DM-85T DIGTAL MULTIMETER
CONTENTS 
GENERAL DESCRIPTION ................................. 
SAFETY PRECAUTIONS.................................... 
FEATURES ................................................ 
OPERATION INSTRUCTION.............................. 
MAINTENANCE .............................. 
TROUBLE SHOOTING .............................. 
 
    General Description 
1.
This is a 3 1/2 digital multimeter with high stablility and performance. It uses a LCD with 
20mm high figure, which makes the reading clearer and the operation more convenient. 
It  can  test  DCV,  ACV,  DCA,  ACA,  resistance,  capacitance,  frequency,  NCV,    duty  cycle, 
temperature,  diode,  and  continuity.  This  meter  also  designed  with  functions  including 
unit  symbol  display,  data  hold,  lighting,  auto  range,  auto  power  off  and  warning 
functions.  To  assure  high  accuracy  and  resolution,  it  adopts  integrated  circuit  8‐bit 
microprocessor and a dual integral A/D conversion as LCD driver, giving high resolution 
and high accuracy. It is an ideal tool for labs, factories and radio‐technology. 
 
2.    Safety Instructions 
The instrument is designed in compliance with IEC1010 standard (safety standard issued 
by  International  Electro  technical  Committee).  Please  read  the  following  safety 
instructions before operation. 
 
2.1 Check the connection and insulation of test leads to avoid electric shock. 
2.2  To  avoid  electric  shock  and  damage  to  the  meter,  do  not  input  voltage  exceeding 
rated value. 
2.3 When measuring voltage higher than DC 60V or AC 40V, please be careful and avoid 
electric shock. 
2.4 Select correct function and range to avoid wrong operation. 
2.5 Move the test leads away from test points when switching to other function. 
2.6 Don't input voltage in current terminal. 
2.7 Don't make any modification to the circuit. It may damage the meter or jeopardize 
safety. 
2.8 Safety symbols: 
Operation Manual
 
  " "  High  voltage,  " "  GND,  " "  Dual  insulation,  " "  Refer  to  manual,  " "  Low 
battery indication. 
 
3.    Features   
3.1 General Characteristics 
3.1.1 Display: LCD; 
3.1.2 Max display: 6000 (3 5/6 digits, automatic polarity, and unit symbol display); 
3.1.3 Measurement method: Analog to digital converter (in micro processor ADC+MCU); 
3.1.4 Sampling rate: approx.3 times/sec. 
3.1.5 Over‐range display: "OL" displayed. 
3.1.6 Low battery indicator: " " 
3.1.7 Working environment: (0~40) , relative humidity: <80%
3.1.8 Store condition: (‐10~50) , relative humidity: <80%
3.1.9 Battery: 2 pieces 1.5V battery ("AAA" 7# battery); 
3.1.10 Dimension: 140×72×37mm (length x width x height); 
3.1.11 Weight: approx. 195g (including battery); 
3.1.12  Accessories:  test  leads,  TP01  thermocouple,  user  manual,  holster,  gift  box,  and 
2*1.5V batteries. 
 
3.2 Technical Features 
3.2.1 Accuracy: ± (a% × reading + digits). To assure accuracy, the ambient temperature 
should  be  (23±5)  ℃,  relative  humidity  <75%.  One  year  accuracy  is  guaranteed  since 
production date. 
 
3.2.2 DC Voltage(DCV) 
Range 
Accuracy 
600mV 
6V 
±(0.5%+4) 
60V 
 
Resolution 
0.1mV 
1mV 
10mV 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DM-85T and is the answer not in the manual?

Questions and answers

Summary of Contents for HTC DM-85T

  • Page 1   “ ”  High  voltage,  “ ”  GND,  “ ”  Dual  insulation,  “ ”  Refer  to  manual,  “ ”  Low  DM-85T DIGTAL MULTIMETER Operation Manual   battery indication.  CONTENTS    GENERAL DESCRIPTION ……………………………  3.    Features    SAFETY PRECAUTIONS………………………………  FEATURES …………………………………………  3.1 General Characteristics  OPERATION INSTRUCTION…………………………  3.1.1 Display: LCD;  MAINTENANCE …………………………  3.1.2 Max display: 6000 (3 5/6 digits, automatic polarity, and unit symbol display);  TROUBLE SHOOTING …………………………  3.1.3 Measurement method: Analog to digital converter (in micro processor ADC+MCU); ...
  • Page 2: Resistance(Ω

    3.2.5 AC Current (ACA)  600V  ±(1.0%+4)  100mV    Input impedance: at 200mv range >40MΩ, at other ranges is 10MΩ.  Range  Accuracy  Resolution  Overload protection: 600V DC/AC rms.    600uA  0.1μA  3.2.3. AC Voltage(ACV)   6000uA  1μA  Range  Accuracy  Resolution  60mA  ±(1.5%+5)  10μA  6V  1mV  ±(0.8%+6)  600mA  100μA  60V  10mV  6A  1mA  600V  ±(1.0%+6)  100mV  10A  ±(2.0%+10)  10mA  Input impedance: 10MΩ.  Maximum voltage drop: 200 mV for full range.  Overload protection: 600V DC/ AC rms. ...
  • Page 3: Operation

    When the resistance under test    Open  circuit  voltage:  is less than 50Ω, buzzer sounds  2.2V    continuously.  3.2.7 Capacitance (C)    Overload protection: 250V DC/AC peak value.  Range  Accuracy  Resolution  WARNING: Do not input voltage at this range!  60nF  ±(2.5%+6)  10pF    3.2.10 Temperature (℃/ )  ℉ 600nF  100pF  Range  Accuracy  Resolution  ±(2.5%+5)  6uF  1nF  <400  ±( ℃ 1.0%+5)  ‐40℃~1000℃  1℃  ℃ ≥400  ±(1.5%+15)  ...
  • Page 4 Low battery indication. Warning: To avoid error readings, which could lead to  ranges.  Under    (Diode/Continuity)  range,  press  SELECT  key  can  choose    possible electric shock or personal injury, please replace the battery in time. diode test) or    (continuity test). Under temperature ranges, press SELECT key to  Null;  Null;  choose Degrees Celsius or Degrees Fahrenheit. Under ACV ranges, press SELECT key  , , ℃ ℉ Duty cycle, Degrees Celsius, Degrees Fahrenheit;  to switch between Frequency or Duty Cycle measurement.  MΩ,kΩ,Ω Megohm, Kilohm, Ohm;  Hertz, Kilohertz, Megahertz;  Hz,kHz,MHz 2) When there is no measurement in 15 minutes, the meter will automatically power  Millivolts, Volts;  mV, V Microamp, Milliamp, Amperes (A).  uA, mA, A off and enter sleep mode. One minute before sleep mode, the buzzer will beep for 5   ...
  • Page 5 ⑥  Holster, Battery door.  1) Do not input a voltage over 600V. Or it may cause damage to the circuit of the  See picture.                            meter, and the built‐in buzzer will alarm.    2) Be careful while measuring a high voltage circuit. DO NOT touch the high voltage  4.2 DCV measurement  circuit.  4.2.1 Insert the black test lead into “COM” terminal, and the red one into “VΩmA”    terminal.  4.4 DCA measurement  4.2.2 Turn the rotary to switch to “ ” ranges. Under Auto Range status, it will display  4.4.1 Insert the black test lead into “COM” terminal and the red one into “VΩmA”  “AUTO” symbol. Press “RANGE/REL” key can change to Manual range, the available  terminal (Max.600mA) or into “10A” terminal (Max.10A);  ranges are 600mV, 6V, 60V, 600V.  4.4.2 Turn the rotary switch to Current ranges, auto range is the default when you turn  4.2.3 Connect test leads to the test point; LCD will display polarity and voltage of the test  on the meter, “AUTO” symbol displayed. Press “RANGE/REL” key can change to Manual  point connected by the red test lead.  range, the available ranges are 600uA, 6000uA, 60mA, 600mA, 6A, 10A.  NOTE:  4.4.3 Connect test leads to the tested circuit; LCD will display polarity and current of the ...
  • Page 6: Capacitance Measurement

      4.7 CAPACITANCE MEASUREMENT  NOTE:  4.7.1 Insert the black lest lead in “COM” terminal and the red one in “VΩmA” terminal;  1) If you are not sure about the range of current under test, please select the highest  4.7.2 Turn the rotary to switch to “ ” range;  range, and then select the proper range based on displaying value.  4.7.3 If the LCD doesn’t display “0”, press “RANGE/REL” to clear the reading;  2) If the LCD displays “OL”, it means the current is over range. Now you need to select  4.7.4 Connect the capacitor to “COM” and “VΩmA” terminal. (Note: the red test leads is  a higher range.  for positive pole +). LCD displays capacitance value.  3) Maximum input current is 600mA or 10A (which terminal the red test lead is  NOTE:  inserted into). Current exceeding rated value will damage the fuse, and may cause  1) Don’t input voltage or current under capacitance ranges;  damage to the circuit of meter.  2) Press “RANGE/REL” to clear the reading before testing to assure the accuracy;    3) There is only the auto range mode under the capacitance range;  4.6 RESISTANCE measurement  4) The  capacitor  must  be  completely  discharged  before  testing  to  avoid  damage  the  4.6.1 ...
  • Page 7: Temperature Measurement

    WARNING:    1) DO NOT input voltage at “Ω ” range.  This function could be affected by different external interference sources, and then the  2) When test circuits, make sure the power is off and all capacitors are discharged. Any  alarm is activated by wrong signal. The measurement result is for reference only.  negative potential or AC signal will activate the buzzer.      Turn the rotary function switch to “NCV” position. When the testing circuit is placed  4.11 TEMPERATURE MEASUREMENT  ” key to select  ℃  or  above the meter, the meter displays the strength of signal, and the buzzer alarms with  4.11.1 Turn the rotary switch to “℃/℉” range. Press “SELECT/ ℉  mode.  “beep beep”.  NOTE:  4.11.2  Insert  the  cold  end  (free  end)  of  thermocouple  in  “VΩmA(+)”  and  “COM(‐)”  1) Even if there is no voltage indication, there may be voltage on the circuit. Do not  terminal, and put the working end (temperature measuring end) of thermocouple on the  rely on NCV detector as the only way to detect voltage.  surface ...
  • Page 8 7) Wipe the case with a damp cloth and detergent. Do not use abrasives and alcohol to    ● The specifications are subject to changes without prior notice;  clean the meter.  ● The content of this manual is regarded as correct. If users find out any mistakes or  8) If  the  meter will  not  be  used  for  a  long time,  please  take  out  the  battery to avoid  leakage damage.  omissions, please kindly contact the manufacturer;  ● The  manufacturer  will  not  be  responsible  for  accidents  and  damage  caused  by  9) When  “ ” ...