Summary of Contents for Ludlum Measurements 2224-1
Page 1
LUDLUM MODEL 2224-1 SCALER / RATEMETER January 2012 Serial Number 125559 and Succeeding Serial Numbers...
Page 2
LUDLUM MODEL 2224-1 SCALER / RATEMETER January 2012 Serial Number 125559 and Succeeding Serial Numbers...
Page 3
RETURN OF GOODS TO MANUFACTURER If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information.
Safety Considerations Environmental Conditions for Normal Use Cleaning Instructions and Precautions Warning Markings and Symbols Calibration and Maintenance Calibration Establishing an Operating Point Meter Calibration Detector Overload Calibration Maintenance Recalibration Batteries Technical Principle of Operation January 2012 Ludlum Measurements, Inc.
Technical Manual Section 1 Section Introduction he Model 2224-1 is a portable microprocessor based radiation survey instrument used to measure and discriminate low-level alpha/beta radiation when used with an alpha/beta scintillation or proportional detector. The data is displayed by an analog ratemeter and a six-digit liquid crystal display (LCD) counter.
The Model 2224-1 serial number is located on the front panel below the battery compartment. Most Ludlum Measurements, Inc. detectors have a label on the base or body of the detector for model and serial number identification.
A mild electric shock may occur if you make contact with the center pin of the input connector. Switch the Model 2224-1 to the ˝ ˝ position before connecting or disconnecting the cable or detector.
) and proceed with use. Principle of Operation The Model 2224-1 is to be used in combination with alpha/beta scintillation or proportional detectors. The instrument uses pulse heights discrimination to distinguish between alpha and beta pulses from the radiation detector.
MODEL 2224-1 Technical Manual Section 3 Section Specifications : four linear range multiples of ×1, ×10, ×100 and ×1000; used in Ranges combination with the 0-500 meter dial, providing an overall range of 0-500 kcpm : beta threshold ( ) is adjustable from -2 to -10 millivolts Thresholds (mV);...
Page 12
MODEL 2224-1 Technical Manual Section 3 : two standard ˝D˝ cell batteries Power : instrument calibration change less than 3% to Battery Dependence battery endpoint : exceeds 350 hours with a fresh set of alkaline ˝D˝ cell Battery Life batteries : -20 to 50 °C (-4 to 122 °F)
MODEL 2224-1 Technical Manual Section 4 Section Description of Controls and Functions Operator Controls (or Range Selector Switch): a six- OFF/BAT/X1000/X100/X10/X1 Switch position rotary switch to select the analog meter range multipliers and check the battery status. When switched to the...
MODEL 2224-1 Technical Manual Section 4 regardless of the switch position and will continue to function when the channel is not selected for display. This allows the operator to view each channel separately or together by simply selecting the appropriate switch position.
Page 15
MODEL 2224-1 Technical Manual Section 4 The ratio is selected from the following table. (O is open and C is closed) SWITCH DIVIDE BY RATIO 1000 Not used Dipswitch 3 & 4 (SPARE) : a one-pole DIP switch used to select tone Dipswitch 5 (TONE) discrimination between alpha and beta count channels.
Page 16
MODEL 2224-1 Technical Manual Section 4 : a multi-turn potentiometer that provides a means to vary the detector current overload set point. : a multi-turn potentiometer used to set the maximum limit to 2000 Vdc. : a multi-turn potentiometer used to adjust the high-voltage test reading to correspond with the actual high-voltage output.
This is typical of industrial or construction sites.) Cleaning Instructions and Precautions The Model 2224-1 Scaler/Ratemeter may be cleaned externally with a damp cloth, using only water as the wetting agent. Do not immerse the instrument in any liquid. Observe the following precautions when cleaning: 1.
Ludlum Measurements, Inc. The Model 2224-1 Scaler / Ratemeter is marked with the following symbols: CAUTION (per ISO 3864, No. B.3.1) – designates hazardous live voltage and risk of electric shock.
1. Connect a Ludlum Model 500 Pulser or equivalent to the Model 2224-1. 2. Switch the Model 2224-1 channel selector to the β position. Adjust the beta threshold ( ) for 3.5 mV and the window ) for 30 mV.
The adjustment is on the front of the meter bezel. It must be adjusted to zero with the range selector switch in position. 2. Connect the Model 500 Pulser to the Model 2224-1 with the appropriate cable. Rotate the Model 2224-1 range selector switch to the ×100 position.
Page 21
1.0 kV on the meter dial. Adjust the output from 500 to 1500 Vdc and confirm that the Model 2224-1 meter corresponds to the external voltmeter within 1 0% of each reading.
MODEL 2224-1 Technical Manual Section 6 11. Replace the Model 2224-1 cover (can) and proceed with use. ETECTOR VERLOAD ALIBRATION Note: The detector operating voltage ( ) must be determined and set before the (overload) adjustment is performed. If the detector operating voltage is re-adjusted, the overload adjustment must be re-adjusted.
Maintenance Instrument maintenance consists of keeping the instrument clean and periodically checking the batteries and the calibration. The Model 2224-1 instrument may be cleaned with a damp cloth (using only water as the wetting agent). Do not immerse instrument in any liquid. Observe the following precautions when cleaning: 1.
Page 24
MODEL 2224-1 Technical Manual Section 6 Note: Never store the instrument over 30 days without removing the batteries. Although this instrument will operate at very high ambient temperatures, battery seal failure may occur at temperatures as low as 37 °C (100 °F).
MODEL 2224-1 Technical Manual Section 7 Section Technical Principle of Operation Refer to the Amplifier/Power Supply Board Schematic, Drawing 390 × 104, for the following: Detector Input/ Amplifier Negative-going detector pulses are coupled from the detector through C021 to amplifier U021. R023 and CR021 protects the input of U021 from inadvertent shorts.
Page 26
MODEL 2224-1 Technical Manual Section 7 U111 to provide a time delay for the μP clock cycle to complete before the next alpha pulse can be recognized by the μP. Beta pulses from pin 1 of U011 are coupled to univibrator U101. Pulses are coupled to the μP from pin 7 of U101 as long as pins 3 and 13 of U011...
Page 27
MODEL 2224-1 Technical Manual Section 7 Detector Overload A voltage drop is developed across R121 and sensed by comparator U012 as detector current increases. When the voltage at pin 3 of U012 goes below pin 2, pin 1 goes low, illuminating the...
Page 28
MODEL 2224-1 Technical Manual Section 7 period. At full meter deflection, the low pulse period, leading edge to leading edge, will be 166 μs, 500 cpm = 130 μs, 400 cpm = 104 μs, 200 cpm = 52 μs, 100 cpm = 26 μs, and 0 = no pulse or +5 V. The pulses are inverted by Q302 on the amplifier/power Supply board and then integrated by R403, C401.
MODEL 2224-1 Technical Manual Section 8 Section Troubleshooting ccasionally you may encounter problems with your LMI instrument or detector that may be repaired or resolved in the field, saving turn-around time and expense in returning the instrument to us for repair. Toward that end, LMI electronics technicians offer the following tips for troubleshooting the most common problems.
Page 30
MODEL 2224-1 Technical Manual Section 8 SYMPTOM POSSIBLE SOLUTION No power (or meter Check battery contacts. Clean them does not reach with rough sandpaper or use an engraver to clean the tips. TEST BAT OK mark) (continued) Check for loose or broken wires, especially between the main board and the calibration board.
Page 31
MODEL 2224-1 Technical Manual Section 8 SYMPTOM POSSIBLE SOLUTION Meter goes full-scale Open the instrument “can” and or “pegs out” check for loose wires. (continued) Ensure that the instrument’s “can” is properly attached. When attached properly, the speaker will be located on the left side of the instrument.
To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products. With many different agencies –...
Need help?
Do you have a question about the 2224-1 and is the answer not in the manual?
Questions and answers