NEA_LEC_EN_ind01_06/01/2021 Nealact Lactate meter User Manual...
Page 2
Dear Nealact User: Thank you for purchasing the Nealact Lactate Meter Monitoring System. This manual provides important information to help you to use the system properly. Before using this product, please read the following contents thoroughly and carefully. If you have other questions regarding this product, please contact the local customer service or place of purchase.
Page 3
IMPORTANT SAFETY PRECAUTIONS READ BEFORE USE Use this device ONLY for the intended use described in this manual. Do NOT use accessories which are not specified by the manufacturer. Do NOT use the device if it is not working properly or if it is damaged. The lactate test strip has not been tested in NICU (Neonatal Intensive Care Unit) This device does NOT serve as a cure for any symptoms or diseases.
BEFORE YOU BEGIN ....................1 Important Information ..............1 Meter Overview ................2 Display Screen ................3 Nealact Lactate Test Strip ............4 SETTING THE METER ..................... 5 THE MEASURING MODE ..................7 Lactate Testing ................7 QUALITY CONTROL TESTING ................8 When Should the Lactate Control Solution Test Be Performed? ..
BEFORE YOU BEGIN Important Information • Severe dehydration and excessive water loss may cause readings which are lower than actual values. If you believe you are suffering from severe dehydration, consult a healthcare professional immediately. • If your lactate is lower or higher than usual, and you do not have any symptoms of illness, first repeat the test.
Meter Overview Test Strip Ejector Eject the used strip by sliding up this button. Bluetooth Indication Light Display Screen Left Scroll Button ( ) Adjust value decrement. Scroll the values in the memory. M Button (M) Enter the meter memory. Right Scroll Button ( ) Enter Bluetooth function and adjust value increment.
Display Screen Test Result Measuring Mode Gen – Sample test Auto QC Mode QC – Control solution test Memory Mode Date / Time Error Message Alarm Symbol Measurement Unit Low Battery Symbol Lactate Symbol Test Strip Symbol Sample Drop Symbol Code of strip...
Test results might be wrong if the contact bar is not fully inserted into the test slot. NOTE: The Nealact meter should only be used with Nealact Lactate test strips. Using other test strips with this meter can produce inaccurate results.
SETTING THE METER Before using your meter for the first time or if you change the meter battery, you should check and update these settings. Entering the Setting Mode (a) Start with the meter off (no test strip inserted). Press and hold at the same time.
Page 11
6. Setting the reminder alarm Your meter has four reminder alarms. The meter will display “On” or “OFF” and “ ”. If you don’t want to set an alarm, press M to skip to select “On”, then press M. this step. Or press With the hour/minute flashing in sequence, press to select the correct hour/minute.
THE MEASURING MODE Lactate Testing The meter provides you with one mode for measuring, “Gen”. You can start with the meter switched off. Insert a test strip to turn on the meter. The screen will display “ ”, a flashing “ ”...
QUALITY CONTROL TESTING When Should the Lactate Control Solution Test Be Performed? ⚫ if it is mandatory following the local regulations in your country, ⚫ if you suspect the meter or test strips are not working properly, ⚫ if your test results are not consistent with how you feel, or if you think the results are not accurate, ⚫...
Page 14
NOTE: ⚫ For the control solution test, your device will tag this measurement as QC test automatically. ⚫ To avoid contaminating the control solution, do not directly apply control solution onto a strip. 3. Read and compare the result After counting down to 0, the control solution test result will appear on the display.
TESTING WITH BLOOD SAMPLE Warning: To reduce the chance of infection: ⚫ Never share a lancet or the lancing device. ⚫ Always use a new, sterile lancet. Lancets are for single use only. ⚫ Avoid getting hand lotion, oils, dirt, or debris in or on the lancets and the lancing device.
NOTE: Choose a different spot each time you test. Repeated punctures at the same spot may cause soreness and calluses. Performing a Blood Test To perform a blood test, you will need: (b), (c), (e) and (f). 1. Insert the test strip to turn on the meter Wait for the meter to display the “...
Page 17
⚫ The confirmation window should be filled with blood before the meter begins to count down. NEVER try to add more blood to the test strip after the drop of blood has moved away. Discard the used test strip and retest with a new one. ⚫...
METER MEMORY The meter stores the 1000 most recent test results along with respective dates and times in its memory. To enter the meter memory, start with the meter switched off. Reviewing Test Results 1. Press and release M. “ ”...
BLUETOOTH PAIRING Data Transmission via Bluetooth You can transmit your data from the meter to your device via Bluetooth. The ProCheck app is designed to assist you easily monitoring your lactate levels. How to Install and Update the ProCheck App You must connect to the internet to download the app.
Page 20
Bluetooth Indicator on the Meter: BLUETOOTH STATUS INDICATOR The Bluetooth function is on and Flashing Blue waiting for connection. The Bluetooth connection is Solid Blue established. WARNING: ⚫ While the meter is in transmission mode, it will be unable to perform a test.
MAINTENANCE Battery Your meter comes with two 1.5V AAA size alkaline batteries. Low Battery Signal The meter will display one of the messages below to alert you when the meter power is getting low. 1. The “ ” symbol appears along with display messages : The meter is functional and the result remains accurate, but it is time to change the batteries.
⚫ Properly dispose of the batteries according to your local environmental regulations. Caring for Your Meter Cleaning ⚫ To clean the meter exterior, wipe it with a cloth moistened with tap water or a mild cleaning agent, then dry the device with a soft dry cloth.
For further information, please refer to the test strip package insert. Important Control Solution Information Use only our Nealact lactate control solutions with your meter. ⚫ Do not use the control solution beyond the expiration date or 3 ⚫...
MEASUREMENT RESULT READINGS Lactate Testing MESSAGE WHAT IT MEANS < 0.3 mmol/L 0.3 to 22 mmol/L > 22 mmol/L Reference Values Lactate Test The meter provides you with plasma equivalent results and are displayed in millimoles of lactate per liter of blood (mmol/L). Desirable range: Lactate 0.3 to 2.4 mmol/L...
SYSTEM TROUBLESHOOTING If you follow the recommended action but the problem persists, please call your local customer service. Error Messages MESSAGE WHAT IT MEANS WHAT TO DO Appears when the Replace the battery immediately. batteries are too low. Appears when a used Repeat with a new test strip.
Troubleshooting 1. If the meter does not display a message after inserting a test strip: POSSIBLE CAUSE WHAT TO DO Batteries exhausted. Replace the batteries. Test strip inserted upside down Insert the test strip with contact or incompletely. bars end first and facing up. Defective meter or test strips.
Page 27
Expired or contaminated Check the expiry date of the control solution. control solution. Control solution that is too Control solution, meter, and test warm or too cold. strips should be at room temperature 20°C to 25°C (68°F to 77°F) before testing. Defective test strip.
SYMBOL INFORMATION SYMBOL REFERENT SYMBOL REFERENT In vitro diagnostic Manufacturer medical device Consult Catalogue number instructions for use Temperature limit Caution Use-by date Humidity limitation Batch code CE mark Serial number RoHS compliance Authorized representative in the European Community This device does not belong to household waste and must be returned to a collection point for recycling electric and electronic devices according to local laws.
SPECIFICATIONS Model No.: Nealact Lactate Meter Dimension & Weight: 102.5 (L) x 56.9 (W) x 21.8 (H) mm, 89 g Power Source: Two 1.5V AAA alkaline batteries Display: LCD with backlight Memory: 1000 measurement results with respective date and time...
PURCHASE INFORMATION Name Cat.number Content Nealact Lactate test strips: TD-4663 Vial of 25 strips Nealact Lactate control solution: TD-4924 1x tube 4mL Nealact Lactate meter: TD-4216 Meter + batteries...