Heathkit VC-3
Voltage Calibrator.

Introduction:

The August 2013 Heathkit of the Month (#51)
covered the IG-4505 Deluxe Oscilloscope Cali-
brator. It also briefly covered the VC-3 Voltage
Calibrator using information gleaned from cata-
logs. To my surprise Bill - K6WHP recently pre-
sented me with a VC-3 complete with manual.
One look at the unusual schematic and I knew it
deserved an article all its own.
The VC-3 (Figure 1), third in the Heathkit line
of oscilloscope voltage calibrators that were
first introduced in 1952, appeared in late 1956.
Figure 2 shows an ad for the Heathkit VC-3
from the 1958 catalog. It remained in produc-
tion through 1962 at a price of $12.50. In those
days the oscilloscope was the most prolific of
Heathkit's products. From the O-1 oscilloscope
(Heathkit's first electronic kit) introduced in
September of 1947 through the end of 1956
Heathkit sold 14 different models, the O-1
through O-11, the OL-1, the OM-1 and OM-2.
All had 5" CRTs except the OL-1 (3"), and all
had AC coupled amplifiers and multivibrator
sweep circuits. DC coupling and triggered
sweep were on the horizon for Heathkit, but
Figure 2: Heathkit VC-3 ad from the main 1958 Catalog
Copyright 2015, R. Eckweiler & OCARC, Inc.
Heathkit of the Month #64 - VC-3 Voltage Calibrator
Figure 1: Heathkit VC-3 Voltage Calibrator
not yet incorporated. The first Heathkit oscillo-
scope featuring these improvements came in
the fall of 1958, the OP-1 Professional Oscillo-
scope Kit. It sold for $179.95, about 2-1/2 times
more than their previous top-of-the line
Heathkit oscilloscope.
The Heathkit VC-3:
Prior to the OP-1, most of the hobby-grade os-
cilloscopes on the market had at best a 1 volt
peak-to-peak test point that one could use to
approximate the voltage of a wave form. Since
the gains were mostly uncalibrated and the at-
tenuator steps a full decade apart, measuring
voltage was difficult without a way to compare
it with a known signal. The VC-3 provides the
ability to make that comparison easily.
The VC-3 puts out a rather impressive ap-
proximate 1 kc square wave that has an accu-
Page 1 of 5

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VC-3 and is the answer not in the manual?

Questions and answers

Summary of Contents for Heathkit VC-3

  • Page 1 DC coupling and triggered proximate 1 kc square wave that has an accu- sweep were on the horizon for Heathkit, but Figure 2: Heathkit VC-3 ad from the main 1958 Catalog Copyright 2015, R. Eckweiler & OCARC, Inc. Page 1 of 5...
  • Page 2 HEATH COMPANY BENTON HAR- BOR, MICHIGAN on two lines at right bottom. The VC-3 is housed in a 4–3/4 H x 7–3/8 W x 4–1/8 D gray cabinet and a dark gray front panel with white nomenclature. The cabinet...
  • Page 3 HOM rev. new Heathkit of the Month #64 - VC-3 Voltage Calibrator input is a small price to pay for being able to producing about 230 VDC relative to the center- make reasonable voltage measurements on an tap of the transformer. The DC is filtered by a 20 otherwise uncalibrated piece of test equipment.
  • Page 4 Heathkit of the Month #64 - VC-3 Voltage Calibrator HOM rev.new and filament will not exceed the filament to cathode maximum rating of 100 VDC. The VC-3 Multivibrator Circuit: The two 6AW8 tube sections are wired as a free running multivibrator. The triode section acting as one half the multivibrator and the pentode section as the other.
  • Page 5 HOM rev. new Heathkit of the Month #64 - VC-3 Voltage Calibrator position, plate voltage is removed from the 6AW8 I’ve gotten numerous requests to write on the preventing calibrator noise on the output. HW-12 & 12A series. If anyone has a PDF copy...

Table of Contents