Use the device for intended clinical purposes. Before each use, Ensure that the AA3 SUPER DELUXE is not damaged and is working properly. No modification of this equipment is allowed.
Fault Diagnosis We hope that this manual meets your needs. The Research & Development department at APPASAMY ASSOCIATES would appreciate any comments you can share with us. Kindly feel free to send your feedbacks, suggestions and clarifications to our e-mail: APPASAMY ASSOCIATES 20, SBI Officer’s Colony, 1st street,...
Number following this symbol indicates the serial number of the Equipment Character following the symbol indicates the Model number of the Equipment Consult User’s Manual Caution, refer to accompanying documents Manufactured By Temperature Limitations Keep away from Sunlight AA3 SUPER DELUXE APPASAMY ASSOCIATES...
Cautions are intended to alert you to importance of following the correct operating procedures where risk of injury to the patient or system. CONTRAINDICATIONS There are no specific contraindications for the use of this device AA3 SUPER DELUXE APPASAMY ASSOCIATES...
Operating Temperature: 21°C ~ 25°C Storage, Usage Period and Transportation Protect the unit against heat when in storage and transportation temperature should be 10°C ~ 45°C Do not expose to temperature above 50°C / 122°F AA3 SUPER DELUXE APPASAMY ASSOCIATES...
1. Cryoprobe lead connector 2. ON/OFF Indicator 3. Pressure Gauge 4. Footswitch Connector 5. Gas Connector FOOTSWITCH DESCRIPTION AA3 SUPER DELUXE Foot switch on fully Press Position PRESS DOWNWARD Foot switch on Release/ Ideal Position AA3 SUPER DELUXE APPASAMY ASSOCIATES...
Unpack the Equipment and save all packing materials. They are specially designed to protect the Equipment and will make repacking easy if you ever need to ship your AA3 SUPER DELUXE After unpacking verifies the Packing list we indicate below.
4. Cryoprobe (Cataract probe/ Retinal probe/ Glaucoma probe/ Intravitreal Probe) 5. Accessories (‘O’Rings-5 no’s and Metal washer-2 No’s) And User Manual AA3 SUPER DELUXE is marked with Model No, Serial No and Trade Mark. INSTALLATION The initial installation will be done by Appasamy Associates Trained Engineers only.
Page 13
(Fig-2) Step-2 Clamp the cylinder connection tubing (yoke end) to the cylinder (Fig-3) and the other end to the Union seen at the back panel gas connecter inlet (Fig-4). (Fig-3) (Fig-4) AA3 SUPER DELUXE APPASAMY ASSOCIATES...
Page 14
Open the cylinder value fully. Observe the pressure gauge readings. The reading should read between 600 p.s.i. to 900 p.s.i. in nitrous oxide gas cylinder and 750 p.s.i. to 1000 p.s.i. with cabon dioxide. (GREEN MARK IN PRESSURE GAUGE) AA3 SUPER DELUXE APPASAMY ASSOCIATES...
Page 15
Vice-versa for 3 to 4 times until the pressure gauge reads zero. This is for removing the gas inside the unit. (Fig-9) Wait for five minutes before disconnecting the gas connector from the cylinder. AA3 SUPER DELUXE APPASAMY ASSOCIATES...
Cylinder is too full open the cylinder for a few seconds before connecting the cryo unit Change the ‘O’ ring if ‘O’ rings are defective leakage persists inform to Leakage from the probe the manufacturer AA3 SUPER DELUXE APPASAMY ASSOCIATES...
WARRANTY APPASAMY ASSOCIATES Warranty for the instrument for a total period of two complete years from the date of delivery. Our obligation under the warranty is limited to replace any part which shall upon examination at our factory be determined to have been defective.