McKesson 413 Series Owner's Manual

McKesson 413 Series Owner's Manual

Advertisement

Quick Links

Digital Thermometers
OWNER'S MANUAL
For use with MFR #'s 16-412GM-00, 16-413BGM-00, 16-413RGM-00,
16-413BGM, 16-413RGM, 16-413BKGM, 16-413RKGM, 16-415GMHT,
and 16-415GMST.

WARNING

• Read instructions thoroughly before using digital thermometer.
• Choking Hazard: Thermometer cap and battery may be fatal if swallowed. Do
not allow children to use this device without parental supervision.
• Do not use thermometer in ear. Designed use is for oral, rectal, and armpit
(axilla) readings only.
• Do not place thermometer battery near extreme heat as it may explode.
• Note: Use of the probe cover may result in a 0.2ºF(0.1ºC) discrepancy from
actual temperature.
• The use of temperature readings for self-diagnosis is dangerous. Consult
your doctor for the interpretation of results. Self-diagnosis may lead to the
worsening of existing disease conditions.
• Do not attempt measurements when the thermometer is wet as inaccurate
readings may result.
• Do not bite the thermometer. Doing so may lead to breakage and/or injury.
• Do not attempt to disassemble or repair the thermometer. Doing so may
result in inaccurate readings.
• After each use, disinfect the thermometer especially in case the device is
used by more than one person.
• Do not force the thermometer into the rectum. Stop insertion and abort the
measurement when pain is present. Failure to do so may lead to injury.
• For children who are two years old or younger, please do not use the devices
orally.
• If the unit has been stored at temperatures over 104ºF (40ºC), leave it in
41ºF~104ºF (5ºC~40ºC) ambient temperature for about 15 minutes before
using it.
PLEASE READ CAREFULLY BEFORE USING
This digital thermometer provides a quick and highly accurate reading of an
individual's body temperature. The digital thermometer is intended to measure
the human body's temperature in regular mode orally, rectally or under the
arm, and the device is reusable for clinical or home use on people of all ages.
To better understand its functions and to provide years of dependable results,
please read all instructions first.
This appliance conforms to the following standards:
ASTM E1112 Standard Specification for Electronic Thermometer for
Intermittent Determination of Patient Temperature,
ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular
requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement,
IEC 60601-1-11 Medical electrical equipment —Part 1-11: General
requirements for basic safety and essential performance –Collateral Standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment and complies with the requirements
of IEC 60601-1-2(EMC), AAMI/ANSI ES60601-1(Safety) standards. And the
manufacturer is ISO 13485 certified.

CONTENTS

MFR #s 16-412GM-00, 16-413BGM-00, 16-413RGM-00, 16-413BGM,
16-413RGM, 16-415GMHT, and 16-415GMST come with 1 thermometer, 1
owner's manual, 1 storage case, and 5 probe sheaths.
MFR #s 16-413BKGM and 16-413RKGM come with 1 thermometer, 1 owner's
manual, 1 storage case, and 20 probe sheaths.

PRODUCT ILLUSTRATION

16-412GM-00, 413 SERIES
Probe
Battery Cap
On/Off Button
LCD
415 SERIES

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 413 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for McKesson 413 Series

  • Page 1 1 storage case, and 5 probe sheaths. MFR #s 16-413BKGM and 16-413RKGM come with 1 thermometer, 1 owner’s manual, 1 storage case, and 20 probe sheaths. PRODUCT ILLUSTRATION 16-412GM-00, 413 SERIES 415 SERIES Battery Cap On/Off Button Probe...
  • Page 2 PRECAUTION *The performance of the device may be degraded should one or more of the following occur: • Operation outside the manufacturer’s stated temperature and humidity range. • Storage outside the manufacturer’s stated temperature and humidity range. • Mechanical shock (for example, drop test) or degraded sensor. •...
  • Page 3 b) Rectal Use: Lubricate silver probe tip with petroleum jelly for easy insertion. Gently insert sensor approximately 1cm (less than 1/2 in) into rectum. Normal temperature between 97.2ºF and 99.9ºF (36.2ºC and 37.7ºC). c) Armpit Use: Wipe armpit dry. Place probe in armpit and keep arm pressed firmly at side.
  • Page 4 • Consult the dealer or an experienced radio/TV technician for help. Disposal of this product and used batteries should be carried out in accordance with the national regulations for the disposal of electronic products. Questions? Call 1-800-777-4908 Distributed by McKesson Medical-Surgical Inc. Richmond, VA 23233 PVN A0318 Made in USA...

This manual is also suitable for:

415 series16-412gm-0016-413bgm-0016-413rgm-0016-413bgm16-413rgm ... Show all

Table of Contents