    DOscan10 Overview Thank you for selecting the DOscan10 dissolved oxygen tester. This manual provides a step-by-step guide to help you operate the instrument, please carefully read the following instructions before use. Installing the Batteries Twist the connector collar counter clockwise, pull the 6-pin connector away from the tester. Insert the two AAA batteries into the battery compartment, note polarity.
  • Page 3: Setup Menu

    DOscan10 Filling the Electrolyte Solution Unscrew the membrane cap from the buttom of the electrode, rinse the inside and outside with deionized water and blot dry. Fill the membrane cap halfway with electrolyte solution. Screw the membrane cap back onto the electrode. Some electrolyte solution will overflow during this process. Check the electrode, make sure that no air bubbles are trapped in the electrolyte solution and the membrane is not creased or damaged.
  • Page 4 DOscan10 Setting the Default Option In the measurement mode, press and hold the Cal key for 5 seconds to enter the setup menu. If necessary, press the Cal key again to select an option. Press the Enter key, the tester saves the current option and moves to the next menu item. Repeat the steps above until the tester returns to the measurement mode.
  • Page 5: Dissolved Oxygen Calibration

    DOscan10 Dissolved Oxygen Calibration The DOscan10 tester allows 1 or 2 points calibration. If you have selected the single point calibration in the setup menu, we recommend that you perform a 100% saturation calibration in the air-saturated water. If the 2 points calibration is selected, the zero oxygen solution needs to be used. During the calibration and measurement, the temperature sensor on electrode must be immersed in solution completely, the solution keeps 0.3 m/s of minimum flow rate to avoid oxygen starvation at the membrane.
  • Page 6: Temperature Calibration

    DOscan10 Temperature Calibration Press and hold the Cal key to enter the setup menu. Press the Enter key until the display shows Press the Cal key, the tester enters the temperature calibration mode. Place the electrode into a solution with a known accurate temperature and wait for measurement is stable. Press the Cal key to modify the temperature value.
  • Page 7: Optional Accessories

    DOscan10 Appendix Preparation of Zero Oxygen Solution Dissolve 500 mg of sodium sulfate (Na ) reagent and a small amount of cobalt (II) chloride hexahydrate (CoCl O) in the 250 ml deionized • water, mix the solution until reagent is completely dissolved. Preparation of Air-Saturated Water Use an air-pump to blow air into deionized water at least 1 hour, while stirring the solution.
  • Page 8: Warranty

    DOscan10 Disposal This tester is required to comply with the European Union’s Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC and may not be disposed of in domestic waste. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.