This also applies if only excerpts of the document are copied or circulated. These same conditions apply to the disclosure of the document in digital format. As of: July - 2018 HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com...
Any other use of the machine or any use beyond its intended purpose is considered non-intended and thus improper. In this case, the device safety and the device protection provided may be compromised. HERMOS AG is not liable for damages resulting from such use.
Reduce the costs of production downtime 4.2 Warranty and liability The „General Terms and Conditions of Sale and Delivery“ of HERMOS AG shall apply. The warranty period is 24 months beginning with the delivery of the device, which is verified by the invoice or other documents.
Risk of injury due to disregarding the safety symbols. Risks exist when disregarding warnings in the operating instructions. Follow all warnings. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com...
(operation, installation)! Operating personnel must be required to immediately report any occurring and identifiable safety deficiencies to their superior! HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
Disconnect the power supply prior to removing or adding components! Observe the basic principles of ESD protection Take the appropriate ESD precautionary measures HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
Other applicable regulations 5.5.1 Regulations and certifications The electrical design and documentation satisfy the DIN / VDE, EN / IEC regulations. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com...
The reading device also continues carrying out protocol messages in polling mode. This may, however, result in delays in the poll rhythm. The polling functionality is optional customer-specific available! 6.2.4 Test module HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
If the read or write process is successful, the yellow LED remains permanently lit. With the existing I / O module, HERMOS LF readers with ASC-W1 protocol also show the success- ful or incorrect reading of the transponder at the respective first output port. See also behavior of the output ports in test mode.
The communication with the device can be carried out via the serial interface (9-pin Sub D) RS232 interface HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
6.4.1 Power supply and current input Description Min. Max. Unit Voltage (reverse polarity protected) V (DC) Current consumption (Read/Write Pulse manner (50ms) Current (passive) HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com...
Do not use the device in areas where it is exposed to vibrations or shocks. The installation location must be adequately illuminated during the installation. Never install the unit during a lightning storm. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
The packaging material consists of cardboard and foil. Dispose of these materials separately under the respective regulations of your country. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
Only mount the components at the designated locations and make sure that the operating and ambient conditions specified in the technical data are always maintained. Dimensions HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
Transponder vertical to the antenna axis: Transponder is outside the antenna range Transponder is within the antenna range HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 E-mail: rfid@hermos.com www.hermos.com...
Signal +24V DC If the device is connected to the power supply, the power LED lights up. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
DIP 3 on: Activated test mode Selection of the test function: off = Read, on = Write HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
The serial interface is implemented as a Sub-D plug connector (9-pin). A serial connection line (1:1 circuit) can be used. Signal (Optional GND) (Optional +5V) HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
ASCII protocol, then please contact the manufacturer HERMOS. Signal Baudrate 19200 Datenbits Stoppbits ASCII: Even Parität SECS/HSMS: None HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
TCP / IP port 3241 in the delivery state. The HERMOS „DeviceDiscoverer“ is available for configuring the network setting. 8.3.1 Change networksettings with DeviceDiscoverer HERMOS components can be found in the LAN network and settings can be easily changed using the „DeviceDiscoverer“. HERMOS AG LFM 4x Reader Rev1.4...
Page 33
WE GENERATE SOLUTIONS Select your network interface if you have several options on your PC. Your network is automatically scanned for all HERMOS reading devices using the „Search Devices“ button. Select the desired reading device in the list to open the network settings.
8.3.2 Firmware update with DeviceDiscoverer Firmware updates can also be performed using the HERMOS „DeviceDiscoverer“. Start the tool with administrator rights and scan the network for all HERMOS devices. To do an update, mark the desired reading device and select „Firware Update”.
WE GENERATE SOLUTIONS 8.3.3 Test Device with DeviceDiscoverer This feature is not yet supported by all HERMOS readers and is only available for the LFM4x as of software version V1.5. Readers can also be tested with the HERMOS "Device Discoverer" if you have one support feature.
Page 36
A success code of 0x00 in the response message indicates the correct execution of the UDP action commands, otherwise an error has occurred. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 37
Via a scan, the first page of every readable 134.2 kHz LF transponder is read out and displayed. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
If the reader receives undefined or random characters, this may result in an accidental protocol change. Automatic protocol detection can be deactivated in parameter 98 (0x62). HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
WE GENERATE SOLUTIONS 9.3 Addressing the antenna port The reader LFM 4x is equipped with 4 antenna ports. Each of these ports has its own antenna tuning with which the connected antenna is tuned to the ambient conditions. In the set protocol, the read or write commands are always executed by port. The 4 antenna ports from 1 to 4 are numbered consecutively and are addressed accordingly in the logging command.
WE GENERATE SOLUTIONS 9.4 Customer Factoryparametersets The reader LFM 4x can be delivered with a few Factoryparametersets. The settings of the Factoryparameterset is part of the factory settings and can only be changed by Hermos. If customer needs another Factoryparameterset setting, please ask Hermos for the necessary steps.
Second start character (ASCII-character „X“ = advanced ASCII protocol) Start 2 High byte of the message length (ASCII-character „0“-„F“) Length 1 HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Checksum 4 Low byte addition of all data (packet header, data and end character). (ASCII character „0“…“F“) HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
0x31 0x32 0x33 0x34 0x35 0x36 0x37 0x38 „3132333435363738“ Data in message (16 ASCII-characters) HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 44
ASCII characters (2 bytes) in hexadecimal format („00“ to „FF“). “2D“(hexadecimal) Example: Value 45 (decimal) HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 45
1 Byte The reader ID defines the address of the antenna port. The reader LFM 4x can be read at all 4 antenna ports (1-4). The antenna port is selected via the reader ID and is continuous. In ASC-W1 -mode the 4 antenna ports with the reader ID „0“ to „3“ are addressed. General messages also possible at address F (HF).
10.4 Protocol Commands Read: Command Description Read Data Automatic Read Write: Command Description Write Data HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 47
RS232 interface. The commands specified for this device are not part of this documentation. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 48
Read/Only transponder „F1“ Read/Write transponder „00“ Read the first page of all transponder types HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
The write process is started only when the sensor is busy, otherwise the error message ‚no tag (4)‘ is sent. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
The values returned for the data item parameter value in the response are decimal values (00-99). The values for the data item Parameter No. are hexadecimal values (0-F). HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
In the case of two-digit hexadecimal parameter numbers (Parameter No.), the parameter number and the parameter value double-digit hexadecimal are also used in the response. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
For two-digit hexadecimal parameter numbers (Parameter No.), the parameter value must always be sent in two digits hexadecimal. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
If TCP / IP is used as the interface, the host will not receive an acknowledgment, because an existing TCP/IP connection is interrupted by the reset. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
More about error codes and the corresponding corrective measures can be found in the chapter Error Codes. Host Device Reader-ID 1 Byte HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
The response code is part of the protocol but is not used for this device. The response code is always ‚0000‘. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
The 8 characters of the software version are represented by 16 ASCII characters. Each character is described in hex format and transmitted by 2 ASCII characters. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
‚no tag (4)‘ is sent. Locking a page cannot be reversed. This page is permanently write protected. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
If the reader can not determine the appropriate calibration, the error „5 - Invalid“ is sent instead of the confirmation. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
For optimum read and write ranges, the value is automatically determined by the reader (I-message). The value is measured in 16 steps (0-F). HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
In parameter 1 „Read Mode“ the sensor can be deactivated. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
B is present. In parameter 1 „Read Mode“ the sensor can be deactivated. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
>> O11310A << 011 The number of available outputs depends on the reader version used. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
6, depending on the Factoryparameterset.) sensor states instead of the 2 output states. Example: Status of all 4 sensors >> Q10 << q100000 HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
01 .. 99 (0,1 s) Default: 05 … (0,5 seconds) time Realisation: a separate parameter for each port HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 66
0 - deactivated 1 – activated Default: 1 Realisation: a separate parameter for each port HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 67
1 … read process when sensor is uncovered Default: 0 Realisation: a separate parameter for each port HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Locked Tag is write protected be written exchange the tag for the new one HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 71
RS232 or Ethernet connection Timeout) Invalid Unknown command was Invalid command Check message syntax received HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
SECS message blocks always have a specified structure that consists of 1-4 length bytes, 10 bytes of message headers and message data. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 80
See example with Reader ID 0x01 and Gateway ID 0x00: Upper Devic-ID (Reader-ID) R-Bit 0 HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 81
For more detailed information about the structure and transmission procedure, see SEMI E4, E5, E37, E99. (SEMI Equipment Communication Standard Message Transfer SECS) HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Only the occurrence of an error is reported. Errors are usually not reset. Bit 8 = 1 Alarm activated Where used S5F1 HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 83
‘:’ a semicolon separates the alarm text from the sensor states ALTX[3] Where used S5F1 HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 84
Sensor type for presence sensor „ECID_35“ Parameter 35 Special features „ECID_36“ Parameter 36 DIP switch activation HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 85
(max. 15 bytes) „HARDWARE“ String of the hardware release (max. 10 bytes) HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 86
WE GENERATE SOLUTIONS „SELF_TEST_RESULT“ Returns the result of the last self-test. A self-test can be triggered by message S18F13 using SSCMD. „MANUFACTURER“ String from the manufacturer „HERMOS“ „ECID_00“ bis „ECID_99“ see data element ECV parameters 0 - 99 Head attribute definitions: „HeadStatus“...
Page 87
Parameter number of the reader. The values are displayed as decimal values. Where used S2F13, S2F15 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 88
The MID corresponds to the DATA (fixed) Where used S18F10, S18F11 Please note the parameters 42-45 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 89
PAGEDATA [1-8] The 8 bytes (one page) of the transponder ID follow. Where used S3F7, S3F13 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 90
Head of the stored SECS message block. Only the last message is saved. This must be confirmed by the host. Where used S9F9 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 91
… Set one of the device LEDs. CPVAL‘s 1 <LEDStatus> 2 <Timeout> 3 <LEDNo> Where used S18F13 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 92
The use of the 4-digit TARGETID is currently not supported. Where used S18F1, S18F3, S18F4, S18F7, S18F9, S18F11, S18F13 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
(Host Reader) • Write MID S18F11 • (Host Reader) Subsystem command S18F13 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 94
The reading device can only be set to „online“ again using the message S1F17 (or reset S2F19); all other messages are cancelled by message SxF0. S1F15 W (header only, no additional elements) HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 95
The reading device transmits the requested attribute (parameter) to the host. S2F14 <L[1] <U1[1] ECV> > HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 96
This message is only displayed if a software reset (RIC=2) has been triggered. S2F20 <B[1] RAC> HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 97
S3F8: Material lost confirmation (MID LOST, ACK) (host -> reading device) The host confirms the message material lost. S3F8 <B[1] ACKC3 > HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 98
S3F14: Material found confirmation (MID FOUND, ACK) (host -> reading device) The host confirms the received MID data. S3F14 <B[1] MIDAC > HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 99
The reading device does not recognise the function number in the header of the message block. S9F5 < B[10] MHEAD > HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 100
S9F9 < B[10] SHEAD > HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
If the ATTRID of the S18F1 message is not known, the corresponding element ATTRVAL receives the value <nothing>. HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 102
TARGETID or acknowledge the result of the request. S18F6 <L,3 <TARGETID> <SSACK> <DATA> > HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 103
TARGETID was successful or failed. S18F8 <L,3 <TARGETID> <SSACK> < L,1 ...<STATUSLISTE> > > HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 104
The reading device can be in maintenance mode (MT) or operating mode (OP) to read the MID with the message S18F9. HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 105
The reading device can be in maintenance mode (MT) to write the MID with the message S18F11. HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 106
This message reports the result of the requested procedure. S18F14 ,W <L,3 <TARGETID> <SSACK> < L,1 <STATUSLISTE> > > HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
If no parameter arrays should be used, please ask the manufacturer (parameter 63)! **): Since software revision V1.5, the reader is always started with the default baud rate 19200Bd because of the protocol recognition. Please ask manufacturer HERMOS if other baud rate is required! Parameter name Description (HEX) (DEZ) The gateway ID is part of the device ID.
Page 108
241: read read-write transponder Default: 0 (read all transponders) Realisation: a separate parameter for each port *) HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 109
. 0 … synchronization deactivated 1 … synchronization activated 1 … activated Default: HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 110
LED2 will be influenced at ASC- Bit 7: W1 mode 0x01 … (0000 0001) Default: HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 111
MIDArea (maximum MID range) and the Carrie- 0x3B CarrierIDLength rIDLength. The default value is customer specific! Default: 0 or 16 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 112
Serial number string The read-only parameter provides the complete serial 0x47 (read only) number string. HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 113
Antenna tuning of the 0x54 addressed antenna Default: 08 Realisation: a separate parameter for each port HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 114
Par. 43 = 0 Par. 44 = 1 Par. 45 = 0 Default: 0 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
S1F1 – Query Software version S2F13 – Query parameter 3 S2F15 – Set parameter 3 to value 5 HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Page 116
S3F7 - Material lost S18F5 - Read data Antenna port: 1 Page: Data length: HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 117
S18F7 - Write data Antenna port: 1 Page: Data length: „ABCDEFGH Data: “ S18F9 – Read MID HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Page 118
(Maintenance) The CPVAL „OP“ can be used to switch back to the normal operating mode. HERMOS AG LFM 4x Reader Rev1.3 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Check parameter syntax and format or wrong plemented or out of range value value HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Never short circuit the fuse! This may result in fire or damages on the device. Only use fuses specified by the manufacturer. HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...
Check the power LED is lit, and make sure that the status indicator does not indicate an error. Try to read in the reader with the HERMOS Device-Discoverer and check the device settings. If the above measures do not correct the error, please contact the manufacturer.
17.01.2019 Protocol LFMS15 (SECS/HSMS) - Storing of FatalErrors in the EEPROM and their retrieval HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com www.hermos.com...
Housing and attachments as plastic waste Electronic components, antennas and cables as electronic waste HERMOS AG LFM 4x Reader Rev1.4 Gartenstraße 19 • 95490 Mistelgau Phone: +49 9279 991 - 0 back to the table of contents E-mail: rfid@hermos.com...