Page 2
GROUND REFERENCE. NO OTHER CONNECTION POINTS ON THE LISN CABINET MUST BE USED FOR SAFETY PURPOSE. NOTICE: The AFJ LISN LT32C/10 is supplied in proper special packaging due to its heavy weight. We strongly recommend you to save packaging for next transport. Distributed by:...
Page 3
3. Make sure that victim has plentiful breathable air around. 4. Wind the victim with a quilt or garments as soon as possible, keeping the patient warm and calm waiting for aid arrival. AFJ LT32C/10 LISN - Operating Manual - rev. 3.8 / 2019-01...
Page 4
AFJ Instruments Warranty Centre, at Customer’s charge. All information in this manual is given in good faith: however, AFJ Instruments shall not be liable for any loss or damage whatsoever, arising from the use of this manual, the products described herein, or any error or omission in either.
Page 5
EN 61000-4-5: 1995 Copyright Acknowledgement: CISPR and EN publications are subject to the Copyright of IEC. Distributed by: Reliant EMC LLC 3311 Lewis Ave, Signal Hill CA 90755, 408-916-5750, www.reliantemc.com AFJ LT32C/10 LISN - Operating Manual - rev. 3.8 / 2019-01...
This device was built and tested according to CISPR 16-1-2. LISN has been supplied from the factory in a technical faultless state. LISN's from AFJ Instruments may use only on power supplies secured by fuses with prescribed safety devices. With no safety condition as:...
A further task of the LISN is the separation of the DUT from the mains concerning the high frequency interference's. Therefore the construction of the LISN LT32C/10 has a simulation impedance of: (50µH + 5 Ohm) // 50 Ohm. The term in brackets is the impedance of the LISN.
The compact form of construction, despite the high current-bearing capacity, makes easy use of the AFJ LISN LT32C/10 for the measurement of high currents directly at the piece of use of the consumer possible. In this way, measurements of mains-borne interferences can be carried out under conditions corresponding to practice.
-20°C to +70°C Dimensions 480 x 550 x 280 mm (W x L x H) Weight 28kg Distributed by: Reliant EMC LLC 3311 Lewis Ave, Signal Hill CA 90755, 408-916-5750, www.reliantemc.com AFJ LT32C/10 LISN - Operating Manual - rev. 3.8 / 2019-01...
LISN, via the appropriate logic signal cable supplied. In such case, totally automatic control of the LISN can be achieved, via the instruments software interface. AFJ LISN LT32C/10 offers current meter to read and measure Switching Operations and EUT absorption current, as per CISPR 14-1 requirements (see Section E).
The protection capabilities of the non-fused protective earth still hold. Distributed by: Reliant EMC LLC 3311 Lewis Ave, Signal Hill CA 90755, 408-916-5750, www.reliantemc.com AFJ LT32C/10 LISN - Operating Manual - rev. 3.8 / 2019-01...
150kHz and 30MHz also has a distinct positive influence on the interference reflection above this range of frequencies with a corresponding design and arrangement of the mains filters. AFJ Instruments will be pleased to help you with the design of a suitable mains interference elimination filter. Distributed by:...
Above 3MHz, the coupling becomes increasingly effective, and external perturbations affect the measurement to a lesser and lesser extent. Distributed by: Reliant EMC LLC 3311 Lewis Ave, Signal Hill CA 90755, 408-916-5750, www.reliantemc.com AFJ LT32C/10 LISN - Operating Manual - rev. 3.8 / 2019-01...
Switching operations are defined in CISPR 14-1 as EUT supply current change; no other specifications are reported. AFJ LISN LT32C/10 offers current meter to read and measure Switching Operations and EUT absorption current, as per CISPR 14-1 requirements. Example of appliances for which the click rate N is derived from the number of switching operations: - Thermostats for room heating equipment;...
Page 17
It is important to keep all connections within the adapter as short as possible, to avoid measuring errors at high frequency. Distributed by: Reliant EMC LLC 3311 Lewis Ave, Signal Hill CA 90755, 408-916-5750, www.reliantemc.com AFJ LT32C/10 LISN - Operating Manual - rev. 3.8 / 2019-01...
SECTION G: LT32C/10 REMOTE CONTROL G.1 • GENERAL To use AFJ LISN LT32C/10 in remote mode, input signal REM has to be grounded, so that REM signal (visualized by LED) became active The signal S0 and S1 coincidence define the line selection as per the following list (Table G.1):...
Need help?
Do you have a question about the LT32C/10 and is the answer not in the manual?
Questions and answers