Dräger Medical 8000 IC Calibration & Testing
Dräger Medical 8000 IC Calibration & Testing

Dräger Medical 8000 IC Calibration & Testing

Advertisement

Incubator 8000 IC / SC / NC
Electrical Safety Test in the USA and Canada according to
CAN/CSA - 22.2 No. 601.1 - M90
1

Visual inspection of basic unit

• Power cord
• Compare fuses to stated ratings on the backside of the Incubator next to the inlet of
the power cord.
2
Safety testing Incubator 8000 IC / SC / NC with DNI Nevada
Model µ-Test 2000
• Fold down flap below electronics module after undoing the two locks.
• Warning: These tests can expose personnel to hazardous electric shock and must
be carried out with caution.
• Note: Do not plug the µ-Test 2000 safety analyzer power cord into a line isolation
monitor as inaccurate readings may occur.
• Ensure that the switch "TEST LOAD" on the backside of the µ-Test 2000 is in the
"IEC 601-1" position.
• Plug the µ-Test 2000 power cord into a live AC receptacle, place the power switch
of the µ-Test 2000 to the "1" or ON position and ensure that the switches marked
"NEUTRAL" and "POLARITY" are in the OFF or CLOSED position.
2.1

Chassis Resistance Testing

• Rotate "MODE" knob to "OHM RESISTANCE" position.
• Attach test lead to "CHASSIS CABLE" input of the tester.
• Plug the Incubator power cord into the test receptacle of the µ-Test 2000.
Incubator 8000 IC/SC/NC
Electrical Safety Test with
DNI Nevada Model µ-Test 2000
Dräger Medical, Inc.
August 2000
Page 1 of 5

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 8000 IC and is the answer not in the manual?

Questions and answers

Summary of Contents for Dräger Medical 8000 IC

  • Page 1 • Compare fuses to stated ratings on the backside of the Incubator next to the inlet of the power cord. Safety testing Incubator 8000 IC / SC / NC with DNI Nevada Model µ-Test 2000 • Fold down flap below electronics module after undoing the two locks.
  • Page 2 Chassis Resistance 0.2 Ohm • Afterwards leave alligator clip at test item 4) or attach the alligator clip to the test item with the lowest resistance. Incubator 8000 IC/SC/NC Electrical Safety Test with August 2000 Page 2 of 5 DNI Nevada Model µ-Test 2000...
  • Page 3 • Remove the red lead with the alligator clip from the Incubator. • Leave all other selections from the previous test the same. • The Incubator is still switched on. Incubator 8000 IC/SC/NC Electrical Safety Test with August 2000 Page 3 of 5...
  • Page 4 Skin Temperature Sensor Simulator 79 01 236 with the switch in "R " position. Short the two yellow "R outlets from the simulator and connect EXT" Incubator 8000 IC/SC/NC Electrical Safety Test with August 2000 Page 4 of 5 DNI Nevada Model µ-Test 2000...
  • Page 5 (According to CAN/CSA - 22.2 No. 601.1 - M90, Sub-clause 19.3, Table IV) • Switch off Incubator and disconnect power cord from the µ-Test 2000 and remove the test equipment Incubator 8000 IC/SC/NC Electrical Safety Test with August 2000 Page 5 of 5...

This manual is also suitable for:

8000 sc8000 nc

Table of Contents