Multifunctional handheld water quantity meter

Advertisement

Quick Links

Y600-A
Multifunctional Handheld
Water Quantity Meter
Manual
Yosemite Technologies Co.,Ltd

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Y600-A and is the answer not in the manual?

Questions and answers

Summary of Contents for Yosemite Y600-A

  • Page 1 Y600-A Multifunctional Handheld Water Quantity Meter Manual Yosemite Technologies Co.,Ltd...
  • Page 2: Table Of Contents

    CONTENTS Preface..............................- 2 - 1 Overview............................- 3 - 1.1 Introduction ..........................- 3 - 1.2 Specifications ..........................- 4 - 2 Setting-up ............................- 5 - 2.1 Configuration ..........................- 5 - 2.2 Installing the batteries ......................- 5 - 2.3 Installing the sensor .........................
  • Page 3: Preface

    Carefully unpack the instrument and accessories from the shipping container, and inspect for possible damage during shipping. Check received parts with items on the packing list. If any parts or materials are damaged or missing, please contact Yosemite Technologies customer service or the authorized distributor immediately.
  • Page 4: Overview

    1 Overview The Y600-A instrument is independently developed by Yosemitech. It provides keyboard and backlight with comprehensive functions, simple operation and brief interface. The instrument is a portable device that can automatically recognize Yosemite's digital sensors including optical dissolved oxygen, 4-electronic conductivity, turbidity, chlorophyll, BGA, OIW and TSS ,and can measure, display, log.
  • Page 5: Specifications

    ±0.2℃ Working temperature 0~ 50℃ -40 ~ 85℃ Storage temperature Calibration method One-point or two-point calibration Sensor cap lifetime One year typical Y600-A Instrument Size 220 x 96 x 44mm Weight 460g Power Two 18650 rechargeable batteries Working temperature 0~50℃...
  • Page 6: Setting-Up

    2 Setting-up 2.1 Configuration Item Number Note Y600-A Instrument Y506-A optical sensor 5m default,optional 18650 rechargeable battery USB cable line Micro USB Wristband Screwdriver To open the battery compartment O-ring Sensor cap sealing part Sponge To moisturize sensor cap Protect Cap...
  • Page 7: Installing The Sensor

    2.3 Installing the sensor The sensor cable has a red slot on the mini-BNC connector which must align with the connector at the base of the instrument. Then insert gently and turn clockwise until the “click” sound is heard. When remove the sensor, push the sensor cable connector in and turn out anticlockwise. Install:...
  • Page 8: Operation

    3 Operation Press key to turn the Y600-A on. When there is no sensor, the measurement interface will display “No Signal!”. If the dissolved oxygen Sensor is connected, the display will show DO values as shown in the figure below.
  • Page 9: Date/Time

    11:25 Menu Menu Press the Menu key to switch to Date/Time menu interface. To select a menu System item, use the “▲▼” key to Calibration highlight the selection and press Data Log the “↙” button. To exit a menu or menu item, press the ESC key.
  • Page 10 Auto Power 11:25 In Auto Power, there are four Auto Power mode: OFF indicates the power supply is normally open, When set to 5 min , 10 min, 15 min, 5 min the instrument will shut off after 5 10 min 15 min minutes, 10 minutes, 15 minutes of no operation,...
  • Page 11 Salinity This menu can setup salinity value. The 11:25 amount of DO increases as salinity level Salinity decreases. instrument Unit: compensate for the deviation of DO value Current: Input: 000..0 caused by salinity. Make adjustments Save with the ▲or ▼keys and then press the “↙”...
  • Page 12 3.3 Calibration The instrument supports DO both single-point and two-points user calibration. Single-point DO calibration means "100.0%SAT" calibration which performed in water saturated air or air saturated water. For best results, we strongly recommend to use air saturated water solution. Two-points calibration consist of 100.0%SAT and 0%SAT.
  • Page 13: One-Point Calibration

    One-Point calibration ⚫ Follow the screen instruction “In 100%”,put the DO sensor into 100% SAT solution, and press the "↙" key to move on. ⚫ Then “Waiting 3 min…”means wait for 3 minutes to stabilize the data. The measured DO value in % unit will flash on the bottom of screen.
  • Page 14: Two-Points Calibration

    11:25 OnePoint In 100% Success 91.56 Locked value at 100% 91.56 Cali Done! Two-points calibration First performed 100% SAT calibration, then start calibrating 0% SAT. Follow the screen instruction “In 0%”,put the DO sensor into 0% SAT solution, and press the "↙" key to move on. Wait for 3 minutes to stabilize the data.
  • Page 15: Sensor Cap

    3.4 Data Log Data Store The Y600-A instrument has 4096 measurements that can be logged to memory and can be programmed to store measurements in two logging formats. When set to Data Log/Data Store/Store 1 Times, the user samples start to be logged a time. When set to Data Log/Data Store/Store 50 Times, the instrument will continue logging 50 times with an interval of 2 seconds.
  • Page 16: Data View

    Data View Max Size Indicates numbers 11:25 measurements to be logged. Data View Use Size Max Size : 4096 Indicates numbers Use Size : 54 measurements have been logged. Current Sel : 54 Current Sel Indicates the selected sample item 190920 15:32:37 that contains the data to be recalled.
  • Page 17: Cal

    Follow the instructions to install the software on your computer and connect the supplied computer interface USB cable to the Y600-A and to a computer USB port. Then the COM PORT can identify the port. Select the download data path and click Load. The file format is. CSV.
  • Page 18: Maintenance

    4 Maintenance 4.1 The instrument maintenance The 600-A instrument require very little maintenance to keep it performing well. If the instrument case needs cleaning, wipe it off with a soft cloth. Please keep the battery compartment dry during using. 4.2 DO maintenance Optical dissolved oxygen sensor is a maintenance free sensor, however to ensure its accuracy specifications, we strongly recommend to perform routine maintenance based on following schedule.
  • Page 19 deionized water or purified water can be added for wetting; if DO membrane is kept in a dry state for a long time, it will lead to inaccurate measurement and greatly shorten the life of DO membrane. Inspect sensor cable :...
  • Page 20: Common Errors And Trouble Shooting

    5 Common Errors and Trouble Shooting Table 5-1 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the Y600-A instrument. If your symptom is not listed, or if none of the solutions solves your problem, please contact us.
  • Page 21 Yosemite Technologies Ltd. Address:18 Dongchang Rd. Building 25,Suzhou Industrial Park, 215126, China :215123 ZIP code :+86-512-87663997 Fax:+86-512-87663911 E-mail:sales@yosemitech.com Web:www.yosemitech.com - 20 -...