meter
SCAN Blood Glucose Meter
Model: G-427B
Distributed by CoachCare
User manual
97 3rd Avenue, New York, NY 10003 USA
Model Number: G-427B
Made in China

Advertisement

Table of Contents
loading

Summary of Contents for SCAN G-427B

  • Page 1 S C A N meter SCAN Blood Glucose Meter Model: G-427B Distributed by CoachCare User manual 97 3rd Avenue, New York, NY 10003 USA Model Number: G-427B Made in China...
  • Page 2 important safety instructionS read this before using  The meter and lancing device are for single patient use. DANGER - Misuse of this electrical device can cause electrocution, burns, fire and other HAZARDS.  Do not share them with anyone including other family member! Do not use on multiple patients! Basic safety precautions should always be taken, including ...
  • Page 3: Table Of Contents

    LIMITATIONS OF USE table of contents  The SCAN G-427B Blood Glucose Monitoring System is not intended for use on neonates. BEFORE YOU BEGIN __________________________________ 5  The SCAN G-427B Blood Glucose Monitoring System is INTRODUCTION TO THE SYSTEM ______________________ 6 not intended for use on artery blood, neonates’...
  • Page 4: Before You Begin

    1. Open the battery cover on the back of your meter. Intended Use 2. If replacing the batteries, please remove the used  The SCAN G-427B Blood Glucose Monitoring System is batteries. comprised of SCAN G-427B Blood Glucose Meter and 3.
  • Page 5: Test Principles

    (when glucose is not changing rapidly). the current produced from glucose oxidation. The strength  The G-427B blood glucose Test Strips are for use with the of the current produced by the reaction depends on the Blood Glucose Meter to quantitatively measure glucose in amount of glucose in the blood sample.
  • Page 6: Contents Of The Kit

    CONTENTS OF THE KIT BLOOD GLUCOSE METER  A meter  Test strips x 25 pcs  User manual  2 AAA alkaline batteries  Lancing device  Sterile lancets x 25 pcs NOTE:  The lancing device can be used several times however the sterile lancet should be changed after each use.
  • Page 7: Meter Screen Display Message

    METER SCREEN DISPLAY MESSAGE TEST STRIP Each strip can be used only once. The Test Strip consists of the following parts: Absorbent hole Apply a drop of blood here. The blood will be drawn in automatically. Confirmation window This is where you confirm if enough blood has been applied to the absorbent hole of the strip.
  • Page 8: Setting The Meter And Deleting The Memory

    SETTING THE METER AND DELETING THE MEMORY SETTING THE UNIT (MG/DL OR MMOL/L) 1. mg/dL → mmol/L With the “mg/dL” displaying, press “M” button until the SETTING THE DATE AND TIME “mmol/L” appears, press “S” button twice again until the 1.
  • Page 9: The Two Measuring Modes

    TESTING WITH BLOOD SAMPLE THE TWO MEASURING MODES The meter provides you with two modes for measuring, CAUTION General and CTL. The CTL mode should be used when a control solution is being tested. See the table below. To reduce the chances of infection: ...
  • Page 10: Overview Of The Lancing Device

    OVERVIEW OF THE LANCING DEVICE Setting Up the Lancing Device A. Device cap B. Handle C. Needle cap D. Needle holder 1. Screw the device cap off. 2. Put the needle into the 3. Screw the needle cap and E. Device body needle holder until the bottom.
  • Page 11: Performing A Blood Glucose Test

    Hold the lancing device firmly against NOTE the side of your finger. Press the release Be sure to use SCAN G-427BS Test Strips, other Test button. You will hear a click, indicating Strips may not produce accurate results. that the puncture is complete.
  • Page 12 Reference Value NOTE You must wait for the end of 5-second count down to TIME OF DAY PEOPLE WITHOUT DIABETES see the data display Make sure the confirmation window of the Test Strip is Before breakfast (fasting) <100 mg/dL completely filled with your blood sample. Two hours after meals <140 mg/dL Source: American Diabetes Association.
  • Page 13: Control Solution Test

    While staying at the lab: You may press “M” button to select CTL mode, for selecting the CTL mode, please refer to the  Make sure that the samples for both tests test are taken “THE TWO MEASURING MODES” section. and tested within 15 minutes of each other.
  • Page 14: Viewing Results On The App And Meter

    NOTE 2. Press “M” again, and the most recent test results with date and time will display. Press “M” once more and the  Do not use the expired control solution. Only use the next most recent test results will appear. Each time you control solution, you can purchase this product press and release “M”, the Meter will recall up to your last through visits the website: www.taidoc.com.
  • Page 15: Care And Storage For Your System

    CARE AND STORAGE FOR YOUR SYSTEM available at www.amazon.com or via a Google search. The SCAN G-427B G-427B meter can be cleaned and  Handle the Meter with care. Dropping or throwing the disinfected by the user twice per day over the 5 year use life Meter may cause damage to the device.
  • Page 16: Specifications

    Meter for one minute. As for the disinfecting procedure, please follow the steps below: 1. Model: G-427B  Wiping the entire outer crust of the Meter with CLOROX 2. Machine size: 3.5 × 2.1 × 0.5 in (9.0 × 5.4 × 1.3 cm) (L×W×H) Healthcare Bleach Germicidal Wipes and keeps the 3.
  • Page 17: Display Messages

    DISPLAY MESSAGES PROBLEMS MESSAGE WHAT IT MEANS ACTION PROBLEMS CAUSES SOLUTIONS The message indicates very 1. Battery may be Blood glucose level is low blood sugar. You should installed improperly. 1.Install the Battery consult with your healthcare lower than 20mg/dl correctly.
  • Page 18: Performance Characteristics

    Between-run Precision PERFORMANCE CHARACTERISTICS Glucose concentration Precision 30~50 51~110 111~150 151~250 251~400 interval(mg/dL) YSI Mean (mg/dL) 41.9 94.0 131.1 199.2 325.3 Within Run Precision Mean (300 Venous Blood Tests per Glucose Level) 40.1 90.5 116.8 193.0 315.6 (mg/dL) Lot 1 Glucose concentration 30~50 51~110...
  • Page 19: Electrical And Safety Standards

    Lay-user Performance Study This meter complies with CISPR 11: Class B (Radiated Only). Your SCAN G-427B Meter result may vary slightly from you actual blood glucose value. This may be due to slight Emissions of the energy used are low and not likely to cause differences in technique and the natural variation in the test...
  • Page 20 Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.