Page 1
PRV. The incorrectly set PRV failed to protect the COBRA set from excessive hot water pressure and this caused the inner liner of the hot water housing to deform and lose contact with the sealing o-ring. As a result hot water leaked into the scrubber housing.
Page 2
Fit the test plug (DB5003843), test nut (DB5003844) and o-ring (E13330) to the hot water relief valve DB5002335 as shown in figure 1. These items are contained in the COBRA Tool and Test Kit. Fit a 3/8” UNF plug to the outlet of the relief valve.
Page 3
JFD are currently undertaking work to improve the flow capacity of the relief valve. This will ensure that the COBRA set is protected from overpressure, even if the COBRA set hot water outlet is totally blocked for any reason. A further guidance note and replacement parts will be issued in the near future to allow clients to carry out this modification.