Please read this manual thoroughly before operating the centrifuge. 2. INTENDED USE This Westlab Smart Micro Centrifuge is used in various laboratories to separate particles from suspension or even macromolecules from solutions according to their density. The different biological substances that are usually separated by centrifugation are microbial cells, organelles, proteins, DNA, and RNA.
Rotor Capacity PCR Strip Rotor Speed Setting USER MANUAL 2000 - 6000 RPM Speed Setting 2000 x g (MAX RCF) CODE 663-958C Speed Accuracy ± 25 RPM Time 0 To 25 Mins & Infinite Size (W x D x H)
ON/OFF SWITCH POWER SOCKET This Westlab Smart Micro Centrifuge is supplied in a packaging box. Open the box, gently remove the centrifuge from the box and plug in the power adaptor pin before connecting to the main power source. The instruction manual and the accessories should be kept with the centrifuge.
CENTRIFUGE for device & user safety. 5. The following are the incorrect methods of loading USER MANUAL tubes in the centrifuge the rotor. CODE 663-958C 0.2 ML 0.4 ML 0.5 ML 1.5 ML / 2.0 ML 1.5 ML / 2.0 ML 1.5 ML / 2.0 ML...
CENTRIFUGE that the rotor is tight enough to avoid any accidents. USER MANUAL Note: Tightening the screw will be in clockwise direction and the loosening will be in counter-clockwise direction CODE 663-958C Sticker Mark Grub Screw To tighten Rotor To loosen...
WESTLAB Push Down SMART MICRO RPM(X100) CENTRIFUGE Lid Release Button USER MANUAL Correct Method CODE 663-958C To open the centrifuge lid, press the “Lid Release” Push Down button down. Lid Release Correct Method Button 13. OPERATION To operate this Micro Centrifuge, the below steps should be followed: 1.
Page 9
The time increments/decrements will be in USER MANUAL minutes and will show on display and indicator. CODE 663-958C Time indicator display Infinite time display The timer can be set from 00 to 25 minutes. If the user...
Page 10
If the display shows“ ” that If the display shows “ ”it means the imbalance detection means the imbalance detection CODE 663-958C is activated is de-activated User can activate and deactivate the unbalance detection function by long pressing the timer knob as shown in above picture.
Resetting is done by switching `Off’ and again switching `On’ the centrifuge by using the On/Off switch located on the rear side of the centrifuge. CODE 663-958C 14. CARE & MAINTENANCE This personal centrifuge and its rotor should be cleaned periodically.
Page 12
The customer shall be responsible for the product or use of products as well as any supervision required for safety. If requested the products must be CODE 663-958C returned to the distributor in well packed and insured manner and all shipping charges must be paid.