BELIMED WD 200 Loading Manual

BELIMED WD 200 Loading Manual

Rack for containers

Advertisement

Rack for containers WD 200
Loading Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the WD 200 and is the answer not in the manual?

Questions and answers

Summary of Contents for BELIMED WD 200

  • Page 1 Rack for containers WD 200 Loading Manual...
  • Page 2 These instructions apply to: Product Rack Rack for containers WD 200 Material 643019 Document | Version 1007-4247 | 02 Language EN-US English Month | Year 12 | 2020 Copyright 2020 © Belimed AG...
  • Page 3: Table Of Contents

    Rack for containers WD 200 | © Belimed Content Introduction Purpose of these instructions ....... 4 Target group .
  • Page 4: Introduction Purpose Of These Instructions

    This manual, including all parts thereof, are protected by copyright law. Utilization of any kind, wheth- er for duplication, translation, or processing in other media, is not permissible without the consent of Belimed AG. The content of this manual must not be used for commercial purposes. Symbols and notices used The following symbols and notices are used to indicate precautionary measures for handling and for occupational safety that are important to avoid harm to health and life.
  • Page 5 Rack for containers WD 200 | © Belimed Introduction CAUTION Indicates hazard with low risk If the hazard is not avoided, it will lead to minor or moderate injury. Possible consequences of the hazard.  Work steps to avoid the hazard.
  • Page 6: Safety Instructions

    Rack for containers WD 200 | © Belimed Safety instructions Declaration of conformity With the EC Declaration of Conformity and the CE mark, Belimed AG affirms that this product com- plies with the essential health and safety requirements. General safety instructions The machine and accessories may only be operated in faultless condition, which is why the following safety instructions must be complied with to avoid risks to patients and operators.
  • Page 7: Warranty

    Defective components must only be replaced by original spare parts from Belimed. Only if original spare parts are used can it be ensured that the spare parts meet safety requirements. If spare parts are used that neither originate from Belimed AG, nor have not been approved by Belimed AG, the warranty is rendered void.
  • Page 8: Summary Of The Safety Instructions

    Safety instructions Rack for containers WD 200 | © Belimed Summary of the safety instructions Summary of all safety instructions for safe operation of the product. Safety instructions are listed in all places where hazards may arise.  Always follow the safety instructions! ...
  • Page 9 Rack for containers WD 200 | © Belimed Safety instructions WARNING Risk of infection due to irregular maintenance For safe and reliable operation, regular maintenance by Belimed-Service is required.  Have regular maintenance work carried out by Belimed-Service. 9 | 18...
  • Page 10: Description

    Rack for containers WD 200 | © Belimed Description Fig. 3-1 Rack 1 Handle 2 Rotating nozzle 3 Roller 4 Coding magnet (option) Option Coding magnets are optionally included with the rack.  Manually install coding magnets depending on the load.
  • Page 11: Loading

     Follow local work instructions. WARNING Risk of infection due to irregular maintenance For safe and reliable operation, regular maintenance by Belimed-Service is required.  Have regular maintenance work carried out by Belimed-Service. CAUTION Risk of infection due to inadequate pre-cleaning Inadequate pre-cleaning can lead to deficient reprocessing.
  • Page 12 Loading Rack for containers WD 200 | © Belimed Fig. 4-1 Example of loading of sterilization container 1 Fixing tube (option) 2 Trough 3 Cover Loading sterilization containers  Load 2 troughs 2.  Place the troughs 2 on their short sides.
  • Page 13: Maintenance And Cleaning Work

    Rack for containers WD 200 | © Belimed Maintenance and cleaning work WARNING Risk of infection due to residues Residual contamination can lead to infections.  Wear the required personal protective equipment, e.g., protective goggles, protective gloves, etc., while performing inspection and cleaning work.
  • Page 14 Maintenance and cleaning work Rack for containers WD 200 | © Belimed Cleaning the rotating nozzle  Unscrew the knurled screw 1.  Remove the washer 2.  Pull out the rotating nozzle 3.  Clean the rotating nozzle 3.
  • Page 15: Optimize Cleaning Performance

    Rack for containers WD 200 | © Belimed Maintenance and cleaning work Optimize cleaning performance WARNING Risk of infection due to soiled wash items Contaminated wash items can lead to life-threatening infections.  Wear the required personal protective equipment, e.g., protective goggles, protective gloves, etc.
  • Page 16: Disposal

    Rack for containers WD 200 | © Belimed Disposal DANGER Risk of infection due to contaminated racks The rack can be contaminated by the last work processes.  Wear personal protective equipment, e.g., protective gloves.  Disinfect the entire rack before disposal.
  • Page 17 Rack for containers WD 200 | © Belimed Disposal 17 | 18...
  • Page 18 Belimed AG | Grienbachstrasse 11 | 6300 Zug | Switzerland | +41 41 449 78 88 | +41 41 449 78 89 | www.belimed.com...