OPERATING INSTRUCTIONS FOR PRIME RANGE
1.
UNPACKING
1.1
Remove all packing material from between the shelves and inner walls of the oven.
2.
MAINS SUPPLY
2.1
240-volt units. Each unit comes supplied with a mains connection lead already fitted with
a correctly rated fuse. The fuse rating and other details for each unit is shown on the
voltage plate riveted to the back of the unit. It is important that, if the fuse needs to be
replaced, it must only be replaced with one of the correct rating.
2.2
110-volt units are supplied with a cable but without a plug or fuse. These units should be
wired in by a suitably qualified electrician to the following:-
BROWN
BLUE
GREEN/YELLOW 'E' Earth pin
DO NOT CONNECT THE OVEN TO A D.C. MAINS SUPPLY OR
3.
OPERATION – Standard Control
3.1
Position the shelves within the work chamber.
3.2
Switch 'ON' the mains switch, indicated by lighting up green.
3.3
If unit is already in use press the 'start/stop button
3.5
Press and hold the 'temperature button
3.5
The 'setpoint display' will continue to show the setpoint as it is edited. The 'temperature'
display will show "S" to indicate the setpoint is being edited.
3.6
Use the increment
degree steps.
3.7
When no keys have been pressed for 10 seconds, or if the `temperature button
pressed again, the new value will be stored and the temperature display will return to normal.
3.8
Set the 'overheat dial' to approximately 10 to 15 degrees above your set point.
to Page 6 for Overheat Scaling.
3.9
Press the 'Overheat reset' button to energise the overheat circuit
3.10 Press the 'start/stop button
18/03/2013 Genlab Ltd
'L' Live pin
'N' Neutral pin
WARNING
SERIOUS DAMAGE WILL OCCUR
' for 2 seconds.
and
decrement buttons to set the desired temperature in one
'
OP INS PRIME RANGE Issue 1
Tanhouse Lane, Riverview Industrial Estate
Widnes, Cheshire WA8 0SR England
Web site: wwww.genlab.co.uk
E-mail:
enquiries@genlab.co.uk
Refer to voltage plate
for fuse requirement
' to extinguish the amber lamp.
Tel: 0151 424 5001
Fax: 0151 495 2197
'
is
Refer
Page 1 of 6

Advertisement

loading
Need help?

Need help?

Do you have a question about the PRIME RANGE and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for GenLab PRIME RANGE

  • Page 1 Set the ‘overheat dial’ to approximately 10 to 15 degrees above your set point. Refer to Page 6 for Overheat Scaling. Press the ‘Overheat reset’ button to energise the overheat circuit 3.10 Press the ‘start/stop button ’ 18/03/2013 Genlab Ltd OP INS PRIME RANGE Issue 1 Page 1 of 6...
  • Page 2 1 degree of the setpoint. To indicate this feature, the timer display will flash the time remaining. 18/03/2013 Genlab Ltd OP INS PRIME RANGE Issue 1 Page 2 of 6...
  • Page 3  When the program finishes: o The oven will turn off. o The yellow ‘running LED will go out. o The green ‘complete’ LED will come on. 18/03/2013 Genlab Ltd OP INS PRIME RANGE Issue 1 Page 3 of 6...
  • Page 4: Preventative Maintenance

    SHOULD NOT be completely sealed. These units are not suitable for use where inflammable solvents are being used where the solvent concentration can reach inflammable or explosive levels. 18/03/2013 Genlab Ltd OP INS PRIME RANGE Issue 1 Page 4 of 6...
  • Page 5 This point will flash as it reached set point as a representation of the PID values tuning to the desired set point. 18/03/2013 Genlab Ltd OP INS PRIME RANGE Issue 1 Page 5 of 6...
  • Page 6 LED indicator (GREEN) Start/Stop: Keypad button. Increment: Keypad button. Decrement: Keypad button. Temperature: Keypad button. Timer: Keypad button. Program: Keypad button Test: Keypad button Reset: Keypad button 18/03/2013 Genlab Ltd OP INS PRIME RANGE Issue 1 Page 6 of 6...