Panasonic MCO-170AICL Operating Instructions Manual page 75

Co2 incubator

Advertisement

STD GAS AUTO CALIBRATION KIT (OPTION)
5. On the STD Gas Setting screen, input CO
2
density of the connected standard CO
gas cylinder.
2
Press Apply key to save the input value. The
display returns to the Tools #1 screen.
Settable range: 4.0 %~21.0 %
Factory setting: 5.0 %
6. Press STD Gas Calibration key to lead the STD
Gas Calibration screen.
7. On the STD Gas Calibration Step1 screen,
system check starts automatically. If the system is
normal, display leads the STD Gas Calibration
Step2 screen.
8. On the STD Gas Calibration Step2 screen, press
OK key to lead the STD Gas Calibration Step3
screen.
74