Ec/Tds Calibration - Hanna Instruments HI 9812 Instruction Manual

Portable ph/ec/tds meters
Hide thumbs Also See for HI 9812:
Table of Contents

Advertisement

EC/TDS
EC/TDS

EC/TDS CALIBRATION

CALIBRATION
CALIBRATION
CALIBRATION
EC/TDS
EC/TDS
CALIBRATION
Accessories needed:
• Use HI 7031 or HI 8031, 1,413 µS/cm (=1.413 mS/cm) EC
solution for HI 9812 and HI 9813 or HI 7032, 1382 ppm
TDS solution for HI 9812 or HI 70442, 1500 ppm TDS
solutions for HI 9813.
Note: the conversion between EC and TDS is made by a built-in circuit,
hence it is requested to calibrate the meter only in EC or TDS
range. The other range is thus automatically calibrated.
PROCEDURE
• Pour approximately 4 cm (1½") of a con-
ductivity calibration solution (e.g. HI
7031/HI 8031) into a beaker. If pos-
sible, use plastic beakers to minimize any
EMC interference.
• Immerse the combined electrode in the solution.
• Wait for a couple of minutes for thermal equilib-
rium to be reached.
• Tap the combined electrode on the bottom, then
shake it lightly while rotating to make sure no air
bubbles remain trapped inside the probe.
• Select the appropriate range, e.g. µS/cm for HI 9812 or the mS/
cm key for the HI9813.
• Turn the EC/TDS calibration knob until the
display shows the EC or TDS reading at
25°C.
11