Guidance and manufacturer's declaration-electromagnetic emissions

This product is intended for use in the electromagnetic environment specified below.
The customer or the user should assure that it is used in such an environment.
Emission test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class B
Harmonic emissions
D
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Complies
IEC 61000-3-3
User's Manual | MDP156
Electromagnetic environment-guidance
The product uses RF energy only for its
internal function. Therefore, its RF emis-
sions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The product is suitable for use in all
establishments, including domestic
establishments and those directly con-
nected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.

Guidance and manufacturer's declaration - electromagnetic immunity

The product is intended for use in the electromagnetic environment specified below. The cus-
tomer should assure that it is used in such as environment.
Immunity test
IEC 60601 test level
Electrostatic dis-
±8 kV contact
charge (ESD)
±15 kV air
IEC 61000-4-2
±2 kV for power sup-
Electrical fast tran-
ply lines
sient/burst
±1 kV for input/ out-
IEC 61000-4-4
put lines
Surge
±1 kV line(s) to line(s)
±2 kV line(s) to earth
IEC 61000-4-5
< 5%
(> 95% dip in) for 0.5
cycle
Voltage dips, short
40%
interruptions and
(60% dip in) for 5
voltage variations on
cycles
power supply input
70%
lines
(30% dip in) for 25
IEC 61000-4-11
cycles
< 5%
(> 95% dip in) for 5
sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
Chapter 4
INDICATIONS FOR USE
Electromagnetic
Compliance level
environment-
guidance
Floors should be
wood, concrete or ce-
ramic tile. If floors are
±8 kV contact
covered with synthetic
±15 kV air
material, the relative
humidity should be at
least 30%.
±2 kV for power sup-
Mains power quality
ply lines
should be that of a
±1 kV for input/ out-
typical commercial or
put lines
hospital environment.
Mains power quality
±1 kV differential
should be that of a
mode
typical commercial or
±2 kV common mode
hospital environment.
< 5%
Mains power quality
should be that of a
(> 95% dip in) for 0.5
typical commercial
cycle
or hospital environ-
40%
ment. If the user of
(60% dip in) for 5
the product requires
cycles
continued operation
during power mains
70%
interruptions, it is
(30% dip in) for 25
recommended that
cycles
the product powered
< 5%
from an uninterrupt-
(> 95% dip in) for 5
ible power supply or a
sec
battery.
Power frequency
magnetic fields
should be at levels
3 A/m
characteristic of a
Power
typical location in a
typical commercial or
hospital
12