Veinlite
NEO
LED Window [A]
End Button [B]
Side Button [C]
Battery Cover [D]
READ INSTRUCTIONS BEFORE USE
TransLite LLC
345 Commerce Green Blvd
Sugar Land TX 77478 USA
+1 281 240 3111
www.veinlite.com
info@veinlite.com
Caution
Veinlite NEO is designed for external
examination only.
To avoid contamination of the device, Veinlite
NEO must be used with the disposable plastic
cover.
To avoid the spread of infection, always change
the disposable plastic cover between patients
and wipe the new cover with a recommended
germicidal wipe.
Always turn the LEDs on after the device is in
position. Do not look directly at the LEDs when
they are illuminated or shine the light from them
directly into anyone's eyes.
Do not immerse the device in liquid.
Do not autoclave the device.
Introduction
The Veinlite NEO is a hand-held battery powered
transillumination device for visualizing veins and
arteries in neonates.
Applications
To assist in venous and arterial access in
neonatal pediatric patients.
Cap [E]
Cover [F]
Operating Directions
Device:
1 Place Veinlite NEO against the skin with the
LED Window [A] in contact with the skin.
2 Depress the End Button [B] to turn
on the LED.
3 Apply gentle pressure to the
Veinlite NEO to maintain optimal skin contact.
4 Device will turn off after cycling through the
colors or you can depress and hold the End
Button [B] for one second to turn device off.
5 When you turn on a device with a low battery
the device will shut itself off.
6 Device will shut off automatically after 2
minutes.
Disposable Covers:
The Veinlite NEO should always be used with a
single-use disposable plastic cover to protect it
from contamination and to prevent the spread
of infection between patients.
1 Place the Veinlite NEO
inside the disposable
plastic cover. Check the
cover to make sure it is intact
and the flap is on the side
away from the patient.
2 These covers are not sterile.
Disinfect the cover with 70%
isopropyl alcohol or other
medically accepted disinfectant.
3 Discard the single-use
disposable plastic cover
after each patient.
The device is shipped with 50 covers.
Additional covers can be ordered from
TransLite or any of its authorized dealers.
Brightness Control:
Depress the Side Button [C] to
change the brightness. Each depression
of the button will cycle through low,
medium, and high brightness levels. The lowest
brightness level is the default setting.