CONTENTS 1.-GENERAL DESCRIPTION 2.-TRANSPORT AND HANDLING 3.-TECHNICAL FEATURES 4.-OPERATION 5.-INSTALLATION 5.1 Installation in the Multifunction Sensor Holder Red. 44-020 5.2 Connections 6.- START-UP 6.1 Sensor Conditioning 6.2 Sensor Calibration 7.- MAINTENANCE 7.1 Cleaning and Calibration 7.2 List of Parts 7.3 Electrode head replace 7.4 Problem-Cause-Solution CE DECLARATION OF CONFORMITY WARRANTY...
1.-GENERAL DESCRIPTION Free chlorine amperometric sensor for drinking water and water treatment. Specifically designed to determine the residual level of inorganic chlorine in water. The chlorine sensor is of the open cell type with no intermediate liquids for the electroche- mical reaction, thus facilitating installation and maintenance.
4.-OPERATION Amperometric analysis is based on the measurement of current intensity. This intensity is produced by the oxidation or reduction of an analyte when a suitable voltage is applied. In the case of free chlorine analysis: HClO + H + 2e Cl + H O As can be deduced from the above reaction, current intensity is proportional to the amount of hypochlorous acid which is present in the measured solution.
5.- INSTALLATION The sensor must be installed where it is possible to ensure a constant flow of water with no chance of air bubbles forming in the measuring cell. It is recommended that it be installed in the Multifunction Sensor Holder (Ref. 44-020), especially designed for this application, and equipped with the following:...
6.- START-UP 6.1 SENSOR CONDITIONING If it has not been used recently or if it is being connected for the first time, the sensor will require a conditionning time. Prior to sensor calibration, insert the sensor correctly into the sensor holder and let the water containing free chlorine flow for 24 hours to ensure that the cell is properly polarized.
7.- MAINTENANCE 7.1 CLEANING AND CALIBRATION Cleaning interval: Every 8 hours and after working for several hours without flow, with water without free waste chlorine, or over 3 mg/l. This cleaning frequency is maintained automatically with the WTR PRO control device. Calibration interval: The first time, after 24 hours.
7.2 LIST OF PARTS CODE NAME 44-015 Chlorine sensor electrode head 44-016 Chlorine sensor body 44104 Ring holder for electrodes base 44106 Hydrodynamic regulator 62300 O-Ring 19x3 FPM 64349 O-Ring 14x2.5 FPM 64410 O-Ring 19x2 FPM 44-016 64410 44-015 64349 44104 44106 62300...
7.3 ELECTRODES HEAD REPLACE Disassembling Not assembling Use a plastic tool to prevent electrodes damage. Do not damage the connection pins of the electrodes head. It is necessary to carry out the step 3 before the assembly, otherwise the connection pins of the electrode head will be damaged.
7.4 PROBLEM-CAUSE-SOLUTION PROBLEMS CAUSE SOLUTION Sensor fails to connect with - Check connections READING mg/l = 0, control device DOES NOT MATCH WITH DPD-1 MEASUREMENT Insufficient flow in sensor - Adjust the flow that holder, or the chlorine sensor reaches the sensor not in contact with the water holder.
Page 13
PROBLEMAS SOLUCIÓN CAUSA LECTURA SUPERIOR El sensor se ha calibrado - Repetir A LA MEDIDA DPD-1 sin esperar el tiempo acondicionamiento del suficiente de sensor y volver a calibrar acondicionamiento Reactivos DP-1 gastados - Repetir medida DPD-1 con reactivos nuevos Medida DPD-1 incorrecta - Incrementar el tiempo de debido a una muestra de...
Page 14
PROBLEMAS SOLUCIÓN CAUSA Interferencias eléctricas - Eliminar la fuente de la externas perturbación. Puede ser útil conectar el agua con una toma tierra. Interferencias de otros - No utilizar más de un elementos oxidantes oxidante para la desinfección el agua. El pH es inestable y no hay - Estabilizar el pH y medida de pH en el...
DECLARACIÓN CE DE CONFORMIDAD I.T.C S.L. Vallès, 26 Polígono Industrial Can Bernades-Subirà 08130 Santa Perpètua de Mogoda Declara que el Sensor de Cloro Libre, identificado con número de serie y año de fabricación cumple la Directiva de Compatibilidad Electromagnética D89/336/CE) siempre que la instalación, el uso y el mantenimientos se efectúen de acuerdo de acuerdo con la normativa vigente y siguiendo las indicaciones del manual de instrucciones.