Advertisement

READ AND SAVE THESE INSTRUCTIONS
Operation Manual
110V | 220V
®
STD
1911

Advertisement

Table of Contents
loading

Summary of Contents for Halosil HaloFogger STD 1911

  • Page 1 READ AND SAVE THESE INSTRUCTIONS ® Operation Manual 110V | 220V 1911...
  • Page 2 Walk Around Tour of the HaloFogger® Rear Handle Status Indicator Light High Precision Nozzle Dry-mist of disinfectant is dispensed through nozzle. Digital TImer Enter minutes to set time for fogging treatment. Start Button Press and hold for 2 seconds to begin unit’s countdown and the treatment process.
  • Page 3: Operation Warnings

    Avoid splashing or overfilling which could harm the user and/or the device. 12. Halosil International is not responsible for any injury or damage caused by using this device outside of the specific parameters detailed in this manual.
  • Page 4 1. Plug the HaloFogger into a power source for 24 hours to fully charge the HaloLog’™ Data Logger’s internal back-up battery. 2. Download the HaloMist™ All-Purpose Disinfectant Use Instructions. Go to www.halosil.com, register your new fogger purchase to download. Setup and Filling THE HALOFOGGER IS DESIGNED FOR USE WITH HALOMIST™ ALL-PURPOSE DISINFECTANT AND SHOULD NOT BE OPERATED WITH ANY OTHER MANUFACTURERS DISINFECTING OR CLEANING PRODUCT.
  • Page 5 A. If you have no handheld H2O2 sensor but a written protocol has been established by your organization or an authorized Halosil Distributor for the room or for the class of similar rooms using a handheld H2O2 sensor, use the time established by that protocol.
  • Page 6 HaloFogger® Operation Manual_1911 Special Instructions - Fogging a Room With a Second Room a. When fogging a room with a second interior room (example: bathroom in a hospital patient room), measure the length, width and height of both rooms, and combine totals to determine final room size. b.
  • Page 7 HaloFogger® Operation Manual_1911 How to Use the HaloLog™ Data Logger Compatible with Apple® iDevices and Android® Devices that have Bluetooth 4.0 The HaloLog™ App free of charge to download. For Apple® iDevices, please visit the App Store. For Android® Devices, please visit Google’s Play Store. What does HaloLog™...
  • Page 8 HaloFogger® Operation Manual_1911 Connecting HaloLog™ App to HaloFogger® 1. Plug HaloFogger into a power source. The Status Indicator Light on top of the fogger will illuminate solid GREEN. 2. Using an iOS device with HaloLog App installed, stand within 10’ of a HaloFogger and open the HaloLog App.
  • Page 9 HaloFogger® Operation Manual_1911 Halo List Screen When the HaloLog App is started, the Halo List screen will appear. The the top of the screen will show “Found HaloFoggers” listed by the fogger’s individual serial number. Only HaloFoggers that are within range to the HaloLog App will be displayed.
  • Page 10 HaloFogger® Operation Manual_1911 Run ID Screen In this screen, specific information about the treatment cycle can be added. By tapping on a text block, a room number, building name or location, the name of the operator and additional comments for that run cycle can be entered. For convenience, the last entered building name is stored and will come up as a default for each HaloFogger.
  • Page 11: E-Mail Screen

    HaloFogger® Operation Manual_1911 To remove the DELETED identifcation from a Run Event ACTION: Tap on the RED highlighted Run Event UnDelete ACTION: Tap A confirmation window will appear. Tap ‘Ok” to continue. The RED highlight and word DELETE are removed. to return to Run Event Screen.
  • Page 12: Utility Screen

    HaloFogger® Operation Manual_1911 Email Screen The HaloLog App will create a .CSV file and open up the Email screen to send. The .CSV file is easily manipulated in MS Excel. Copy and paste new records downloaded to an existing file. Each Run ID log can be additionally coded within MS Excel. Utility Screen The Utility Screen allows the data exchange between the HaloFogger and the iDevice to be monitored in real time.
  • Page 13: Maintenance

    HaloFogger® Operation Manual_1911 Maintenance Nozzle Cleaning and Alignment To keep your HaloFogger® working at optimum performance, it is recommended that you clean the nozzle on a regular basis (example: clean nozzle once-a-week if fogger is used once-a-day or more.) Step 1. Before performing any maintenance to the HaloFogger, disconnect the power cord from the electrical outlet.
  • Page 14: Troubleshooting

    HaloFogger® Operation Manual_1911 Troubleshooting The following table of potential indicators provides a quick reference of the descriptions, probable causes and most common solutions regarding the operation of the HaloFogger®. If the information in this table does not describe or solve the problem you are having, contact Customer Service at 1 302-454-8102, Monday through Friday, 9:00AM to 5:00PM EST.
  • Page 15: Replacement Parts

    HaloFogger® Operation Manual_1911 Replacement Parts To order replacement parts, contact Customer Service at 1 302-454-8102, Monday through Friday, 9:00AM to 5:00PM EST. Part Part Number 110V Power Cord (8’) 0750-110VPC 220V Power Cord (8’) 0750-220VPC Nozzle Assembly 0750-000 (includes retainer cap, nozzle insert, main bolt and O-ring)) Re-Order Number...
  • Page 16 Photocopy and post outside of room being treated.
  • Page 17: Warranty And Registration

    Halosil International and its authorized distributors assume no liability for any accident, injury, death, loss or other claim related to or resulting from the use of this device. In no event shall Halosil International or its authorized distributors be held liable for incidental or consequential damages relating to or resulting from the use of this device or any of its parts.
  • Page 18 Learn how to reduce infection rates at your facility by visiting 91 Lukens Drive www.halosil.com. There you can find out more about the Halo New Castle, DE 19720 USA Disinfection System™, download literature and case studies, and find a Tel +1 302-454-8102 distributor near you.

Table of Contents