The Fiagon Navigation System is a system that is used in surgical computer-assisted instrument navigation. Please refer to these instructions for use to operate the Fiagon system correctly and safely. You will find an overview and explanation of all relevant steps to set up the unit that are necessary for a navigated operation.
The navigation system may only be used with the components and accessories that the manufacturer indicates as suitable. The navigation system may only be used with Fiagon Navigation software and accessories that the manufacturer indicates as compatible with the system.
Patient data protection HIPAA Privacy Rule Compliance: Users of the Fiagon Navigation System have the option of assigning a password before system operation in order to prevent unauthorized access to individually identifiable health information.
Read these notes carefully. General safety notes All Fiagon products have been developed and manufactured in compliance with strict safety standards. Nevertheless, a proper use and regular maintenance of the medical system is required. It is indispensable that the users of this system carefully read and pay attention to all risks and safety phrases described in the instruction manual.
60601-1:2005 Clause 16 · Prior to each use, the system should be checked for loose components and damage. DO NOT use if these conditions exist. If service is required, contact your Fiagon Navigation sales representative immediately. · Use of an uninterrupted power supply (UPS) to the device is mandatory to ensure the device remains on during voltage dips and unintended interruptions for interventions that do not allow for delays in the procedure or absence of image guidance.
Do not connect equipment other than storage media to the USB interface. The Network connection must be unplugged from the Fiagon Navigation System during surgery. Make sure Storage media are free of viruses before connecting to the system.
Page 15
The installation in the tower as well as the connection to the monitor and video source are subject to the initial installation. Consult the Fiagon service or your dealer. This procedure is done by Fiagon’s authorized personnel. WARNING: Use of an uninterrupted power supply (UPS) to the device is mandatory to ensure the device remains on during voltage dips and unintended interruptions for interventions that do not allow for additional delay in the procedure or absence of image guidance.
The navigation instrument has the focus. Full screen Video mode This mode is configured in the applications (modules) offered by Fiagon that utilize a video signal of a medical video device (endoscope or microscope). In this mode the connected video signal is displayed in full screen on the display. The mode is activated when no navigation information is available for 3 seconds.
Troubleshooting – unintended power shut down Power shut down of Fiagon navigation system can occur by e.g. unintended removal of power plug, necessary change of the location of the device during surgery, power loss due to voltage dips (Dip>95% over 5s duration) or a power shut down due to electrical failures.
Spilling liquids on any electrical device attached to the system should warrant inspection by an authorized Fiagon service technician. Inspection instructions After cleaning, visually inspect the system under good light conditions for remaining debris. Repeat cleaning procedures if required.
Updates of this instruction manual will be published as additional pages as soon as updated information or corrections are available. Updates will be provided along with a software update of the current version (e.g., update of Fiagon Navigation 3.6.5 to Fiagon Navigation 3.7.5).
The manufacturer acts in compliance with international and national standards and laws. Information about the compliance can be provided on request. Please contact your regional consultant or contact us under the following address: Fiagon GmbH Neuendorfstraße 23b 16761 Hennigsdorf Germany...
Page 29
Instructions for Use EMC compliance statement The Fiagon Navigation System needs special precautions regarding EMC and needs to be installed and put into service according to the following described EMC information. Electromagnetic emission The Fiagon system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.
Page 30
Instructions for Use Electromagnetic immunity The Fiagon system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment. Immunity Test IEC 60601-1-2 Test...
Page 31
Instructions for Use Electromagnetic immunity The Fiagon system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment. Immunity Test IEC 60601-1-2 Test...
Page 32
Recommended separation distances between portable and mobile RF communications equipment and the Fiagon system The Fiagon system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Fiagon system can help prevent...
Page 33
Instructions for Use The Fiagon navigation should only be connected with the following accessories and cables to comply with the standards mentioned above. Interface Type Cable Type / Max. length Fiagon Order No. Remarks accessory Monitor Out/ DVI – D, shielded...
Contact a service representative if you plan to use a different device. Before using the Fiagon system for the first time in clinical routine, it is necessary to check its compatibility with the CT scanner or DVT. If problems arise regarding the loading process of image data sets, please contact your service representative.