Important: Retain this card for reference and carfully read the operating instructions before use
INSTALLATION
Check you have air and water
Connect your 4-hole tubing
from your unit 4 hole tubing
to Polisher connector
ADDING POWDER
Remember ! Do not add above fill line
ADJUSTING THE CONTROLS
Use the fingertip test to achieve correct powder/water ratio
DELDENT
Ltd.
Petach Tikva, Israel
19 Keren Kayemet St. 49372 Petach Tikva, Israel
Tel: 972-3-9049668 Fax: 972-3-9049681
e-mail: info@deldent .com
www.deldent.com
screw up tight to
prevent leakage
Over filling will cause clogging
Start at the lowest end and gradually build
water intensity. Inadequate water will produce
a spluttering, intermittent stream.
Use only enough water for a steady stream.
CORRECT ADJUSTMENT 'FINGERTIP' TEST
1. In the 'polishing mode', open the water control knob (H
2.Hold nozzle approximately 1 cm (3/8") from the finger tip.
3. Activate unit, and gradually reduce the amount of water in the stream
by means of the water control knob.
4. A point will be reached when a spot of powder begins to collect
on the fingertip. At this point inadequate water is present in the stream.
5. Increase the water flow until the spot of powder on the fingertip just disappears.
TROUBLE SHOOTING
1.
Disconnect handpiece
2.
Depress foot pedal and check flow from handpiece connector
Your foot control will now
activate the Polisher
3.
Clean components in ultrasonic bath in warm water or mild vinegar solution
4.
Dry thoroughly
Warning:
Always use a dry air supply. Drain water filter on your compressor daily.
REPLACING SPRAY HEAD TUBING
Less powder = less intense
cleaning
AUTOCLAVING HANDPIECE
Rinse/Polish mode selector
Polish = water + air + powder
Rinse = water + air only
For USA technical support call:
1-860 246 0359
O) fully.
2
Unblocking a handpiece
Air
Water
Avoid sharp bends in tubing
Always autoclave disassembled