Summary of Contents for Stryker System 5

  • Page 2 INSTRUCTIONS FOR USE REF 4208 US Patents: 5,263,972; 5,747,953; 6,013,991 4100 E. Milham Kalamazoo, Michigan (USA) 49001 European Authorized Rep: 269-323-7700 RA/QA Manager 1-800-253-3210 Stryker France ZAC Satolas Green Pusignan Av. de Satolas Green 69881 MEYZIEU Cedex 10/05 4208-001-700 Rev-C www.stryker.com France...
  • Page 3: Intended Use

    Important Information Intended Use The Stryker System 5 Battery Powered Heavy Duty Sag- WARNING - CAUTION - NOTE ittal Saw is an oscillating cutting device used for cutting away bone and bone related tissue. Please read this manual and follow its instructions carefully.
  • Page 4 • DO NOT immerse the handpiece or battery pack. Wa- use. ter may enter the casing and damage the electrical components. *For further information, contact your Stryker Instruments sales representative or Customer Service at 1-800- 253-3210. Outside the USA, contact your nearest Stryker subsidiary.
  • Page 5: Features And Functions

    Features and Functions • Battery Latch - Depress the latch to release the battery pack from the handpiece. • Battery Pack - Contains a rechargeable battery that provides power to the handpiece. • Blade Lock Button - The button allows the insertion and release of the blade. •...
  • Page 6 Instructions Blade Installation Sagittal Head Positioning 1. Pull the sagittal head forward and away from the WARNING: handpiece body; then rotate it to the desired cutting angle (see figure 2). Always place the trigger switch in the SAFE position before attaching or removing any accessory to prevent NOTE: The sagittal head has eight possible cutting inadvertent running of the handpiece.
  • Page 7 Instructions (cont’d) 2. Test the operation of the handpiece by rotating the Battery Pack Installation trigger switch to either the STD or FAST position and squeezing the trigger. NOTES: 3. Set the trigger switch to the SAFE position until you •...
  • Page 8 *This product is not field repairable. In case of operating difficulties, all Stryker products must be returned to Stryker Instruments for repair. For more information, contact your Stryker Instruments sales representative or call Stryker Customer Service at 1-800-253-3210. Outside the...
  • Page 9: Periodic Maintenance

    Storage and Handling To ensure the longevity, performance and safety of this equipment, package in original packaging materials when storing or transporting. Periodic Maintenance INTERVAL ACTIVITY Prior to each use. Inspect, operate and test the handpiece to ensure that it is working properly.
  • Page 10: Cleaning Recommendations

    Cleaning Recommendations 3. Rinse instruments under running water. Hold the WARNINGS: handpiece upright to prevent water from running into the battery contact area. • Handpieces and batteries must be cleaned and steril- ized before first and every use. 4. If water leaks into the handpiece, tip the handpiece •...
  • Page 11: Sterilization Recommendations

    Sterilization Recommendations WARNINGS: • Handpieces and batteries must be cleaned and ETO: sterilized before first and every use. • 100% ETO • Remove the blade and battery from the handpiece • 120-135°F (49-57°C) prior to sterilization. • Wrapped in an instrument tray or fully perforated sterilization box •...
  • Page 12: Specifications

    Specifications* Model: REF 4208 Sagittal Saw Size: 8.9 in. [226 mm] height (with battery) 1.6 in. [41 mm] width 6.5 in. [165 mm] length Weight: 3.4 lbs. [1.54 kg] (with battery) Speed: 10,000 CPM (STD mode); 12,000 CPM (FAST mode) Excursion: 5°...
  • Page 13 Guidance and manufacturer’s declaration - electromagnetic emissions The System 5 handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of the System 5 handpiece should assure that it is used in such an environment.
  • Page 14 Guidance and manufacturer’s declaration - electromagnetic immunity The System 5 handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of the System 5 handpiece should assure that it is used in such an environment.
  • Page 15 Guidance and manufacturer’s declaration - electromagnetic immunity The System 5 handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of the System 5 handpiece should assure that it is used in such an environment.
  • Page 16 Recommended separation distances between portable and mobile RF communications equipment and the System 5 handpiece The System 5 handpiece is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the System 5 handpiece can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the System 5 handpiece as recommended below, according to the maximum output power of the communications equipment.
  • Page 17 On request, Stryker Instruments will provide a loaner unit for your use while repairs are being made. Please clean and sterilize all potentially contaminated products being sent in for repair, credit, or return of a loaner unit. The policy of Stryker Instruments is not to accept or process potentially contaminated products that do not meet this requirement.