Labnet Enduro™ Power Supplies Instruction Manual Catalog Numbers: High Amperage (250V) E0203 E0203-230V General Purpose (300V) E0303 E0303-230V This manual is available in additional languages at www.labnetlink.com.
Power Supply. To prevent accidental electric shocks to the user, always use gel box systems that have safety lids. Labnet recommends against the use of gel box systems and/or power leads that have unshielded banana plugs, thus minimizing any potential shock hazard to the user.
The warranty does not cover in-transit damage. To ensure safe, reliable operation, always operate the Labnet Enduro™ Power Supply in accordance with the manufacturer’s instructions. Always wear protective gloves and safety glasses when working in a laboratory environment.
(250V and 300V) or constant power applications and programming mode concurrently (only 250V). The Labnet Enduro 250V Power Supply is designed to run applications at maximal power and efficiency. This instrument is ideal for DNA/RNA electrophoresis, SDS-PAGE, native PAGE, Western Blotting applications.
1. Check the label located near the AC inlet to ensure that the unit is compatible with locally provided voltage. 2. Place the Labnet Enduro Power Supply on a level laboratory bench. Keep the area around the power supply clear to ensure proper ventilation of the unit.
NOTE section of 9.1. 1. Press the Power Switch on the rear side of the instrument to turn on the Labnet Enduro Power Supply. Upon start-up, the Display Screen on the front of the instrument will illuminate. The factory default settings (or last settings used) will be displayed.
AMP:3.00A 9.2 Basic Programming Protocol (250V only) A basic Programmable Mode operating procedure of the Labnet Enduro™ 250V Volt Power Supply is provided below. We recommend reading the guidelines provided in this manual for best results before starting an operation. The Programmable Mode allows you to vary levels in voltage (V), current (mA), and power (W) during specified periods of time as discrete changes (STEP) for up to 10 steps, depending upon your electrophoresis needs.
Discontinuous buffer systems and, to a lesser extent, continuous systems increase resistance during the run. If you use the current limiting setting on the Labnet Enduro Power Supplies, the voltage will increase as resistance increases to satisfy Ohm’s law (V = IR). If no voltage limit is set and a local fault condition occurs, such as a poor connection, very high local resistance may cause the voltage to increase to the maximum capacity of the power supply.
Power supply is turning the Power switch to the On position. alarm: Display shows overheating. “OVER TEMP” If you cannot restart the instrument, turn off the power, disconnect the power cord from the outlet, and contact Scientific Support. LABNET ENDURO POWER SUPPLIES...
12.0 Ohm’s Law Conversions Electrophoresis is the migration of a charged particle under the influence of an electrical field. The power supply output parameters voltage, current, and power are related by the following two equations: Voltage (V) = Current (I) x Resistance (R); (V = I x R) Power (P) = Current (I) x Voltage (V);...
For additional fuses contact Scientific Support. To replace the fuse: 1. Turn off the main power switch at the rear of the Labnet Enduro Power Supply and detach the power cord from the rear of the instrument. 2. Open the fuse compartment located inside the Power Entry Module by inserting a small flat blade screwdriver into the slot below the ON/OFF switch.
For your reference, make a note of the serial and model number, date of purchase, and supplier here. Serial No. __________________________________________ Date Purchased _______________________________________ Model No. _________________________________________ Supplier _____________________________________________ 16.0 Equipment Disposal According to Directive 2012/19/EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment (WEEE), this product is marked with the crossed-out wheeled bin and must not be disposed of with domestic waste.
Page 14
To request certificates, please contact us at www.labnetlink.com. Warranty/Disclaimer: Unless otherwise specified, all products are for research use only. Not intended for use in diagnostic or therapeutic procedures. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications. For additional product or technical information, visit www.corning.com/lifesciences or call 800.492.1110.