Ultrasonic test cell

Advertisement

TDC1000 Ultrasonic "Test Cell" Assembly Instructions
Rev 1.0

Supplies required:

1. CYA glue w/Latex
2. CYA Accelerator
3. Plastic "test cell" container
4. 400 grit sandpaper
5. Alcohol
6. Hot glue gun
7. Ultrasonic transducers w/attached wires
8. Spray duster/ air gun

Assembly Steps:

1. Roughen 1cm square area on lower side of plastic container with sandpaper
2. Degrease with alcohol
3. Dry area with air gun
4. Apply CYA glue to 1cm square area