GMI D1062S Instruction Manual

GMI D1062S Instruction Manual

Vibration transducer interface din-rail

Advertisement

Quick Links

D1062S
INSTRUCTION MANUAL
Vibration Transducer Interface
Din-Rail Model D1062S
D1062 - Vibration Transducer Interface
ISM0094-7

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the D1062S and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for GMI D1062S

  • Page 1 D1062S INSTRUCTION MANUAL Vibration Transducer Interface Din-Rail Model D1062S D1062 - Vibration Transducer Interface ISM0094-7...
  • Page 2 Characteristics General Description: The single channel DIN Rail Vibration Transducer Interface D1062S provides a fully floating dc supply for energizing vibration transducers, accelerometers or 2-3 wires sensors located in Hazardous Area, and repeats the sensor input voltage in a totally isolated circuit located in Safe Area to drive vibration monitors or analyzers for rotating machinery control and supervision purposes.
  • Page 3 Ordering information Model: D1062S Power Bus enclosure Front Panel and Features  Input from Zone 0 (Zone 20), Division 1, installation in Zone 2, Division 2.  0 to –20 V Input/Output Signal.  Wide band signal transfer.  Input and Output short circuit proof.
  • Page 4 Parameters Table In the system safety analysis, always check the Hazardous Area/Hazardous Locations devices to conform with the related system documentation, if the device is Intrinsically Safe check its suitability for the Hazardous Area/Hazardous Locations and gas group encountered and that its maximum allowable voltage, current, power (Ui/Vmax, Ii/Imax, Pi/Pi) are not exceeded by the safety parameters (Uo/Voc, Io/Isc, Po/Po) of the D1062 Associated Apparatus connected to it.
  • Page 5 NON HAZARDOUS LOCATIONS, CLASS I, DIVISION 2, CLASS II, DIVISION 1, GROUPS E, F, G, CLASS III, DIVISION 1, GROUPS A, B, C, D T-Code T4, CLASS I, ZONE 2, GROUP IIC T4 CLASS I, ZONE 0, GROUP IIC MODEL D1062S Programmable Current Source...
  • Page 6 NON HAZARDOUS LOCATIONS, CLASS I, DIVISION 2, CLASS II, DIVISION 1, GROUPS E, F, G, CLASS III, DIVISION 1, GROUPS A, B, C, D T-Code T4, CLASS I, ZONE 2, GROUP IIC T4 CLASS I, ZONE 0, GROUP IIC MODEL D1062S Current Programmable Selection...
  • Page 7 Warning D1062 is an isolated Intrinsically Safe Associated Apparatus installed into standard EN50022 T35 DIN Rail located in Safe Area/ Non Hazardous Locations or Zone 2, Group IIC, Temperature Classification T4, Class I, Division 2, Groups A, B, C, D, Temperature Code T4 and Class I, Zone 2, Group IIC, IIB, IIA Temperature Code T4 Hazardous Area/Hazardous Locations (according to EN/IEC60079-15, FM Class No.
  • Page 8 Installation D1062 is a vibration transducer interface housed in a plastic enclosure suitable for installation on T35 DIN Rail according to EN50022. D1062 unit can be mounted with any orientation over the entire ambient temperature range, see section “Installation in Cabinet” and "Installation of Electronic Equipments in Cabinet" Instruction Manual D1000 series for detailed instructions.