TempHion pH/ORP (Redox) / ISE Smart Sensor and Data Logger Instructions TempHion pH/ ORP (Redox) TempHion ISE For Sales & Service Contact 2650 E. 40th Ave. • Denver, CO 80205 Phone 303-320-4764 • Fax 303-322-7242 1-800-833-7958 PROUDLY www.geotechenv.com MADE IN THE...
TABLE OF CONTENTS TEMPHION INSTRUCTIONS General Information General Information ..............................Page 4 Dimensions ..................................Page 4 Specifications ................................Page 5 Initial Inspection and Handling ..........................Page 6 Do’s and Don’ts ................................Page 6 TempHion Reference Electrode ..........................Page 7 General Precautions ..............................Page 7 Installation Connecting External Power ............................Page 8 Connecting a TempHion to a Computer ......................Page 8 Cable Wiring ..................................Page 9 Installing Aqua4Plus or Aqua4Plus Lite Software ...................Page 9...
TEMPHION INSTRUCTIONS GENERAL INFORMATION Seametrics TempHion™ Smart Sensor Several TempHions, or a combination of TempHions microprocessor-based submersible sensor with built-in and other Seametrics Smart Sensors, can be networked data logging, storing thousands of records in non-volatile together and controlled from one location directly from a memory.
TEMPHION INSTRUCTIONS GENERAL INFORMATION Specifications* Housing & Cable Weight 0.8 lb. (0.4 kg) Body Material Acetal & 316 stainless or titanium Wire Seal Fluoropolymer and PTFE Material Cable Submersible: polyurethane, polyethylene, or ETFE (4 lb./100 ft., 1.8 kg/30 m) Field Connector Standard Temperature Operating Range...
TEMPHION INSTRUCTIONS GENERAL INFORMATION Initial Inspection and Handling Upon receipt of your smart sensor, inspect the shipping package for damage. If any damage is apparent, note the signs of damage on the appropriate shipping form. After opening the carton, look for concealed damage, such as a cut cable. If concealed damage is found, immediately file a claim with the carrier.
TEMPHION INSTRUCTIONS GENERAL INFORMATION TempHion™ Reference Electrode General Precautions TempHion’s patented reference electrode is the key to The rest of this manual includes step-by-step instructions TempHion’s superior downhole/high pressure performance. for setting up the TempHion calibrating it, and using it in TempHion uses a long capillary pathway, initially filled the field.
TEMPHION INSTRUCTIONS INSTALLATION Connecting External Power Interface Cable The TempHion comes with two 1.5V AA internal batteries. Screw- RS485/RS232 If auxiliary power is desired, you can use a 6–16 VDC supply Adapter Sensor that can provide 15 mA. Connect to Vaux++ (pin 1 - white) Serial Port and Ground (pin 5 - blue) or contact Seametrics for auxiliary PC or Laptop...
TEMPHION INSTRUCTIONS INSTALLATION Cable Wiring Installing Aqua4Plus or Aqua4Plus Lite Software If you buy your cabled sensor with a connector installed The TempHion comes with the Aqua4Plus or Aqua4Plus (the normal configuration), no further wiring is needed. Lite host software that is installed on your PC or laptop. Use For reference purposes, the first two diagrams below show this software to program the datalogger, to retrieve data the pinout from the connector for various scenarios.
TEMPHION INSTRUCTIONS INSTALLATION Installing the Sensor The black reference reservoir at the lower end of the Note about Calibration TempHion Smart Sensor is shipped filled with Seametrics reference solution. (If your reference reservoir is not filled, The TempHion must be calibrated before deploying. see the Maintenance section in this manual.) The black The TempHion has one temperature channel and four reference assembly has two grooves.
SETTINGS AND CALIBRATION TEMPHION INSTRUCTIONS General Calibration Information IMPORTANT CALIBRATION NOTES! The TempHion should be calibrated before first use and periodically thereafter. It should also be calibrated if moving Data Files • Calibration can only be done when to a different sampling environment where readings will be there is no data stored on the sensor.
SETTINGS AND CALIBRATION TEMPHION INSTRUCTIONS pH Channel Calibration Two-Point Calibration Preparation • Seametrics recommends pH buffers of 4, 7, and 10 for — Compute First Calibration Point — calibration. For a one-point calibration, select the buffer closest to the expected values in your samples. For a two- •...
TEMPHION INSTRUCTIONS SETTINGS AND CALIBRATION ISE Channel Calibration Introduction to ISE Calibration One-Point Calibration Seametrics recommends using the “known addition — Computing Calibration Value— method” for preparing calibration solutions. Using this • Rinse sensor with distilled water and pat dry. method, the sensor is placed in 100 mL of distilled or de- •...
Page 14
TEMPHION INSTRUCTIONS SETTINGS AND CALIBRATION ISE Channel Calibration (continued) Two-Point Calibration — First Calibration Point— — Applying Calibration Values— • Rinse sensor with distilled water and pat dry. • Observe the M and I values in the right hand section of the calculator.
TEMPHION INSTRUCTIONS SETTINGS AND CALIBRATION ORP (Redox) Channel Calibration Note on units: The unit “Eh” refers to readings in millivolts — Applying Calibration — referenced to a hydrogen electrode. In • Re-open the Settings and Calibration window. other words, Eh represents millivolt •...
TEMPHION INSTRUCTIONS OPERATION Collecting Data with Aqua4Plus & Aqua4Plus Lite Retrieving Data from the Sensor/Datalogger • Click on the file you want to retrieve. Following is a brief overview on using Aqua4Plus/ Aqua4Plus Lite to collect data. Please refer to the software •...
OPERATION/DIRECT READ MODBUS & SDI - 12 TEMPHION INSTRUCTIONS DIRECT READ ( MODBUS OR SDI - 12 ) ® While the TempHion comes with Seametrics’ easy to use Aqua4Plus or Aqua4Plus Lite software, you can also use standard Modbus RTU or SDI-12 equipment to easily ®...
DIRECT READ MODBUS & SDI - 12 TEMPHION INSTRUCTIONS Power Consideration Register Addresses with Firmware prior to 2.5 Zero based One based If your sensor does not have internal batteries and is not Temperature 62592 62593 powered continuously by an auxiliary power supply, then pH (in mV) 62594 62595...
DIRECT READ MODBUS & SDI - 12 TEMPHION INSTRUCTIONS Reading via SDI-12 SDI-12 commands always return values for all four TempHion channels – temperature, pH, ORP, and ISE channels. Most TempHions are configured either as temperature, pH, and ISE or just temperature and ISE (such as Chloride or Bromide). You will need to disregard the readings for the channel(s) you do not have configured on your unit.
Page 20
DIRECT READ MODBUS & SDI - 12 TEMPHION INSTRUCTIONS Request Measurement with CRC <a>MC! <a>0024<cr><lf> Request all measurements <a>D0! <a>+21.345+7.181+.053+.459 {crc}<cr><lf> Read: temperature, pH (in pH units), (ISE in ppm), ORP (in Eh) <a>MC1! <a>0024<cr><lf> Request all measurements <a>D0! <a>+21.345-134.458+100.48+.84.404 {crc} <cr><lf>...
TEMPHION INSTRUCTIONS MAINTENANCE and REFERENCE Emptying and Cleaning the Reservoir Sensor Your TempHion is normally shipped from the factory There are no user-serviceable parts, other than the batteries already filled with reference solution. However, from time and reference solution. If problems develop with sensor to time you will need to replace the reference solution.
TEMPHION INSTRUCTIONS MAINTENANCE and REFERENCE and the sensor may read erratically! M & I Range for pH Calibration There are three variations of TempHion pH configuration. Depending on which way yours is configured, your M and I Excess filling solution values should be between those values shown in the chart spills over, forcing out below when you calibrate the sensor.
TEMPHION INSTRUCTIONS MAINTENANCE and REFERENCE Reconditioning Battery Changing Tips • Never place a tool on the sensor body, it is very Reference Electrode thin and will deform causing leaks at o-ring seals • If stored dry or in KCl, refill per maintenance section and potentially crushing the circuit board! earlier in this manual.
Page 24
TEMPHION INSTRUCTIONS MAINTENANCE and REFERENCE By Hand—cabled version only With Pliers or Vise Grips—good for field use 1. Tightly grasp the top cap in one hand. Cabled Sensor 2. Brace your hand against something such as a 1. Grasp the pliers on the swaged area (do not grab table or the ground.
Page 25
TEMPHION INSTRUCTIONS MAINTENANCE and REFERENCE Removing the Housing 4. Hold the top cap assembly at 90° to the housing opening as shown. Depress the spring with your 1. With your free hand, grasp the sensor body. Press fingertip and tuck the wiring into the cutaway on down to compress the square ring.
Page 26
TROUBLESHOOTING TEMPHION INSTRUCTIONS Problem Probable Causes Things to try… Software will not Loose cable Make sure all cable connections are secure. communicate with sensor Contacts in connector loose Be sure all wires are securely fastened inside the round connector. USB driver not installed See Connecting the TempHion to a Computer in the Installation section or see the USB/RS485 Adapter Installation application note on our...
Page 27
TEMPHION INSTRUCTIONS WARRANTY/DISCLAIMER LIMITED WARRANTY/DISCLAIMER - Seametrics TempHion SMART SENSOR A. Seller warrants that products manufactured by Seller when properly installed, shall be free from defects in material and workmanship. Seller’s obligation under this warranty shall be limited to replacing or repairing the part or parts or, at Seller’s option, the products which prove defective in material or workmanship within TWO (2) years from the date of delivery, provided that Buyer gives Seller prompt notice of any defect or failure and satisfactory proof thereof.
Page 28
Seametrics • 19026 72nd Avenue South • Kent, Washington 98032 • USA LT-14315r14-20180306 (P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • seametrics.com 3/6/2018...