Operation Overview

During operation, air is drawn in at the top, is measured, and exits at the bottom. Do not block the
air intake and exit openings. When operating, the display shows the ion count present in the air,
updating twice per second. The 10-foot long ground wire should be used to connect the meter to
earth ground for most types of measurements, especially if you are measuring high ion levels.
At power on or when changing polarity, the meter stabilizes for about four seconds, after which
ion measurements begin. For accurate readings, avoid measuring near charged plastic such as
synthetic fabric (clothing). Excess charges will reduce both positive and negative ion counts in
the vicinity.
In a typical environment, ions do not mix well. There can be clusters of negative ions in one area
with almost no ions just a few inches away. Thus, it is normal for the ion level to fluctuate
irregularly as air is drawn into the meter. To provide a more stable reading, the meter imposes
five second averaging. An alternative average of 2 or 14 seconds can be selected.
Ion peak levels are continually saved in the meter. By pressing the "Peak Mode," the saved
maximum ion level that occurred (since the last time "Reset Peak" was depressed) is displayed.
"Record", initializes automated recording for later retrieval via the USB port. The AlphaApp PC
application provides for retrieval and management of the meter data. The application may be
https://www.alphalabinc.com/alphaapp/
downloaded at:
"Scale" switches the meter's range between 10 to 2 million ion cm
3
and 100 to 200 million
ions cm
.
3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AIC2 and is the answer not in the manual?

Questions and answers

Bill Kunde
May 10, 2025

Error message High Offset Clean Chamber, who do I clean the chamber?

1 comments:
Mr. Anderson
May 20, 2025

To clean the chamber for the AlphaLab AIC2 when receiving a High Offset Clean Chamber error message, blow compressed air through the air intake. If the issue is due to moisture condensation, allow the meter to dry out.

This answer is automatically generated

Summary of Contents for AlphaLab AIC2

  • Page 1 Air Ion Counter AIC2 2.1mm Power Adapter Jack USB Port Earth Ground Jack Operation Overview During operation, air is drawn in at the top, is measured, and exits at the bottom. Do not block the air intake and exit openings. When operating, the display shows the ion count present in the air, updating twice per second.
  • Page 2 “Polarity.” Note that when the polarity is changed, the meter re-stabilizes. “Scale” selects the alternative 200 million range on the dual scale model. On the standard (single range) AIC2 the scale is fixed at 2 million. The 200 million scale Selected Polarity...
  • Page 3 Peak Mode Pressing "Peak Mode" displays the real time ion levels (top displayed value) as well as the highest ion level measured by the meter since the last time the peak was reset. While in this mode, momentarily pressing "Peak Reset”...
  • Page 4 Power Options and Battery Level Operation time with the installed 4 AA alkaline batteries is typically 16 hours. To access the batteries, remove the two screws on the battery cover at the back of the meter. The meter may be operated with a 6 or 9VDC, 200mA regulated or unregulated power adapter with a 2.1mm barrel plug, positive center.
  • Page 5 Selectable Ion measurement/record intervals; 0.5, 1.0, 2.5, 5, 10, 15, 30, or 120 seconds  96 hours of nonvolatile storage in the AIC2 at the 1 second measurement interval  Ion level verses time graphs with optional selectable automatic scaling and time compression ...
  • Page 6 SPECIFICATIONS Ion Polarity Positive or Negative Ions - selectable 2 Million or 200 Million Ions cm - selectable Ion Range Ion Resolution 10 Ions/c (2M Scale), 100 Ions/cm (200M Scale) Ion Accuracy +/- 20%, repeatability 5% Averaging 5 or 14 seconds - selectable Peak Value Continually saved Peak value display - selectable...
  • Page 7 10.6 x 11.8 x 2.7 inches (27 x 30 x 6.8cm) The warranty period is one year from date of delivery. Warranty: Designed and manufactured in the USA by AlphaLab Inc., 3005 South 300 West, Salt Lake City, UT 84115 www.alphalabinc.com Tel (801)487-9492 Fax (801)487-3877 Manual V2.6 October 2020...

Table of Contents