VIDEO-CYSTOSCOPE M o d e l : D R 0 6 0 3 8 0 B - 2 USER MANUAL MANUAL P/N:IA-MN-213-3(Version 3.1) Hangzhou Hawk Optical Electronic Instruments Co., Ltd.
Page 2
Operate the device according to the specific procedures. The engineer at Hawk is always prepared to assist in answering any of your questions or in meeting your specific requirements. We also welcome any comments you may have concerning about our equipment.
Page 3
Conformance Standards 1.The following classifications are in accordance with the IEC/EN 60601-1:6.8.1: 1) According to 93/42/EEC Medical Device Directive, this is Class IIa medical Device. 2) According to IEC/EN 60601-1. • Equipment is Class I, BF Part. 3) According to IEC 60529:2001-02. The protection level of monitor is IPX0.
1.Safety Introduction Safety is everyone’s responsibility. The safe use of this equipment is largely depend upon the installers, users, operators, and maintainers. It is imperative that personnel study and familiar with this entire manual before attempting to install, use, clean, service or adjust this equipment and affiliated accessories.
Page 6
Such tampering will make customer forfeit any rights to claim under warranty. ● This equipment may only be used together with accessories supplied by HAWK. If the customer makes use of other brand accessories, use them only if their safe usability under technical safety aspects has been proved and confirmed by HAWK or the manufacturer of the accessory.
Page 7
● Immediately turn off and unplug any equipment that gives off smoke, sparks, strange noises, or odors . 1.2.2 Environment requirements 1.2.2.1 Avoid the following environments for operation or storage: Where the device is exposed to water vapor. Don’t operate a device with a wet hand. Where the device is exposed to direct sunlight.
The safety using of videoscope not only depends on videoscope itself, but also depends on other devices and accessories which use with this videoscope. To ensure compatibility with these devices and accessories, we recommend that the company uses only HAWK videoscope and accessories.
1.10 Label and Symbol description The Label Symbol Description is listed as below. Table2:Label and Symbol Label Icon Meaning ~ AC Power Power supply Turn on (Main power) Turn off(Main Power) Type BF Applied Part “Note” “Warning” Fuse Manufacture Date Manufacturer and Address European Union representative and Address Do not throw into wastebin...
1.11.2 Declaration of conformity The device is made by Hawk company meet the requirements of EN60601-1-2: 2007. Table 3:Guidance and manufacturers declaration -- Electromagnetic Emissions The device is intended for use in the eietromagnetic environment specified below. The customer or the user of the Irrigation Aspiration Unit should assure that it is used in such an environment.
Page 11
humidity should be at least 30% Electrical fast ±2 kv for power ±2 kV for power Mains power quality should be that of transient/burst supply lines supply lines a typical commerdial or hospital lEC 61000-4-4 ±1 ±1 environment input/output lines input/output lines ±1 kV line (s)to ±1...
Page 12
Portable and mobile RF communications equipment should be used no closer to any part of Model 006,including cables,than recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d=1.2 Conducted RF d=1.2 80MHz~800MHz 3 Vrms 61000-4-6 3Vrms...
close or piled up to other equipment, it should observe and test the system can be run normally. ● Although the device has been designed to interfere with the protection of such electromagnetic radiation. When it connects with RF surgical set, you should restart the device as soon as electromagnetic interference occurs.
This equipment may be used for examination, diagnosis and operation. Do not use this equipment for any other purposes or in any other settings. 2.4 Feature 2.4.1 The videoscope has a unique waterproof function, it can be cleaned thoroughly and disinfected. The instrument channel can be cleaned in whole course, so that minimize the risk of infection among patients with each other.
18) Housing 19) AV output connector: connect AV output cable and output image to TV and monitor. 20) USD connector: transfer image to computer 21) SD Slot: insert SD card 22) Screen 23) Standby button: press this button for turn on the device 24) Power indicator: indicator on means switch on the device, indicator off means switch off the device 25) Decrease button: Decrease the parameter and up the menu, review the picture.
Page 18
and green indicator in the low right is light. This process takes a few seconds, please wait patiently. After power on, entering the preview mode to show the current image. If shows on the low left of screen, it can take photos and videos. Date and time will show on the top right, video time and capacity show on the low right, electricity capacity and show on the low left.
Page 19
then press photo/confirm button to enter into the file, press menu button to back. 3.5.7 Setup: After enter into sub-menu of Setup,There are 6 sub -menus: Video time, storage, Date time, Memory Info, LCD Brightness, Language, press button to select, then press photo/confirm button to confirm.
4)Memory Info: To record inside memory and SD Memory capacity. 5)Brightness LCD: Seven steps can be selected, use to decrease and increase the brightness. 6)Language: English and Chinese. 3.5.8 Review: In the preview mode, press menu and check photo files and Video file according to 5.i, Press the Menu button to enter the main menu in preview mode, press Photo/confirm button to confirm and play video, press Photo/confirm button or back/switch button to stop playing.
Page 21
Note When equip and remove 3.5” monitor, be careful to prevent the fall, fall damaged! a. Connecting LCD Screen to Video Endoscope: Aim at red dot of LCD Screen and Video Endoscope, connect them, then screw down lock nut and fix LCD Screen.
changes to white, and immediately release the button. 4.1.5 Please aim the videoscope at an object (such as your hand), if there is abnormal dot on the monitor screen, blurred images or striped images. If the image is not clear, check the cable connection, or wipe the objective lens with clean gauze moistened with alcohol.
Note ● In order to ensure smooth operation, after each operation, should first visual inspection equipment external mechanical damage and normal work can be, and as far as possible in the operation chamber spare a controller. ● CCD connector and head are vulnerable optical instruments, any curve and collision may cause the breakage of lenses or the deviation of optical axle to result in obscure image.
Page 24
● If the device in question is running, the power switch must be turned off immediately, unplug the cable, interrupt the operation. And readjust the device by the technician, if it still have problems, please contact HAWK service department for repair. Note Do not bend too hard, if the operator feel the resistance is too large, bending is not in place or cannot bend, the videoscope should be returned for repair.
5 Reprcessing 5.1 Reprcessing a) Please refer to HAWK's REPROCESSING MANUAL as attached- b) information if product need special treatment before reprocessing. 5.2 Checking 5.2.1 Leakage Test Connect the connector of leak detector to connector of endoscope, aiming red point of connector of leak detector to red point of connector of endoscope, and then press the connector down, the leak detector is fixed on the endoscope.
Then go through the following list before you call for service. If the problems still exit, turn off the device and move monitor from the videoscope, and contact local service technician or appointed HAWK service facility. 5.3.1 Monitor Phenomenon...
Resistance of bending wheel is normal If the device doesn’t work after you do above corrective actions, please hand this device to HAWK service and the device must conform to the following requirements. ●It should clean and sterilize the device before packing it.
Page 28
View of filed 110° Diameter of distal part Fr 17.3 Diameter of insertion part Φ5.7mm Instruments channel Φ2.2mm Working length 380mm Depth 3——50mm Bending angle Up 210° / down120° Temperature:+10℃ to +40℃ Running environment Humidity: 30 - 75% Pressure: 700 - 1060hPa. Temperature:...