CALIBRATION SOLUTIONS • Manual is provided for quick and easy installation and to best utilize PHT-026 • Has large digital Display system for easy reading • A FREE AC POWER ADAPTOR for Australia, UK, US...
Page 3
• Operating Temperature: 0 °C ~ 50 °C • Automatic Temperature Compensation (ATC): 0 °C ~ 50 °C • AC Power Adaptor (for UK, US, Australia or Europe) • Dimensions: 155mm x 86mm x 22mm • Weight: 223g Parts: 1 Display 2 Charger / Power Supply connector 3 ON/OFF switch 4 pH electrode input terminal...
Page 4
If it will be used for the first time, the meter has to be charged first. After plugging in the AC adaptor to the PHT-026 meter, slide the power switch to the left to turn ON the device. Charge the built-in rechargeable battery for two hours before you start using PHT-026.
Page 5
• EC, CF, PPM calibration Press the “MODE” button, choose EC for operation. Immerse the meter in 12880µs/ cm (12.88EC) calibration solution. Allow the reading to stabilize. Use the mini screw driver to turn the CAL Calibration knob until “12.88EC” appears. Rinse the electrode with distilled water, and absorb moisture by using filter-paper.