Geotech High Vacuum Sampler Installation And Operation Manual

Advertisement

Quick Links

Geotech High
Vacuum Sampler
Installation and Operation Manual
Rev 05/06/13 Part # 12151001

Advertisement

Table of Contents
loading

Summary of Contents for Geotech High Vacuum Sampler

  • Page 1 Geotech High Vacuum Sampler Installation and Operation Manual Rev 05/06/13 Part # 12151001...
  • Page 2: Table Of Contents

    Table of Contents SECTION 1: SYSTEM DESCRIPTION ................3 SECTION 2: SYSTEM INSTALLATION ................6 SECTION 3: SYSTEM OPERATION .................. 7 SECTION 4: SYSTEM MAINTENANCE ................8 SECTION 5: SYSTEM TROUBLESHOOTING ..............9 SECTION 6: SYSTEM SPECIFICATIONS ............... 11 SECTION 7: SYSTEM SCHEMATIC ................12 SECTION 8: REPLACEMENT PARTS AND ACCESSORIES..........
  • Page 3 DOCUMENTATION CONVENTIONS This uses the following conventions to present information: An exclamation point icon indicates a WARNING of a situation or condition that could lead to personal injury or death. You should not proceed until you read and thoroughly understand the WARNING message.
  • Page 4: Section 1: System Description

    Section 1: System Description Function and Theory The High Vacuum Sampler is an air-sampling device that uses a compressor to draw air from a sampling source and convey that sample into sealable sample bags. When an external 12 VDC power source connects to the unit, the compressor is capable of producing a vacuum of up to 27 in-Hg.
  • Page 5 Figure 1-2 Figure 1-3 – Example of ALTEF 1 Liter Bag Envirotecnics Tel: +34 872 080 542 envirotecnics@envirotecnics.com www.envirotecnics.com...
  • Page 6 Figure 1-4 – Example of Tedlar 1 and 3 Liter Bags Envirotecnics Tel: +34 872 080 542 envirotecnics@envirotecnics.com www.envirotecnics.com...
  • Page 7: Section 2: System Installation

    Securely plug the circular plug connector on the opposite end of the cable to the corresponding receptacle on the face of the controller. Turn on the High Vacuum Sampler briefly and verify its operation. Place your thumb over both ports: Air should be blowing out the AIR OUT port and vacuum should be present at the VACUUM IN port.
  • Page 8: Section 3: System Operation

    Attach all tubing to the appropriate ports and proceed with collecting air samples. When operating the High Vacuum Sampler the vacuum gauge should never exceed -25 in Hg. Should the gauge ever hover between -22 and -25 in Hg, it would mean there is a significant clog somewhere along the input line from the sample source, or the vacuum at the source is greater than the High Vacuum Sampler’s potential.
  • Page 9: Section 4: System Maintenance

    Unit must be returned to Geotech Environmental Equipment for any service. In order to ensure a long service life, keep the High Vacuum Sampler clean. Often a soft, damp cloth can be used to remove dust and dirt from the exterior surfaces of the enclosure. In...
  • Page 10: Section 5: System Troubleshooting

    • If the 15 Amp circuit breaker is tripped, push to reset. If it trips again, call Geotech Service at 1-800-833-7958 to arrange for the equipment to be sent back to a factory authorized repair location.
  • Page 11 If there is vacuum then check the tubing between the VACUUM IN port and the sample source for cuts, kinks and holes. If there is no vacuum then call Geotech Service at 1-800-833-7958 to arrange for the equipment to be sent back to a factory authorized repair location.
  • Page 12: Section 6: System Specifications

    Section 6: System Specifications Power Requirements: 72-84 W, 12-14 VDC input at 7.5 Amps Nominal Operating Current: 6 Amps DC Over Current Protection: 15 Amps Maximum Operating Vacuum: 27 in Hg Ambient Operational Temperature: 50°F-104°F (10°C-40°C) Dimensions: 6.70" H x 4.80" x 11.40" W (17.02cm H x 12.92cm D x 28.96cm W) Weight: 11.4lbs (5.17kg)
  • Page 13: Section 7: System Schematic

    Section 7: System Schematic Figure 7-1 – High Vacuum Sampler Wiring Diagram Envirotecnics Tel: +34 872 080 542 envirotecnics@envirotecnics.com www.envirotecnics.com...
  • Page 14: Section 8: Replacement Parts And Accessories

    Section 8: Replacement Parts and Accessories Parts List Parts Description 82150020 HIGH VACUUM SAMPLER PPP103001 QCK CNCT,NCKL,1/4X1/8MPT,PUSH/ PULL 87050501 TUBING,PE,.170x1/4,FT POLYETHYLENE 87050509 TUBING,FEP,.170x1/4,FT 51150063 POWER SUPPLY,EXTERNAL,AC ADAPT 12V,100W 81150019 ADAPTOR,CONTROLPRO-WELL CAP INCLUDES 10FT TUBING 57500008 ASSY,POWER CORD,DC W/ AMP 17500035...
  • Page 15 DOCUMENT REVISIONS EDCF# DESCRIPTION REV/DATE Previous Release 12/17/12 Edited description for spare part # 87050509. 1570 05/06/13 Added Revision History Table - SP Envirotecnics Tel: +34 872 080 542 envirotecnics@envirotecnics.com www.envirotecnics.com...
  • Page 16: The Warranty

    Geotech agrees to repair or replace, at Geotech’s option, the portion proving defective, or at our option to refund the purchase price thereof. Geotech will have no warranty obligation if the product is subjected to abnormal operating conditions, accident, abuse, misuse, unauthorized modification, alteration, repair, or replacement of wear parts.
  • Page 17 Geotech Environmental Equipment, Inc. 2650 East 40th Avenue Denver, Colorado 80205 (303) 320-4764 ● (800) 833-7958 ● FAX (303) 322-7242 Envirotecnics Global Service SL Oficinas y Fabricación:Calle dels Pirineus s/n (esquina Espinau) 17460 - Celrà - Girona Tel: 872 080 542 Fax: 872 080 543 Delegación Madrid: Avda.