TPC P555 Operation Manual

Air polishing system

Advertisement

Quick Links

P555
Operation Manual
Air Polishing System
TPC
851 S. Lawson St.
City Of Industry, CA 91748
P626-810-4337 Fax 626-810-4245
www.tpcdental.com
REV. 1-1-13

Advertisement

loading
Need help?

Need help?

Do you have a question about the P555 and is the answer not in the manual?

Questions and answers

Summary of Contents for TPC P555

  • Page 1 P555 Operation Manual Air Polishing System 851 S. Lawson St. City Of Industry, CA 91748 P626-810-4337 Fax 626-810-4245 www.tpcdental.com REV. 1-1-13...
  • Page 2 Warning! This device is intended for use by a dentists and or dental hygienists only. Do not use on the following patients. 1) Those that require restriction on sodium intake. 2) Those with grave ulcers in digestive organs. 3) Those with hepatic dysfunction. 4) Those with lung or cardiac dysfunction. 5) Those with hyperemia, hemorrhage, or inflammation in the oral cavity. 6) Those with damage or abnormality in oral cavity.
  • Page 3 1. Caution before use This product is used for jetting powder. It may clog by moist air or by water. Do not use a delivery system in which the air pressure may increase rapidly when it clogged. How to check the air pressure of the delivery system. •...
  • Page 4 4. Sterilization The manufacturer recommends autoclaving this device Disassemble Air prophy unit into the following components. • Powder case and nozzle. • Powder case into the powder case body and the powder case cap. • Powder case cap into the O-ring, powder case cover, thin gasket and serrated cover ring. Autoclave e for 20min, at 121°C, or 15min, at 132°C.