TxIsoRail 4-20 mA
ISOLATED TEMPERATURE TRANSMITTER – OPERATION MANUAL V1.0x H
SPECIFICATIONS
Senso r inpu t : Use r defined. The suppo r ted senso r s a r e listed in Table 1,
along with thei r maximum r anges.
The r mocouples: Types J, K, R, S, T, N, E and B to IEC 60584 ( ITS - 90 ) .
Impedance >> 1 M
Excitation: 180 µA, 2 o r 3 - wi r e connection ( fo r 2 - wi r e
P t 100:
senso r s, tie, te r minals 2 and 3 togethe r) .
α= 0.00385, acco r ding to IEC 60751 ( ITS - 90 ) .
Vol t age:
0 to 50 mVdc, 0 to 10 Vdc. Impedance >> 1 M
No t e: 0 - 10 Vdc input type r equi r es an inte r nal jumpe r
*
switching.
0 to 20 mA, 4 to 20 mA. Impedance 15.0
Sensor Type
Range
Thermocouple K
-150 to 1370 °C / -238 to 2498 ºF
Thermocouple J
-100 to 760 °C / -148 to 1400 ºF
Thermocouple R
-50 to 1760 °C / -58 to 3200 ºF
Thermocouple S
-50 to 1760 °C / -58 to 3200 ºF
Thermocouple T
-160 to 400 °C / -256 to 752 ºF
Thermocouple N
-270 to 1300 °C / -454 to 2372 ºF
Thermocouple E
-90 to 720 °C / -130 to 1328 ºF
Thermocouple B
500 to 1820 °C / 932 to 3308 °F
Pt100
-200 to 600 °C / -328 to 1112 ºF
0 to 50 mV
Voltage
*
0 to 10 V
0 to 20 mA
Current
4 to 20 mA
Table 1 – Transmitter input sensors'
Ou t pu t: 2 - wi r e 4 - 20 mA, linea r with r espect to the measu r ed signal.
Resolu t ion : 0.001 mA ( 14 bits )
To t al Accu r acy :
Bette r than 0.3 % of the maximum span fo r
the r mocouples and 0.2 % fo r Pt100 and voltage;
Response Time : < 500 ms
Isola t ion : Between the senso r and the 4 - 20 mA loop ( 1000 V / 1 min ) .
Powe r Supply: 12 to 35 Vdc, ac r oss the t r ansmitte r ;
Maximum Load ( RL ) : RL ( max. ) = ( Vdc – 12 ) / 0.02 [Ω]
We r e: Vdc= Powe r supply voltage
Ope r a t ing Tempe r a t u r e: - 40 to 85 ° C
Humidi t y: 0 to 90 % UR
Elec tr omagne t ic Compa t ibili t y: EN 50081 - 2, EN 50082 - 2
In t e r nal p r o t ec t ion agains t pola r i t y inve r sion.
Cold junc t ion compensa t ion f o r t he r mocouples.
Ω
Ω .
Ω
( + 1.9 Vdc ) .
Minimum measurement span
100 °C
100 °C
400 °C
400 °C
100 °C
100 °C
100 °C
400 °C
40 °C
5 mV
1 V
2 mA
2 mA
CONFIGURATION
If the t r ansmitte r is al r eady configu r ed as r equi r ed by the application ( senso r
type, r ange, etc ) , it may be installed and used r ight away. Howeve r , if a
distinct configu r ation is r equi r ed, this can be done th r ough the TxCon f ig
so ft wa r e and the TxCon f ig In t e rf ace.
The TxConfig inte r face and softwa r e can be pu r chased f r om the manufactu r e r
o r at its autho r ized dist r ibuto r s and r ep r esentatives. Updates fo r the softwa r e
a r e available at ou r website. To install the TxConfig softwa r e, r un the
Tx_se t up.exe file and follow the inst r uctions.
The TxConfig inte r face connects the t r ansmitte r to the PC, as shown in
Fig. 1.
Fig. 1 – TxConfig Interface USB connections
Once the connection is accomplished, the softwa r e shows the configu r ation
options of the t r ansmitte r model attached. Access the Help fo r usage
inst r uctions.
The TxConfig sc r een in shown in Fig. 2.
Fig. 2 – TxConfig main screen
The fields in the sc r een mean:
1. Inpu t Senso r : Choose the desi r ed tempe r atu r e senso r among the
available options. See Table 1.
2. Measu r ing Range: Defines the beginning and the end of the r ange.
When the Low Scale Limit is configu r ed with a value highe r than the
Full Scale Limit, the cu rr ent output will have a dec r escent behavio r
( 20~4 mA ) .
The values configu r ed in these fields can not be beyond the senso r
measu r ing r ange. The minimum span value has to be obse r ved as well
( see Table 01 ) .
1 / 3
Need help?
Do you have a question about the TxIsoRail and is the answer not in the manual?
Questions and answers