive. External housing • VEGAWAVE...
  • Page 18: Supplement

    Depending on the sensor Ambient temperature on the sensor -20 … +80 °C (-4 … +176 °F) housing Ambient, storage and transport tempera- -40 … +80 °C (-40 … +176 °F) ture on the instrument housing External housing • VEGAWAVE...
  • Page 19 Electrical protective measures Protection rating Ʋ Sensor housing IP66/IP67 (NEMA Type 4X) Ʋ Instrument housing - Instrument IP66/IP67 (NEMA Type 4X) socket Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF. External housing • VEGAWAVE...
  • Page 20: Dimensions

    9 Supplement Dimensions External housing - direct cable outlet 90 mm (3.54") 110 mm (4.33") ~ 66 mm (2.6") Fig. 8: External housing with direct cable outlet to the vibrating element External housing • VEGAWAVE...
  • Page 21 ~ 59 mm (2.32") 69 mm (2.72 ") ø 80 mm (3.15") ø77 mm (3.03") M20x1,5/ ½ NPT M20x1,5/ ½ NPT Fig. 10: Housing versions - instrument housing Instrument housing - plastic Instrument housing - stainless steel, electropolished External housing • VEGAWAVE...
  • Page 22 Notes External housing • VEGAWAVE...
  • Page 23 Notes External housing • VEGAWAVE...
  • Page 24 Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2021 VEGA Grieshaber KG Am Hohenstein 113 Phone +49 7836 50-0 77761 Schiltach E-mail: info.de@vega.com...