Lutron Electronics PM-9102 Operation Manual

Lutron Electronics PM-9102 Operation Manual

200 mbar, differential input manometer

Advertisement

Quick Links

200 mbar, differential input
MANOMETER
Model : PM-9102
OPERATION MANUAL
Your
purchase
MANOMETER marks a step
forward for you into the
f i e l d
o f
measurement.
this
MANOMETER
complex
and
instrument,
its
structure will allow many
years of use if proper
operating techniques are
developed.
Please
the following instructions
carefully and always keep
this manual within easy
reach.
of
this
p r e c i s i o n
Although
is
a
delicate
durable
read

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PM-9102 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Lutron Electronics PM-9102

  • Page 1 200 mbar, differential input MANOMETER Model : PM-9102 Your purchase this MANOMETER marks a step forward for you into the f i e l d p r e c i s i o n measurement. Although this MANOMETER complex delicate...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS 1. FEATURES............1 2. SPECIFICATIONS..........1 2-1 General Specifications........1 2-2 Electrical Specifications........3 3. FRONT PANEL DESCRIPTION.......4 3-1 Display............4 3-2 Power Off/On Button........4 3-3 Hold Button..........4 3-4 Zero Button..........4 3-5 " MAX./MIN. " Button........4 3-6 Unit Button....
  • Page 3: Features

    1. FEATURES * Dual & differential input, 200 mbar max. range. * Application : Industrial, laboratory, heating, ventilation, medical hospital, used for air or not corrosive and not ionized gas & liquid. * Sensor is built inside the housing. * Single lugs for pipe connection. * 8 kind display units ( mbar, psi, , mm Hg, inch Hg, meter...
  • Page 4 Display units mbar, psi, , mm Hg, inch Hg, meter , inch , Atmosphere. Function Dual & differential input, data hold, zero/relative, memory. Zero adjust Push button on the front panel. Sensor * Sensor is built inside the housing. * Piezoelectric sensor. * Used for dry, non-corrosive and non-ionic air and gas only.
  • Page 5: Electrical Specifications

    2-2 Electrical Specifications Unit Max. range Resolution mbar ± 200 mbar mbar ± 2.9 0.001 ± 0.204 0.001 mm Hg ± 150 mm Hg mm Hg inch Hg ± 5.906 inch Hg 0.002 inch Hg meter ± 2.040 meter 0.001 meter inch ±...
  • Page 6: Front Panel Description

    3. FRONT PANEL DESCRIPTION Fig. 1 3-1 Display 3-7 P1 input socket 3-2 Power Off/On Button 3-8 P2 input socket 3-3 Hold Button 3-9 RS-232 Output 3-4 Zero Button Terminal 3-5 " MAX./MIN. " Button 3-10 Battery Compartment 3-6 Unit Button /Cover 3-11 Plug/ quick coupler...
  • Page 7: Measuring Procedure

    4. MEASURING PROCEDURE 1) Power on the meter by pressing the " Power Off/On Button " ( 3-2, Fig. 1 ). 2) Select the desired temperature units ( mbar, psi, mm Hg, inch Hg, meter , inch , Atmosphere ) by pushing the "...
  • Page 8: Auto Power Off Disable

    7) Data Record ( Maximum, Minimum reading ) * The DATA RECORD function displays the maximum and minimum readings. To start the DATA RECORD function, press the " MAX./MIN. Button " ( 3-5, Fig. 1 ) once. " REC " symbol will appear on the LCD display. * With the "...
  • Page 9: Rs232 Pc Serial Interface

    5. AUTO POWER OFF DISABLE The instrument has built-in " Auto Power Shut-off " in order to prolong battery life. The meter will switch off automatically if none of the buttons are pressed within 10 min. To de-activate this feature, Select the memory record function during measurement, by pressing the "...
  • Page 10: Battery Replacement

    The 16 digit data stream will be displayed in the following format : D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicate the following status : Start Word D12 & D11 Annunciator for Upper Display mbar = 86 psi = 23...