Translation of the original instructions OPERATING UNIT Type BE10 BE10V Ex-BE10B Ex-BE10BV Ex-BE10C Ex-BE10CV Please read this manual before starting up. To be retained for future reference.
1. Concerning this manual This manual • is part of this product • is valid for all mentioned series • describes the safe and proper use in all operating stages 1.1 Terms User: Single person or organisation who is using the products e.g. customer, user, assistant 1.2 Target groups Target group Task...
What to do Cross reference → Requirement Tab. 4: Symbols and meaning 1.5 Latest state Please find the latest state of this operating instruction under www.lutz-pumpen.de. 1.6 Copyright The content of this operating instruction and the images contained in them are subject to the copyright protection of Lutz Pumpen GmbH.
2. Modular Flow Meter System TS The flow meter operates according to the positive displacement principle. In a housing fitted with inlet and outlet is a measuring chamber through which the fluid is forced to flow on its way through the housing. In the measuring chamber, a nutating disc blocks the fluid`s clear path. As it flows through the measuring chamber, the fluid pushes the nutating disc to one side. It is the action of displacing the impediment which produces the wobble movement of the disc and circular motion of the nutating disc shaft. By means of a driver, a magnet with several poles is set into rotation, which is able to activate a magnetically switched contact through the lid of the housing.
3. Modular Flow Meter System HDO The flow meters of series HDO are oval gear flow meters. This has proven to be a reliable and highly ac- curate method of measuring flow. Exceptional repeatability and high accuracy over a wide range of fluid viscosities and flow rates are features of the HDO series flow meter design. The low pressure drop and high pressure rating means the HDO series flow meter is suitable for both gravity and pump (in line) applications. For practically-orientated application, the flow meter system is made up of various modules: À Operating unit Electronic evaluating unit with touchscreen for operation of the flow meter. Á Protection cap Protects the modules basic flow meter, relay module or power sup- ply unit when the operating unit is decentrally installed. Pulse converter Amplifies the signals from the basic flow meter. Is required if the signals need to be transmitted further than 5 m.
4. Safety The manufacturer will not be held responsible for any damages resulting from non-compliance of the overall documentation, in particular for damages in case of non-compliance of the operating instructions. 4.1 General safety information Observe following regulations before accomplishing all activities. Product safety The operating unit is constructed in conformity with the state of the art and approved safety-related standards.
Page 9
Personnel qualification • Ensure that personnel tasked with work on the operating unit have read and understood these instructions and all associated applicable documents, particularly safety, maintenance and servicing information, before they start work. • Clarify responsibilities, competencies and monitoring of the personnel. • Ensure that all work is carried out by technical qualified personnel only: - Assembly, servicing, maintenance work - Work on the electrics...
5. Proper use The device can be used in the following environments: - on roofed surfaces - in roofed and closed rooms - altitude up to 2000 m - ambient temperature -10°C up to 60°C - maximum relative humidity 80% - degree of polution 2 NOTICE Device failure due to environmental influences!
5.1 Technical data Type BE10 BE10V Ex-BE10B Ex-BE10BV Ex-BE10C Ex-BE10CV Display LCD-Display 63 x 63 mm, 128 x 128 Pixel Protection class IP55 Power supply Lithium battery 3.6 V (max. 200 hours count mode) or via relay module, power supply unit...
5.2.1 Electrical data Internal power supply 3.6 V (DC); Approved battery type for the power supply: • Tadiran SL-760, 3.6V, size AA Change only outside of hazardous location. Operating unit input Explosion protection class intrinsic safety Ex ib IIB/IIC; (housing connector or Max.
5.3 Safety measures when handling the battery NOTICE Lithium batteries represent a safe and reliable energy source when used properly. However, these potential consequences may occur if they are used incorrectly or misused: • The leakage of battery fluid or • the escape of gas or, in extreme circumstances, • a fire and/or •...
5.5 Prevention of obvious misapplication (examples) • Do not connect the operating unit with other electronic devices if their usability is not confirmed by the manufacturer. • Do not use operating units BE10 and BE10V neither in explosive hazard area nor with electric connection in explosive hazard area. 6. Scope of supply When unpacking the goods, check the articles that: •...
7. Marking Type of operating unit ATEX-Identifi cation Technical data Serial number Year of construction (last two digits of the serial number e.g. -20 for 2020)
Blast wave: Flying parts can kill you. ► Change of battery only outside of hazardous location. ► Use only the delivered battery. NOTICE ► After you have inserted the battery, the Lutz company logo is displayed. The operating unit now waits until a basic flow meter is connected. The operating display appears approx. 30 seconds after the basic flow meter is connected. If it takes more than 30 seconds to connect the basic flow meter, the display switches itself off. After you have touched the user interface, the display switches itself on again.
Page 17
4. Loosen the screw and remove the cover of the battery bracket. 5. Insert the new battery. Check the correct polarity. The Lutz company logo is displayed. 6. Mount the cover on the bat- tery bracket. 7. Connect the cover of the bat- tery bracket with the screw.
8.2 Turning the operating unit / pulse converter / protection cap NOTICE ►Use the supplied Allen key SW3 for the screws. ►Only tighten the screws hand-tight. The operating unit, pulse converter and protection cap can be turned to the required position in 90°-steps. Mounting steps for operating unit (The assembly of the pulse converter and the protection cap must be made as follows): Flow meter system TS (→...
8.3 Separate installation of the operating unit Due to the modular construction of the flow meter system, it is possible to install the operating unit with or without relay module separately to the basic flow meter itself. A protection cap or a pulse converter, an intermediate plate and a data line are required in addition. NOTICE ►Use the supplied Allen key SW3 for the screws. ►Only tighten the screws hand-tight. ►Use a pulse converter instead of a protection cap if the data line is longer than 5 m.
Page 20
10. Connect the plug with the circuit board of the pulse converter (or with the ribbon cable of the protection cap). 11. A ttach the pulse converter / protection cap onto the flow meter and mount the caps. 12. Connect the operating unit and pulse converter / protection cap using the data line. Fig.
8.4 Hose installation of the operating unit NOTICE ► Use the supplied Allen key SW3 for the screws. ► Only tighten the screws hand- tight. Mounting steps: 1. Remove the caps and release the screws. 2. Carefully lift the operating unit.
9. Operation Operating display Menu AUTO STOP Menu level Menu 4/5 LCD contrast Safety prompt Flow monitoring Communication Simulation Function level Fault detection Monitoring time Change value Cancel...
Page 23
Explanation of the display Top title Button to call up the menu level Symbol - Liquid is registered Symbol - External actuation is active 5 Symbol - Valve function On or Off point is active Button for help function Symbol - Overrun correction is active Symbol - Programming lock is active Symbol - Relays 1 and 2 are active 10 Unit of measurement 11 Button...
9.1 Manual mode In manual mode, you carry out filling with a manual nozzle. Prior to the filling process you set the partial volume meter to 0 and the desired volume is measured whilst you operate the nozzle. Setting the operating mode (→ chapter 9.5) Menu Operating mode: Manual mode RESET 9.1.1 Resetting the part volume meter Filling Menu Press the "RESET"...
9.2 Operation with volume preset In this mode you can enter the required filling volume on the flow meter. This volume is automatically added to a container once you start the process. The flow meter controls a pump and/or a solenoid valve for the fill process via the relay module. Setting the operating mode (→ chapter 9.5) Menu Symbol for operating mode volume preset AUTO Press the "START" button to start the filling process. Press the "STOP" button to stop the filling process. Press "Change volume" to change the preset volume. Change START volume 9.2.1 Starting the filling process Menu AUTO...
Menu AUTO The filling process is finished when the preselected volume has been filled. Change START volume 9.2.2 Stopping the filling process Menu AUTO STOP Interrupt Press "Continue" to continue the started filling process. Continue Press "Change filling volume" to change the preset volume. The new filling Change filling volume volume is reduced by the volume already added. Press "END" to end the started filling process.
9.2.3 Changing the volume preset Menu AUTO Change START volume Filling volume ← ← Press "← ← C" to delete the last entry on the display. Cncl Use "0 to 9” and "." to enter the fill volume. Filling volume Press "Cncl" to cancel the entry and return to the previous screen. ←...
9.3 Call up the menu level Menu RESET You can use this function to view all present settings (→ chapter 9.4). You can use this function to determine an op. mode, manual mode, volume preset or display mode (→ chapter 9.5). Menu 1/5 Current setting You can use this function to determine the displayed meas.
Page 29
You can use this function to switch fault detection on or off. You can specify a monitoring time for the flow meter. If the flow meter does not detect any fluid during this monitoring time, the current filling process is terminated and an error is reported. This allows issues such as an empty sampling vessel, dry running, air pockets, and mechanical faults on the flow meter to be detected (→ chapter 9.14). You can use this function to set the switching behavior of relay output 2 during the filling process. You can switch on a connected solenoid valve early or with Menu 3/5 a delay in this way (→ chapter 9.15). Fault detection You can use this function to switch the leakage alarm on or off.
Page 30
You can use this function to switch the measuring activity of the flow meter on or off. This is helpful for commissioning and maintenance work (→ chapter 9.24). In the flow meter system, multiple operating units, relay modules, and basic flow meters can be connected up. A relay module and a basic flow meter form a unit. The operating unit only works with a basic flow meter or with a unit Menu 5/5 consisting of relay module and basic flow meter in all cases. You can use this function to couple up a basic flow meter and a relay module to form a unit. In Meas. on/off the case of multiple units you select which one is actuated by the operating Coupling modules unit. This function is only active if multiple basic flow meters or relay modules Programming lock are connected up (→ chapter 9.25). Factory settings SW version You can use this function to protect important flow meter settings from being changed by unauthorized persons.
9.4 Current settings You can use this function to view all present settings. Set operating mode. Abbreviations: MAN (Manual mode), AUTO (Volume preset), Disp. (Display mode) Set display: The front letter stands for the large display; the rear letter for the smaller display.
Page 32
Safety prompt: On/Off Digital input: On/Off Current setting 4/5 Safety: Alarm output: On/Off Dig. inp.: Alarm: RS 485-Interface: On/Off RS 485: Analog: Analog output On/Off Pulse outp.: Measurement: Off Pulse output On/Off Measurement – Fluid flowing through is measured: On/Off Basic flow meter type: ST10, SL10, LM10, UN10, VA10, HDO250P, HDO250, HDO400, HDO500, Other Current setting 5/5 Basic meter:...
9.5 Setting the operating mode The modular flow meter system can be operated in different operating modes: Menu Menu 1/5 Current setting Operating mode Display options Calibrate RESET Total volume Manual mode: In manual mode, you carry out filling with a manual Op. mode nozzle. Prior to the filling process you set the partial Manual mode volume meter to 0 and the desired volume is measured whilst you operate the nozzle. Volume preset Display mode Volume preset:...
9.6 Display options The measured values can be shown on a large and small display on the operating display. You can define which measured values are displayed under "Display options." The top row in the selection field refers to the large display and the bottom row to the small display. The selection options are different in manual mode and automatic mode. The last definition before a change was made to the display mode remains valid for this operating mode. Explanation of the displays: Manual mode Display opt. 1/2 Partial volume counter Partial volume/ The partial vol. meter counts the volume of a partial volume (lot, filling proc., Total volume...
Menu Cal. process Cal. process On the next screen, enter Filling process the value read complete? from the vessel. Continue Value on vessel Value on vessel In this screen, enter the value read from the vessel. Press "← ← C" to delete the last entry on the display. Use "0 to 9”...
Page 37
Menu Cal. process AUTO volume Press the "START" button to start the filling has been added. process into a measurement vessel. Cancel START Menu Cal. process Cal. process AUTO On the next screen, enter Filling process the value read complete? from the vessel. Continue STOP Value on vessel In this screen, enter the value read from the vessel.
9.7.3 Manual input of the calibration factor Menu Menu 1/5 Calibrate Current setting Comparative measurement Operating mode Manual input Display options Calibrate Cancel RESET Total volume In this screen, enter the calibration factor. Cal. factor Cal. factor Use "↓" and "↑" to select one of the last 10 set cal. factors. The last value is 1.2345 Station1 the factory cal.
9.8 Deleting the total volume Menu Menu 1/5 Total volume Current setting Total volume Operating mode 25634.2 L delete? Display options Calibrate Cancel RESET Total volume Syst. info Really delete? This display only appears when the safety prompt is activated (→ chapter 9.20). Cancel Syst.
9.10 Select unit of measure Menu Menu 2/5 Language Unit of measure Decimal places Economy RESET Backlight Unit of measure Press "Volume unit" to select the unit of measure for displaying vol.- Volume unit dependent values. Time unit Press "Time unit" to select the unit of measure for displaying time-depend- ent values.
You can use this button to select from the following units of measure for Volume unit volume-dependent values: Liter, m³ (cubic meter), kg (kilogram), t (ton), Liters US-Gal US-Gal (US gallon), IMP-G (imperial gallon), ft³ (cubic foot), or lb (pound). IMP-G The factory setting is "Liter".
9.11 Change displayed decimal places Menu Menu 2/5 Language Unit of measure Decimal places Economy RESET Backlight Press "000000" to display the values without decimal places. Decimal places Press "11111.1" to the values with one decimal place. This is the factory setting.
9.12 Change energy save mode Menu Menu 2/5 Language Unit of measure Decimal places Economy RESET Backlight Switch off display completely Economy The function with the lowest current consumption. The flow meter com- Switch off display compl. pletely switches the display off if it is not used for more than 1 minute. The Switch off display comes back on when fluid is detected again or if you touch the user display part. interface. Switch off display never Switch off display partially Cancel The function with the average current consumption. The flow meter switches off the top title/menu bar and the buttons if it is not used for longer than 1...
9.14 Fault detection On/Off/Change value The filling process can be controlled automatically by the flow meter, since faults may develop in the con- necting lines, such as running dry, blockages, defective solenoid valves, etc. After the start of the filling process, the flow meter monitors the time during which no liquid is registered. The flow meter switches off when a certain time limit is exceeded. A fault report appears on the display. The interrupted filling process can be resumed or completed when the fault has been remedied. The monitoring time can be freely selected between 1 and 99999 seconds. A time of 3 seconds is set by the manufacturer.
9.15 Valve function Menu Menu 3/5 AUTO Fault detection Valve function Leakage alarm Overrun corr. Change START Ext. actuation volume Valve function Description of the function "On point" → chapter 9.15.1 On point Description of the function "Off point" → chapter 9.15.2 Off point Cancel 9.15.1 On point The function "On point"...
Page 46
Valve function On point Off point Cancel Press "Change value" to set the On point for the solenoid valve. After the On point start of the filling process, the solenoid valve is opened with the set delay. Solenoid valve 5.23 s Press "On" to switch time-delayed opening of the solenoid valve on. This switch on after pressing "START".
9.15.2 Off point The function "Off point" can be used to switch off the solenoid valve on relay output 2 early in the fill- ing process. This can offset e.g. long valve closing times. With a second valve on relay output 1, pressure peaks and foam formation can be avoided. A throttle element, such as a ball valve in line with the valve disconnected last, to regulate the flow volume is beneficial. Fig. 9 - Filling operation with two solenoid valves Rel. 1 Rel.
Page 48
Valve function On point Off point Cancel Press "Change value" to set the Off point for the solenoid valve. The sole- Off point noid valve is closed as per this value before the end of the filling process. Sol. valve 5.23 s Press "On" to switch the early closing of the solenoid valve on. This can switch off before filling end. offset e.g. long valve closing times. With a second valve on relay output 1, Change value pressure peaks and foam formation can be avoided. Press "Off" to switch this function off.
9.16 Leakage alarm On/Off Menu Menu 3/5 AUTO Fault detection Valve function Leakage alarm Overrun corr. Change START Ext. actuation volume Press "On" to switch the leakage alarm on. Leakage alarm Press "Off" to switch the leakage alarm off. This is the factory setting. Press "OK" to save the selection and return to the menu level. Press "Cancel"...
9.18 External actuation On/Off Menu Menu 3/5 AUTO Fault detection Valve function Leakage alarm Overrun corr. Change START Ext. actuation volume Press "On" to switch external actuation on. Ext. actuation Press "Off" to switch external actuation off. This is the factory setting. Press "OK" to save the selection and return to the menu level. Press "Cancel"...
9.20 Safety prompt On/Off Menu Menu 4/5 LCD contrast Safety prompt Flow monitoring Communication RESET Simulation Press "On" to switch the safety prompt before the start of volume preselec- Safety prompt tion on. This is the factory setting. Press "Off" to switch the safety prompt before the start of volume preselec- tion off.
Lower limit Press "← ← C" to delete the last entry on the display. Use the buttons "0-9" and "." to enter the value. Press "Cncl" to cancel the entry and return to the previous screen. ← ← Cncl Press "OK" to save the value and finish making entries. Upper limit Press "←...
Communication 1/2 You can use this function to determine the behavior of the digital input. Digital input Alarm output You can use this function to determine the behavior of the alarm output. RS 485 You can use this function to determine the behavior of interface RS 485. Analog output Cancel You can use this function to determine the behavior of the analog output.
9.22.2 Alarm output On/Off Communication 1/2 Digital input Alarm output RS 485 Analog output Cancel Press "On" to switch the alarm output on. Alarm output Press "Off" to switch the alarm output off. This is the factory setting. Press "OK" to save the selection and return to the menu level. Press "Cancel" to retain the setting and return to the previous menu level. Cancel 9.22.3 RS 485-Interface On/Off Communication 1/2...
9.22.4 Analog output Communication 1/2 Digital input Alarm output RS 485 Analog output Cancel Analog output Press "Off" to switch the analog output off. 0-20 mA NAMUR This is the factory setting. 4-20 mA NAMUR 0-20mA 4-20mA Press "0-20 mA NAMUR", "4-20 mA NAMUR", "0-20mA", 0-10 V DC "4-20mA"...
9.22.5 Pulse output Communication 2/2 Pulse output Cancel Press "Change value" to define the pulse value for the pulse output. Pulse output The pulse value is the volume that flows through the flow meter before a Pulse value: pulse is triggered. 10.5 L/Pulse Press "On" to switch the pulse output on. Change value Press "Off" to switch the pulse output off. This is the factory setting. Cancel Press "← ← C" to delete the last entry on the display. Pulse value Use the buttons "0-9"...
9.23 Simulation of the outputs You can use this function to recreate certain conditions at the flow meter system outputs. This is helpful for commissioning and maintenance work. Menu Menu 4/5 LCD contrast Safety prompt Flow monitoring Communication RESET Simulation Simulation 1/2 You can use this function to simulate the behavior of relay output 1. Relay output 1 Relay output 2 You can use this function to simulate the behavior of relay output 2.
9.23.4 Simulation of RS 485 interface Simulation 1/2 Relay output 1 Relay output 2 Alarm output RS 485 Cancel Simulation RS485 Press "On" to switch the simulation on interface RS 485 on. Press "Off" to switch the simulation on interface RS 485 off. Press "Cancel" to end the simulation. Cancel 9.23.5 Simulation of analog output Simulation 2/2...
9.23.6 Simulation of pulse output Simulation 2/2 Analog output Pulse output Cancel Press "Off" to switch the simulation on the pulse output off. Press "0 Hz" to switch the pulse output to the rest state. Simulation Pulse Simulation Pulse Use the buttons "2 Hz" to "1 kHz" to switch on the respective signal 100 Hz frequency on the pulse output.
Meas. on/off Press "OK" to save the selection and return to the menu level. Press "Cancel" to retain the setting and return to the previous menu level. Cancel Menu This display appears instead of the measured values when the measurement is switched off.
9.25.1 Select relay module Coupling modules Coupling modules Name Relay module Ser.No. 12345 000001 Basic meter Ser.No. 12345 000001 Cancel Press "None" to disconnect the coupling of a relay module and the basic Relay module 1/2 flow meter from the previous screen. None You can use this button to connect the relay module with the displayed serial number to the basic flow meter from the previous screen.
9.26 Programming lock On/Off/Change value The programming lock is switched on and off with a numerical code with Menu a maximum of 4 digits. The following functions cannot be changed through the programming lock: Select operating mode, Display options, Calibration, Delete total volume, Language, Unit of measurement, Decimal places, Economy, Fault detec- RESET tion, Valve function, Leakage alarm, Overrun correction, External actua- tion, LCD contrast, Safety prompt, Flow monitoring, Communication,...
9.27 Resetting to factory settings Menu 5/5 Menu Meas. on/off Coupling modules Programming lock Factory settings RESET SW version The following values and functions are reset: Factory settings Display options, Calibration factor (1), Unit of measurement (L/min), Decimal Reset device to factory places (1), Economy (Switch off part.), Backlight (Off), Fault detection (On), settings?
9.29 Help function Menu You can obtain more information on the functions of the operating unit on the display. To do so, first press the button with the question mark. The frame around the button changes to confirm. RESET Menu Then press the display area for which you would like more information. RESET Help 5 Press the "RESET" button to clear the part.
10. Maintenance and servicing 10.1 Reset function If the operating unit no longer responds in count mode or when buttons are pressed, you can restart the system. To do so, press both "Reset" buttons on the operating panel for 5 seconds at the same time.
Page 67
Mounting steps: 1. Remove the caps and release the screws. 2. Carefully lift the operating unit. Pay attention to cable length! 3. Release the plug. 4. Loosen the screw and remove the cover of the bat- tery bracket. Fig. 11b - Change battery 5.
Taking back of batteries and rechargeable batteries is free of charge. 11. Repairs Repairs should only be made by the manufacturer or authorized Lutz-dealers. Only use original Lutz spare parts. 12. Traceability Products manufactured by Lutz Pumpen for potentially explosive atmospheres are identified by an indi- vidual batch number which allows them to be traced.
Annex 1. Terminal assignment of cables and connectors Function white brown + 6V green Earthing yellow grey Reed switch pink blue Coding black violet grey / pink red / blue SLEEP/ white / green brown / green RESET/ Earthing Fig. A1 - Data outlet operating unit Fig.
2. Trouble shooting 2.1 Fault report Operating mode Fault report Cause Remedy General Communication with • The connection • Check bus line another module not between the modules possible. is interrupted CHECK BUS! • Program error in the • Restart the system by pressing system "Reset"...
Page 71
Operating mode Fault report Cause Remedy Flow monitoring Flow rate too low • Pump damaged • Check pump • Line strainer blocked • Clean strainer • Hose line kinked • Check hose line • Solenoid valve • Check solenoid valve damaged Flow monitoring Flow rate too high...
2.2 Other faults Operating mode Fault Cause Remedy General No liquid registered • Air in the piping • Prevent air pockets in system/dry running connecting lines, use a solid hose system • Pump is not • Check pump operating • Line strainer blocked •...
Page 73
Operating mode Fault Cause Remedy General Liquid leaks from the • Seal defective • Replace seal housing of basic flow meter • Hairline cracks in • Check installation conditions housing and/or on for mechanical stresses and screw connection correct them due to mechanical influences •...
Page 74
• Battery contacts • Clean contacts corroded After insertion of • Basic flow meter is • Check basic flow meter the battery the missing Lutz company logo • Check connecting wires and appears longer than 1 plug connections minute No indication after • Batteries fitted • Open battery compartment...
Translation of the original declaration of conformity We herewith declare under the sole responsibility that the following product complies with the EU Directives listed. Manufacturer: Lutz Pumpen GmbH Erlenstraße 5-7 D-97877 Wertheim Operating unit Product: Type: Relevant European ATEX : 2014/34/EU ●...