INSTALLATION & OPERATING MANUAL SELECTORR SERIES CLAW VACUUM PUMPS C30 ® Please read the manual before operating the vacuum pump. TABLE OF CONTENTS 1.0 INSTALLATION General Description Unpacking Location Power Requirements Vacuum Connections Oil Filling on Gear Box 2.0 SAFETY General Notices Warning Labels and Explanation Location of Labels...
INSTALLATION AND OPERATING MANUAL This manual covers the C30 contact-less operating claw type vacuum pumps. The model number is stamped into the nameplate with serial number: C30. Please identify the model number and serial number when ordering parts. 1.0 INSTALLATION General Description The C30 vacuum pump is a dry and contactless machine, enclosed in an acoustic sound shield, designed to have cooling air passed through the sound shield by a fan.
If the pump is supplied with a motor starter, it is preset at the factory according to customer specifications. It is advisable to check that these settings are in line with the voltage at your location. If the voltage is different, please contact Ohio Medical for motor and starter information.
Remove the plastic protective cap from the inlet port prior to connection of pump to the system. Should process gas contain dust or other foreign particles, a suitable inlet filter should be connected to the inlet port. Consult Ohio Medical for recommendations or refer to the “Vacuum Pump Filters” tab at www.ohiomedicalparts.com.
• Only authorized and trained operators should operate the vacuum pump. • If the pump is not properly working, it should be stopped immediately. • Ohio Medical shall have no liability for any accident or failure arising from non- compliance with instructions in this manual.
The standard version is for use of dry air only, and may not be used in hazardous areas. Handling of humid air or gases containing aggressive chemicals is possible only with specially configured units. Consult Ohio Medical for assistance. Excessive back pressure on the unit may result in excessive current draw.
MAINTENANCE The C30 vacuum pump requires very little maintenance. To ensure optimum performance, the following maintenance steps should be followed: Gear Box Oil 4.1.1 Oil Level Check the oil level on a monthly basis. Under normal circumstances, it should not be necessary to add oil between oil changes. A significant drop in oil level means there is an oil leak.
PROBLEM SOLVING Problem: The pump does not reach capacity. 5.1.1 Possible Cause: Inlet screen (mesh) of the inlet filter clogged with debris. Remedy: check inlet filter element and clean screen (mesh) by compressed air or wash it. 5.1.2 Possible Cause: Pipe work is too long or small. Remedy: Use a bigger diameter pipe and shorten the lines length if possible.
Problem: The pump is running abnormally hot. 5.6.1 Possible Cause: Not enough air ventilation to pump. Remedy: Make certain a sufficient amount of fresh air is supplied to the pump. Check for dirty or blocked mesh, clean mesh as necessary. Problem: The pump will not operate (seized up).
Page 13
C30 PART LIST Description Gasket, Inlet Flange Silencer Inline Gear Box Housing Gear Box Cover (rear) Accoustic Mat for Shield Cover, End Round Head Bolt / M5 x 12 Accoustic Mat for Shield Cover, Front Washer, Spring Lock, 6mm Pump Housing Cover 1 (End Plate) Plate (vertical) Washer, Spring Lock, 16mm Fan Housing...