Summary of Contents for SENSIRION Winnipeg Ventilator

  • Page 1: Table Of Contents

    Quick Reference Guide Contents Warnings Circuit Setup Pre-Use Pressure Leak Test Ideal Body Weight (MALE) Ideal Body Weight (FEMALE) Inspiratory Time %O2 Rise Time Alarm Troubleshoo�ng Full IFU: www.CEVen�lators.ca Call Toll Free: 1-833-229-1036 52160-B...
  • Page 2 Warnings Warnings All flows, volumes and pa�ent data parameters are Inspiratory, NOT Expiratory The set pressure is Peak Inspiratory Pressure (PC or PSV). ΔP = Set Pressure - PEEP Example Se�ngs Results 15 cmH Pressure 15 cmH Peak Inspiratory Pressure 5 cmH 5 cmH Manual PEEP Valve...
  • Page 3: Circuit Setup

    Circuit Setup Expiratory Line Port Inspiratory Port (with internal airway pressure tap) Airway Pressure Port Oxygen Source Oxygen Input (50 psi / 345kPa) Pa�ent Interface Pa�ent HME With Filter 2 X 72 in (183 cm) Hose Airway Pressure Tap Exhala�on Valve External PEEP Valve General Informa�on To TURN ON the ven�lator, push the power bu�on on the...
  • Page 4: Pre-Use Pressure Leak Test

    Pre-Use Pressure Leak Test Instructions Silence Alarms Select Standby Mode Block (occlude) the pa�ent interface connec�on Press and hold the Standby and Inspiratory Hold bu�ons simultaneously for 2 seconds Verify that the Pressure graph reaches 50 cmH See the Insp. Tidal Volume display for test status and result. Insp.
  • Page 5: Ideal Body Weight Male

    Ideal Body Weight Male Measure the (MALE) Pa�ent Height Determine Tidal Volume using IBW Chart MALE – IDEAL BODY WEIGHT AND TIDAL VOLUME CHART mL/kg TIDAL VOLUME ( ) AT: PATIENT HEIGHT FT’ IN” IN” 4' 0" 22.4 4' 1" 24.7 GREY AREA INDICATES TARGET VOLUMES 4' 2"...
  • Page 6: Ideal Body Weight Female

    Ideal Body Weight Female Measure the (FEMALE) Pa�ent Height Determine Tidal Volume using IBW Chart FEMALE – IDEAL BODY WEIGHT AND TIDAL VOLUME CHART mL/kg TIDAL VOLUME ( ) AT: PATIENT HEIGHT FT’ IN” IN” 4' 0" 17.9 4' 1" 20.2 GREY AREA INDICATES TARGET VOLUMES 4' 2"...
  • Page 7: Inspiratory Time

    Inspiratory Time Tidal Volume (mL) Page 7 of 14...
  • Page 8: I:e

    Inspiratory Time (Seconds) Page 8 of 14...
  • Page 9: O2 Rise Time

    %O2 Rise Time at Pa�ent Connec�on Page 9 of 14...
  • Page 10: Alarm Troubleshoo�Ng

    Alarm Troubleshoo�ng When you see Do This... this alarm... • Assess pa�ent status • Inspect the expiratory limb for defec�ve or damaged components that may be occluding the circuit • Press the alarm reset bu�on • If alarm is not resolved remove pa�ent from breathing circuit and apply manual breaths with a manual resuscitator.
  • Page 11 Alarm Troubleshoo�ng When you see Do This... this alarm... • Remove pa�ent from breathing circuit and apply manual breaths with a manual resuscitator. Discon�nue the ven�lator and send for service. • Obtain alterna�ve ven�la�on, and remove ven�lator from use and request service •...
  • Page 12 Alarm Troubleshoo�ng When you see Do This... this alarm... • Assess the pa�ent for: • Increased or thicker mucus or other secre�ons blocking the airway • Coughing, gagging or figh�ng the ven�lator’s mandatory breath delivery • A bronchospasm • Assess the pa�ent for dysynchrony: •...
  • Page 13 Alarm Troubleshoo�ng When you see Do This... this alarm... • Assess pa�ent status • Check the pa�ent circuit (including filters) and the endotracheal tube • Assess the low PEEP alarm se�ng • Assess the high PEEP alarm se�ng • Check the external PEEP valve se�ng, check the valve for malfunc�ons •...
  • Page 14: Call Toll Free: 1-833-229-1036

    Alarm Troubleshoo�ng When you see Do This... this alarm... • Assess pa�ent status • Re-assess ven�la�on mode according to clinical assessment and clinical treatment plan •Assess the trigger sensi�vity se�ng •If proceeding with PSV, select PSV mode and manually reset alarm by selec�ng the alarm Reset bu�on •...