BG Test Module Precautions
• This module should only be used in vitro and can only be used with MicroTech's
Exactive EQ blood glucose test strips. Use of other brand test strips will result in
incorrect test results.
• The blood glucose test function can only be used to determine blood glucose
levels using whole blood samples. Do not use serum or plasma samples.
• The blood glucose test function is not intended for neonatal applications.
• Test results may not be accurate for blood hematocrit of more than 55% or be
low 30%.
• Blood containing high levels of vitamin C or other reducing agents can lead to
inaccurate results.
• The blood glucose meter test range is 1.1-33.3mmol/L (20-600 mg/dL).
• Triglycerides above 3,000 mg/dL and cholesterol above 500 mg/dL will lead to
inaccurate test results.
• Patients with serious illness (such as severe dehydration or ketoacidosis) are
not suitable for measuring blood sugar using the blood sugar test system.
• The blood glucose meter is only suitable for clinical screening tests or family
self-monitoring. The test results cannot be referred to as confirmed cases. In
order to ensure the accuracy of the results, the test results can be further con
5