Dual display to show the internal resistance and voltage of the battery simultaneously. – The RE856 has 99 sets of composite comparator function, which can be set at resistance and voltage values to get the reliable detection of battery deterioration.
RE856 OPERATION MANUAL Switches & Input Terminals LCD Display key: To turn the beeper on or off – To decrease the displayed value – Hold this key for 2 seconds to change the temperature units – key: To store one set of readings to memory –...
RE856 OPERATION MANUAL Screen Display Auto power off function is enabled Measured battery total capacity Resistance zero-adjustment function is on The calculated remaining capacity of the battery Display is locked Shown when the comparator function is on Auto-hold function is on...
Accessories Clip-type test lead with temperature sensor, pin-type test lead, DCA current adaptor, zero- adjustment board, operation manual, batteries, AC adaptor, USB cable, CD PC software (also available at www.valen.com.au), carrying case. AUS 1300 734 253 NZ 0800 734 253...
RE856 OPERATION MANUAL Electrical Specifications Conditions to Guarantee Accuracy Temperature 23°C ± 5°C Humidity 80% RH or less (no condensation) Temperature coefficient 0.10 x (specified accuracy)/°C (<18°C to >28°C) Zero-adjustment After zero adjustments for each range Resistance measurement Range Resolution...
RE856 OPERATION MANUAL OPERATION Preparation The following safety information must be observed to ensure maximum personal safety during the operation of this tester. To avoid electric shock when replacing the batteries, first disconnect the leads from the object to be measured.
RE856 OPERATION MANUAL Operation Read the battery internal resistane, DC voltage and temperature directly from the display. When using the clip type test lead, the temperature sensor is detected, and the temperature is automatically displayed. Press the key for 2 seconds to change the displayed temperature unit.
RE856 OPERATION MANUAL Pin-type Test Lead Zero-Adjustment When the resistance reading is lower than 200 digits and stable, the zero-adjustment is automatically carried out. The would then stop flashing, and the beeper sound would also stop. Remove the pin-type test leads from the zero-adjust board and start the measurement, connect the leads to the battery to test.
RE856 OPERATION MANUAL Moving Average Function If the resistance measurement value is unstable, this can be corrected with the moving average function. Press and hold down the key, then rotate the resistance range switch to any range to turn on the meter, to samples select mode, ‘SET’...
RE856 OPERATION MANUAL Disable Auto Power-off Function If no switch operation for 10 minutes, the meter automatically powers off. Auto data memory and continuous data logging mode will auto-disable the auto power-off function. For continuous measurement, in some cases it may be necessary to disable the auto power-off function.
RE856 OPERATION MANUAL Comparator Table Resistance Low Limit Resistance High Limit Resistance Voltage Middle High Voltage Comparison Beeper Beeper Beeper Value High Beeper Beeper Voltage comparison value = (high limit voltage value + low voltage value)/2. Start/Stop Controls for the Comparator...
RE856 OPERATION MANUAL Power on the battery charger, the meter will display the charge current value with a plus sign and the battery accumulated charging capacity ( Connect the battery load to the battery and use the DC current adaptor. Clamp the current transducer (jaw) around the ‘+’...
Page 17
RE856 OPERATION MANUAL To Auto Memorise the Reading Press the key for 4 seconds; the display will show SWT INTV. Using the keys, select the desired interval time from 3 seconds to 255 seconds. Press the key to exit. Press the key for 2 seconds to auto memory mode;...
Repairs or servicing are not covered in this operation manual and should only be performed by qualified personnel. Please return the meter to VALEN if this is the case. Periodically wipe the case with a dry cloth. Do not use abrasives or solvents on this instrument.
RE856 OPERATION MANUAL If a ‘User Account Control’ windows opens asking ‘Do you want to allow the following program from an unknown publisher to make changes to this computer?’ click ‘Yes’ Online Operation Run the Software Click ‘Run BatCapTesterS.exe icon on the dektop.
RE856 OPERATION MANUAL Click the button in the top left corner of the main screen. A dialogue box will open - input a file name then click to begin saving data to the file just created. Click the button to stop recording.
RE856 OPERATION MANUAL Data Convert Excel Open Microsoft Excel and locate the saved file in Excel format, for example, test.xls. The ‘Text Import Wizard’ then appears. Follow the steps from 1 to 3 to complete. Graph Open a saved data file in the software program and click the button forth from the left.
RE856 OPERATION MANUAL Sampling Time PC Sampling Time PC sampling rate is the rate at which the PC collects readings while connected to the meter. Click the button fifth from the left on the main menu. Meter Sampling Rate Meter sampling rate is the rate at which the meter stores readings. Click the button fifth from the left on the main menu.
RE856 OPERATION MANUAL Set Comparator Click the button fifth from the right on the main menu. Click to set the meter. Erase EEPROM To erase the EEPROM, click the button second from the right on the main menu. Click to confirm.