CEMARLINE MAXX-700 User & Installation Manual

CEMARLINE MAXX-700 User & Installation Manual

Laser patient positioning system

Advertisement

Quick Links

The MAXX-700 User Installation Manual
Version 2.0 - January 2020

Advertisement

Table of Contents
loading

Summary of Contents for CEMARLINE MAXX-700

  • Page 1 The MAXX-700 User Installation Manual Version 2.0 - January 2020...
  • Page 2: Table Of Contents

    Installation ............................8 Pre-installation ........................8 Power Supply .......................... 8 Typical Room Layout Example ..................10 Location of Installation ......................11 Powering Up The MAXX-700 System ................12 Troubleshooting ..........................13 Operation Failure ........................13 Other Problems ........................13 Maintenance ..........................14 Repair .............................
  • Page 3: Introduction

    1.1 Intended Use The MAXX-700 lasers are intended to be used to align patients for medical diagnostic imaging and radiation therapy. They are intended to be used in a clinical or hospital setting by professionally trained medical staff.
  • Page 4: Laser Safety

    If the equipment or the lasers have been moved, it is advised to realign the lasers to the isocenter before use. CAUTION - The laser light is a risk for eye damage. The laser used in the MAXX-700 patient positioning system is a Class 2 laser. To prevent eye damage, do not allow patients to stare into the laser beam.
  • Page 5 There are different possible safety labels depending on whether the lasers are being used in Europe or North America. Safety Labels For Europe The wavelength is 450 nm for a blue laser. The wavelength is 532 nm for a green laser. The wavelength is 635 nm for a red laser.
  • Page 6: General Terms

    1.3 General Terms Name Description The Sagittal Laser The sagittal laser is a single vertical line projected from the laser. The Crosshair Laser The crosshair laser is a crossed line projected from the laser. The Isocenter The isocenter is a center point where the X, Y, and Z planes coincide. The PAN-ID The PAN-ID is the network identifier number of the wireless network within a room.
  • Page 7: Technical Specifications

    1.5 Technical Specifications Category Feature Specification X Axis - Horizontal Movement 20 mm (0.8in) +/-10 mm (from factory center) Y Axis - Vertical Movement 20 mm (0.8in) +/-10 mm (from factory center) Roll Axis - Horizontal Rotation 10º +/-5° Pitch Axis - Vertical Rotation 6º...
  • Page 8: Installation

    Installation Once the MAXX-700 unit has been received and the package has been opened it must be properly installed before operation. 2.1 Pre-installation The MAXX-700 Patient Positioning System provides an optical aid for rapid and accurate patient positioning and repositioning for therapy and virtual simulation machines.
  • Page 9 Figure 2-1 Power Supply for the European Union Figure 2-2 Power Supply for North America The Power Plug Figure 2-3 Laser Unit Bottom The power supply connector pushes into the base of the laser unit with a screw-on safety ring to keep the power supply connector in place during operation. •...
  • Page 10: Typical Room Layout Example

    2.3 Typical Room Layout Example Unit #3 Unit #1 Unit #2 Unit #4 Page 10...
  • Page 11: Location Of Installation

    Lasers can be installed to either mounting plates (Figure 2-4) or adjustable mounting brackets (Figure 2-5 and Figure 2-6). Mount the front plates to the walls first. After attach the MAXX-700 laser units to the plates. Figure 2-4 Mounting Plate -...
  • Page 12: Powering Up The Maxx-700 System

    Plug in the power cable into each laser unit. Please observe the following events (A to D) illustrated in Figure 2-7on Page 12. WARNING - Repeatedly turning on and off the MAXX-700 units can cause damage to the hardware. B. The...
  • Page 13: Troubleshooting

    Troubleshooting 3.1 Operation Failure If any operation cannot be completed successfully, the laser unit's network LED will start blinking to verify that there is a timeout. In most cases, this is due to a wireless communication problem. If linking to the unit is successful, you can repeat the previously failed operation. If linking fails, please follow the steps below: 1.
  • Page 14: Maintenance

    WARNING–No modifications may be made to the remote control as this will result in the Cemar Electro Inc. warranty becoming null and void. 4.2 Maintenance, Cleaning, and Disposal Once the MAXX-700 laser unit is installed, CEMARLINE lasers require no maintenance other than cleaning beam splitters and lenses according to standard optical practices.
  • Page 15 • Do not mix this device with general waste upon disposal. For proper treatment, recovery, and recycling, please contact your disposal local supplier for further information. In a case that a laser fails to function properly, please contact our customer service department as soon as possible.
  • Page 16: Product Warranty

    Product Warranty Cemar Electro Inc. warrants its patient positioning equipment to be free from defects in materials and workmanship under normal use and service for 2 years from the date of the shipment. The sole obligation of Cemar Electro Inc. under this warranty is to repair or replace without charge or to refund the purchase price, at the option of Cemar Electro Inc., of any parts which its examination shall have disclosed to be defective, provided that the buyer shall have given to Cemar Electro Inc., a written notice of the claimed defect no later than 7 days after the end of the warranty...
  • Page 17: Customer Service

    Customer Service Cemar Electro Inc. is committed to satisfying our customer’s needs. If you have any questions, comments, or suggestions regarding our products and services, please call or email us. Please contact a Cemar Electro Inc. distributor or a representative for a quotation or for a detailed description of our ordering policies, warranties, delivery policy, conditions of sale, damaged goods policy, and returned goods policy.
  • Page 18: Compliance Information

    7.2 Compliance Information EU Radio Equipment Type Compliance Declaration The MAXX-700 series which comprises wireless communication equipment complies with Directive 2014/53/EU. The full text of the EU Declaration of Conformity is available at the following internet address: http://www.cemarmedical.com Page 18...
  • Page 19: List Of Symbols

    List of Symbols This device complies with the Medical Devices Directive & EMC Directive of the European Economic Community. Do not mix this device with general waste upon disposal. For proper treatment, recovery, and recycling, please contact your dealer or supplier for further information. Date of Manufacture Manufacturer Part Number...