Contact Information ® C O N G R A T U L A T I O N S on your ProSun purchase! Please make sure you read this manual carefully before using this equipment and specifically the maintenance schedule. Keep this manual available for future reference. If you have any questions please contact ProSun directly.
Table of Contents Important Customer Information ................5 Who Should Use, Read, & Understand this Manual ........5 Indications For Use …………..…............5 Intended Use ..…........….......... 5 Warning Signs & Symbols …................. 5 Preparing the Product For Use ................8 Export ….......................
Page 4
How to Operate the Equipment ……………………………………………………….. 34 Choosing the Correct Exposure Time …………………………………………… 34 Normal Equipment Operation & Software …………………………………………… 35 Operating Modes ………………………………………………………………..35 Standby Mode ………………………………………………………...... 35 Tanning Mode ………………………………………………………...... 35 Cool Down Mode ………………………………………………………………..35 Service Mode ………………………………………………………………………… 35 Starting the Tanning Session ……………………………………………………..
Any service work outside of the preventative care table in this manual should be done by a ProSun authorized technician. This is a Class I ME Equipment, with Type B Applied Parts used for continuous use for a maximum of 9 minutes.
Page 6
High Voltage exists and is dangerous for human contact. UV Radiation hazard for Human Health Caution Symbol means to proceed with caution and follow instructions careful ESD-Sensitive devices, do not touch without using necessary proper precautions. Type B Applied Parts Symbols found on equipment and packaging: Follow and read instructions thoroughly prior to using equipment.
Page 7
Protective earth connections. Recyclable Material Polystyrene Cardboard Mixed Paper Wood Waste of electrical and electronic equipment must not be disposed as unsorted municipal waste. It must be collected separately, and must be disposed as per local regulations. Alternating Current. Single Phase Three phase alternating current.
ProSun does not assume any liability if there is non-compliance outside of the intended sale territories of the United States or Canada. The exporter assumes their own risk by removing this equipment outside of the United States or Canada.
Coordinate with ProSun installation department to set up delivery and installation details. In a commercial setting an external timer is required. ProSun recommends to use this accesory (T-Max Flushmount G2).
Page 10
ProSun and all authorized parties shall reserve the right to refuse to install equipment into any room which has not been properly prepared in accordance with the electrical specification and room requirements.
Commissioning of the tanning equipment The equipment is initially commissioned by a certified ProSun technician. The device will be properly assembled and handed over ready for us. In order to properly commission the tanning equipment, the business owner must ensure that the location is ready prior to the equipment being installed.
General Safety Information General Safety Notes: Any adult with any of the below issues should not use this tanning equipment. If you are unsure consult your doctor prior to use. Contraindication: This sunlamp product is contraindicated for use on persons under the age of 18 years.
Page 13
Allow at least 48 hours to pass between each tanning session. Tan based on the exposure sticker on the unit or in this manual. Skin Type I should not tan. As an owner you must properly instruct all users on using the exposure table located on the front of the bed and only tanning for the minutes based on their skin type.
Page 14
This tanning bed is an over the counter tanning device that emits ultra violet light in the UVB and UVA region of the spectrum intended for the cosmetic tanning of the human skin in adults. Only one user should tan in the unit at a time. Wear protective eyewear during the duration of the sesion to protect the eyes against UV radiation.
The Tanning Process The tanning unit you have purchased is designed to concentrate optimal levels of UVA, in conjunction with very low percentages of UVB, on the outermost layers of the skin, so as to stimulate the production of Melanic pigment, which is slightly pink in its dormant state, these causes it to turn brown after excretion.
Tanning Sensitivity Before using a tanning device always check with your doctor first. *The following may increase skin sensitivity: • Perfumes and colognes containing furocoumarins. • Compounds from natural products such as plants and fruits (their natural oils can sensitize the skin to sunlight.
Components 3 step body cooler XL Light for Extra long 2 meter lamps. Touch Screen Sound system with pre- installed speakers, MP3 integration. The tanning device consists of 4 panels ( 2 main panels and 2 door panels) and a control column in the center. In the diagram above, the different components have been outlined and described.
Be sure to always have these labels on your equipment, if they become ilegible or are missing contact ProSun for replacement. Labels on units are differently sized than in descriptions below. Each label is listed in the following pages for clear description of what is written on each label.
Page 21
2. V6 Label The V6 label is attached to the front right panel on the outside of the unit. 3. 18 and Older Label Age warning label stating no one under the age of 18 should use this product. This label is located on the front of the unit visible to the user.
Page 22
6. Danger Label 5. Warning Label This label is located on the front of the This label is located on the front of the unit unit on the door panel, visible to the on the door panel, visible to the user. user.
Page 23
8. Serial Label The serial label is located on the back of the unit and is the identifier for the unit. This includes the serial number and conformity marking. The below is a sample serial label, please check your equipment for your equipment specifications. 9.
Technical Data Electrical Specifications Luxura V6 48 Ultra Intensive Power consumption rating: 32 Amp Electrical Rating: 230V 3~ 50A 60Hz Rated Fuses: 50 amp 3-pole Continuous Operation: Continuous operation for a maximum of 9 minutes. Classification: Type B Applied Parts...
Please insure that your electrician and/or installer, reads this user’s manual carefully prior to the installation of your ProSun unit. You will need a buck booster if voltage requirements are not met. Service personnel should be authorized by ProSun to perform work on equipment and should have knowledge of equipment and read the full manual.
Electrical Connections Live Components Risk of fatal injury due to electrical shock when touching live parts. DANGER! Risk of electrical Shock To avoid the risk of electrical shock, this equipment must only be connected to a supply mains with protective earth. WARNING! Both the power supply and the related connection work must comply with national standards.
Tanning Bed Standards FILL IN STANDARDS INFORMATION HERE This tanning equipment is in conformity with performance standards 21 CFR 1040.20 and the standards for electromagnetic compatibility (EMC): IEC 60601-1-2 (2014) Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The Luxura V6 48 Model 48V6 is intended for use in the electromagnetic environment specified below.
Page 28
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity The Luxura V6 48 Model 48V6 is intended for use in the electromagnetic environment specified below. The customer or the user of the Luxura V6 48 Model 48V6 should assure that it is used in such an environment.
Page 29
“Product Name” should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the “ProSun Tanning Equipments. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
Page 30
Recommended separation distances between portable and mobile RF communications equipment and the Luxura V6 Model 48V6 Non life supporting ME Equipment The Luxura V6 48 Model 48V6 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Luxura V6 48 Model 48V6 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the V6 48 Model 48V6 as recommended below, according to the maximum output power of the communications equipment.
Before your tanning session Check for any conditions mentioned in the manual that may prevent you from using the tanning equipment. Make-up and cosmetics Do not wear make-up when tanning. Clean skin absorbs the UV Light properly. Remove all make-up prior to tanning. Make-up clogs the pores in your face. The UV light from a tanning device re-opens these clogged pores.
If used the warranty will be voided! ProSun® recommends the use of Lucasol One Step when cleaning acrylic sheets. Acrylics in tanning equipment are extremely porous and should not be cleaned/disinfected with non- approved products.
Operation of Equipment Obligations of the business operator • As the business operator you are responsible for informing the user of the risks of tanning. • All danger and safety labels should be attached to the equipment and clearly visible to the user.
How to operate the equipment Prior to operating the equipment check that the unit has been cleaned and check for any damage to the acrylics. No other people should be in the vicinity of the equipment when in use, especially no children under the age of 18.
Normal Equipment Operation and software Before turning on the equipment make sure you have proper warning signs at your location. We recommend permanently placing a Warning sign on the wall close to the tanning bed such as the one below: WARNING •...
Tanning Mode The following description is for a unit that has all possible features, some of these features may not be available for your unit or market. DANGER! Excessive dose of UV radiation Excessively prolonged tanning sessions may cause severe damage to skin and eyes.
Stopping the Tanning Session If the time in the session runs out naturally then the unit will go into a 3 minute cool-down. However if you need to stop the tanning session prior to end of the sesion you can do so by hitting the START/STOP Button.
Customer Concerns Failure to use protective eyewear may result in severe burns or other eye injury. If discomfort develops, stop using the tanning equipment immediately and seek medical advice from your physician. • Do not continue to use the tanning device if unexpected effects, such as itching, occur within 48 hours after the first tanning session.
Waste of electrical and electronic equipment must be disposed as unsorted municipal waste. It must be collected separately, and must be disposed as per local regulations. Contact ProSun International LLC or an authorized representative for information concerning the decommissioning of your equipment.
Control Functions Not all options shown below are applicable on this device and in the US Market. Function Description Additional Information Time Remaining This option will show the time After the session is over this will also remaining in the session show the cool down time Start/Stop Switches the UV lamps on and off...
Note: If you need to switch between timer and stand-alone mode If an external timer or coinbox does not have a ‘remote start’ function please contact ProSun or a certified electrician if you need help. and instead the appliance must be started with the start/stop button, this can be configured by your dealer.
Page 42
BODY COOLING Press the body cooling button. The LED above the button begins to flash. The LEDs above plus/minus will flash. Set the desired value. The standard value is 1. After 5 minutes, you can't turn off the body cooling. VOLUME (NOT AVAILABLE IN ALL MARKETS) Press the volume button.
Page 43
Service Mode When the unit is in standby mode, you can enter into the service or maintenance parameters. STATUS MENU Hold the body cooling and plus buttons for five seconds. You now see the unit status menu. The hour counters and software versions can be viewed here. RESETTING THE HOURS COUNTERS Select LH.
PARAMETER MENU Hold the body cooling and minus buttons for ten seconds. Enter the PIN code; it is set to 000 by default. You can change the value of the digit with plus and minus. Press the body cooling button to proceed to the next digit. Press start/stop to confirm the PIN code.
Page 45
DEFAULT Unit Start time (minutes) 0 = Coin unit / 1 - maximum time = Stand alone Maximum time Time at which the decorative lighting is switched on 30 (min.) Special mode 1 = Restore factory settings / 6 = Reset TL hour meter Pin code Unique code for entering the parameter menu Usefull life lamps...
Reading the hour counters and software STATUS MENU Hold the body cooling and plus buttons for five seconds. You now see the unit status menu. The hour counters and software versions can be viewed here. RESETTING THE HOURS COUNTERS Select LH. Hold the start/stop button for three seconds.
When one of the above-mentioned fault messages appears on the display, you should no longer use the tanning appliance and the unit should be disconnected from the power. Contact ProSun for further instrcutions. (This does not apply to the LH message.) If there an error message such as the above shows on your display.
Accidental reactivation can result in serious accidents. Immediately after deenergizing the equipment, secure all switches or fuses used for deenergizing against being switched on again. For more maintenance information please visit our website at: www.prosun.com/customer-service 48 MARK76V6002 Rev 4...
The table below is a preventative care table that can be done by any unskilled person, if any service outside of this table is needed please contact ProSun immediately and this would should only be done by a certified technician.
NOTE: For thorough cleaning and disinfection, the product you are using has to remain in contact with the surface that is being cleaned and disinfected for a specific amount of time, as mentioned on the instructions of the product. The specifc amount of time will depened on the type of cleaner/disinfectant that you are using.
Maintenance by skilled persons: ProSun recommends having an annual inspection by a certified technician to review connections, check for lose wiring, check the power cords and review all other high voltage components. At this time the certified technician can dust all electronic components.
Damage to equipment by user or Contact ProSun for replacement. Non- cleaning skilled person can replace acrylic. Any troubleshooting outside of the above mentioned guide should be done by a certified technician. Contact ProSun to schedule service with a certified technician. 52 MARK76V6002 Rev 4...
Replacement of Components WARNING: Before doing any work on this machine, whether it be an unskilled person or a certified technician, ALWAYS disconnect the power at the main circuit breaker(or, if the main power cable has a wall plug, pull it out). WARNING! IMPORTANT: Only use manufacturer-authorized replacement components.
Removing Acrylics WARNING: Before obtaining access to terminals, all power supply circuits must be unplugged. This can be done by an unskiled person who has read and understood the manual. 54 MARK76V6002 Rev 4...
Replacing the Lamps Remove acrylics before replacing lamps. This can be done by an unskiled person who has read and understood the manual. Lamp replacement: Holding the lamp at both ends, turn it counter-clockwise until the bi-pins can slide easily out of the lamp holder socket, and slowly take it out of the unit.
Assembly Introduction Required for assembly (not included): • Preferably 2 wrenches to tighten bolts and nuts • Short ladder or step stool • Power adapter kit (230V for commercial use, for warranty to apply voltage must be between 218-230 V) *Make sure you write the Serial Number (located on the Control Panel) down on the previous page and store the manual somewhere you can easily access.
Page 57
Symbols Warning symbol: People with long hair should wear a hairnet! Warning symbol: Wear the protective goggles provided Warning symbol: Attention please! Warning symbol: Sunbed should be disconnected from power source Warning symbol: Remove plug from socket Warning symbol: Not for household use Symbol: Facial tanner Symbol:...
PROSUN’s sole discretion. Prior to making any repairs PROSUN will advise the Purchaser of the nature of the problem and the cost to the Purchaser based on the proration of the defective system or component.
Page 81
CONSEQUENTIAL DAMAGES WITH RESPECT TO THE PRODUCTS DELIVERED HEREUNDER OR SUBSEQUENT USE THEREOF. 7. Limitation of Liability: PROSUN shall not be liable for any loss, damages, or penalty resulting from failure to perform any of its obligations under this Agreement due to major force or any cause beyond its reasonable control.
Page 82
ProSun International, LLC 2442 23rd Street North St Petersburg, FL 33713 www.ProSun.com 800-874-2776 Contraindication: This Product Is Contraindicated For Use On Persons Under The Age Of 18 Years. “Contraindication: This Product Must Not Be Used If Skin Lesions Or Open Wounds Are Present.” “Warning: This Product Should Not Be Used On Individuals Who Have Had Skin Cancer Or Have A Family History Of Skin Cancer.”...
Need help?
Do you have a question about the Luxura V6 48 Ultra Intensive and is the answer not in the manual?
Questions and answers