Incubator 8000 IC / SC / NC
Electrical Safety Test in the USA and Canada according to
CAN/CSA - 22.2 No. 601.1 - M90
1
Visual inspection of basic unit
• Power cord
• Compare fuses to stated ratings on the backside of the Incubator next to the inlet of
the power cord.
2
Safety testing Incubator 8000 IC / SC / NC with DNI Nevada
Model µ-Test 2000
• Fold down flap below electronics module after undoing the two locks.
• Warning: These tests can expose personnel to hazardous electric shock and must
be carried out with caution.
• Note: Do not plug the µ-Test 2000 safety analyzer power cord into a line isolation
monitor as inaccurate readings may occur.
• Ensure that the switch "TEST LOAD" on the backside of the µ-Test 2000 is in the
"IEC 601-1" position.
• Plug the µ-Test 2000 power cord into a live AC receptacle, place the power switch
of the µ-Test 2000 to the "1" or ON position and ensure that the switches marked
"NEUTRAL" and "POLARITY" are in the OFF or CLOSED position.
2.1
Chassis Resistance Testing
• Rotate "MODE" knob to "OHM RESISTANCE" position.
• Attach test lead to "CHASSIS CABLE" input of the tester.
• Plug the Incubator power cord into the test receptacle of the µ-Test 2000.
Incubator 8000 IC/SC/NC
Electrical Safety Test with
DNI Nevada Model µ-Test 2000
Dräger Medical, Inc.
August 2000
Page 1 of 5